2-Methoxyethanol (Methyl Cellosolve, 2ME)
2-Methoxyethyl Acetate (Methyl Cellosolve Acetate, 2MEA)
2-Ethoxyethanol (Cellosolve, 2EE)
2-Ethoxyethyl Acetate (Cellosolve Acetate, 2EEA)

OSHA Method ORG-53 | January 1985

For problems with accessibility in using figures and illustrations, please contact the Salt Lake Technical Center at 801-233-4900.
These procedures were designed and tested for internal use by OSHA personnel. Mention of any company name or commercial product does not constitute endorsement by OSHA.

The tables on this page are best viewed on a desktop, notebook, or tablet computer.

Method no.: 53
Matrix: Air
Procedure: Samples are collected by drawing known volumes of air through charcoal tubes. Samples are desorbed with 95/5 (v/v) methylene chloride/methanol and analyzed by gas chromatography using a flame ionization detector.
Recommended air volume
and sampling rate:

10 L at 0.1 L/min
  2ME 2MEA 2EE 2EEA
OSHA PEL* ppm(mg/m3) 25(80) 25(120) 200(740) 100(540)
Target conc. ppm(mg/m3) 5(16) 5(24) 5(19) 5(27)
Reliable quantitation limits:
ppm(mg/m3)
0.1(0.3) 0.01(0.05) 0.02(0.06) 0.01(0.05)
Standard error of estimates
at target concentrations:
(Section 4.4)
5.4% 5.2% 5.4% 5.1%
*All have "Skin" notation.
Special requirements: Samples for 2MEA and 2EEA should be refrigerated upon receipt by the laboratory to minimize hydrolysis.
Status of method: Evaluated method. This method has been subjected to the established evaluation procedures of the Organic Methods Evaluation Branch.
Date: January 1985

Chemist: Carl J. Elskamp

Organic Methods Evaluation Branch
OSHA Analytical Laboratory
Salt Lake City, Utah

  1. General Discussion
    1. Background
      1. History

        The procedures used by OSHA to determine airborne concentrations of 2ME, 2MEA, 2EE, and 2EEA are based on NIOSH methods S79(Ref. 5.1), S39(Ref. 5.2), S361(Ref. 5.3), and S41(Ref. 5.4). All four of these methods specify collection on activated charcoal sampling tubes and analysis by gas chromatography (GC) using flame ionization detection. The methods for 2ME and 2EE specify 95/5(v/v) methylene chloride/methanol for the desorption solvent, while the methods for 2MEA and 2EEA utilize carbon disulfide. NIOSH validated these methods over the approximate ranges of 10 to 50 ppm 2ME, 10 to 50 ppm 2MEA, 100 to 400 ppm 2EE, and 50 to 200 ppm 2EEA.

        ACGIH has proposed to reduce the time-weighted TLVs for these four analytes to 5 ppm (Refs. 5.5 - 5.8), thus target concentrations at this level were chosen for this evaluation. In NIOSH's evaluation of the 2MEA method, it was reported that the desorption efficiency dropped off with mass of analyte collected. (Ref. 5.9) At a loading of 4.83 mg the average desorption efficiency was 81.9%. at 2.42 mg it was 80.9%, and at 1.21 mg it was 72.3%. For 10-L air samples, this range equals 100 down to 25 ppm. Similar results were reported for 2EEA. (Ref. 5.10) At a loading of 10.7 mg the average desorption was 84.2%, at 5.26 mg it was 83.9%, and at 2.53 mg it was 68.0%. For 10-L air samples, this range equals 200 down to 50 ppm. Since the selected target concentration of 5 ppm is much lower than these two ranges, it was anticipated that the desorption efficiency at this level would be unacceptably low, thus a better desorption solvent was needed for 2MEA and 2EEA.

        An average desorption efficiency of 97.6% was reported by NIOSH for the 2ME method over the range of 65 to 260 ppm for 10-L air samples with no apparent decrease with decreasing concentration. (Ref. 5.11) Thus a methylene chloride/methanol mixture was considered for a desorption solvent for all four analytes.

        Initial tests were done to determine if the recommended 95/5 (v/v) mixture was the optimum ratio of methylene chloride to methanol. Desorption studies were done on all four analytes using various mixtures if the two solvents ranging from 99/1 (v/v) methylene chloride/methanol to pure methanol. The 95/5 and 90/10 mixtures gave essentially the same efficiencies of around 100% for all four analytes, while the efficiencies for 2ME and 2EE were lower for a 99/1 mixture and the efficiencies for 2MEA and 2EEA were lower for a 50/50 mixture. The desorption efficiencies for all four analytes were much lower when pure methanol was used compared to the 95/5 and 90/10 mixtures. To maintain consistency with the NIOSH methods for 2ME and 2EE, a 95/5 mixture was chosen.

        Many GC column packings that are used for analysis of solvent samples are suitable for glycol ethers. These include 20% SP-2100/0.1% CW1500 on 80/100 Supelcoport, 10% SP-1000 on 80/100 Supelcoport, and 0.1% SP-1000 on Carbopack C. Another packing that was found to work particularly well was 0.2% CW1500 on 80/100 Carbopack C. This was used throughout the evaluation. For convenience in packing, a glass column was used. It is believed that a stainless steel column will perform just as well.

        Breakthrough studies were performed as discussed in Section 2.4. The breakthrough volumes are much higher than the recommended sample volume, but to maintain consistency with the many other solvent methods that specify charcoal sampling tubes, 10 L was chosen. This also provides a safety margin when other solvents are present in the sampled atmosphere. The presence of other solvent vapors will diminish the total capacity of the charcoal for the analytes.

        All four analytes were found to be stable on charcoal except when 2MEA and 2EEA were stored at ambient temperatures. Under this condition there is some loss of these two analytes through hydrolysis. This was confirmed by analyzing some of the 2MEA samples for 2ME, and 2EEA samples for 2EE, that had been stored at ambient temperatures for 15 days. The amounts of 2ME and 2EE found accounted for the corresponding loss of 2MEA and 2EEA. Although the loss is only on the order of 10 to 15% after 15 days compared to Refrigerated samples, it is recommended that samples for 2MEA and 2EEA be refrigerated until analyzed as a precautionary measure.

        All four analytes were successfully evaluated by collection on charcoal, desorption with 95/5 (v/v) methylene chloride/methanol, and analysis by GC using flame ionization detection.

      2. Toxic effects (This section is for information only and should not be taken as basis for OSHA policy.)

        As reported in the Documentation of Threshold Limit Values (Refs. 5.5 - 5.8), all four analytes were investigated by Nagano et al (Ref. 5.12) in terms of potency for testicular effects. They concluded that on an equimolar basis, the respective acetate esters were about as potent as 2ME and 2EE in producing testicular atrophy and leukopenia (an abnormally low number of white blood cells) in mice. Based on this study and because 2MEA and 2EEA hydrolyze to 2ME and 2EE respectively in the body, ACGIH suggests lowering the time-weighted TLVs for all four analytes to 5 ppm.

        The following is quoted from NIOSH Current Intelligence Bulletin 39. (Ref. 5.13)

        The National Institute for Occupational Safety and Health (NIOSH) recommends that 2-methoxyethanol (2ME) and 2-ethoxyethanol (2EE) be regarded in the workplace as having the potential to cause adverse reproductive effects in male and female workers. These recommendations are based on the results of several recent studies that have demonstrated dose-related embryotoxicity and other reproductive effects in several species of animals exposed by different routes of administration. Of particular concern are those studies in which exposure of pregnant animals to concentrations of 2ME or 2EE at or below their respective Occupational Safety and Health Administration (OSHA) Permissible Exposure Limits (PELs) led to increased incidences of embryonic death, teratogenesis, or growth retardation. Exposure of male animals resulted in testicular atrophy and sterility. In each case the animals had been exposed to 2ME or 2EE at concentrations at or below their respective OSHA PELs. Therefore, appropriate controls should be instituted to minimize worker exposure to both compounds.

      3. Workplace exposure

        2ME- It is used as a solvent for many purposes: cellulose esters, dyes, resins, lacquers, varnishes, and stains; and as a perfume fixative and jet fuel deicing additive. (Ref. 5.5)

        2MEA- It is used in photographic films, lacquers, textile printing, and as a solvent for waxes, oils, various gums and resins, cellulose acetate, and nitrocellulose. (Ref. 5.6)

        2EE- It is used as a solvent for nitrocellulose, natural and synthetic resins, and as a mutual solvent for the formulation of soluble oils. It is also used in lacquers, in the dyeing and printing of textiles, in varnish removers, cleaning solutions, in products for the treatment of leather, and as an anti-icing additive for aviation fuels. (Ref. 5.7)

        2EEA- It is used as a blush retardant in lacquers; as a solvent for nitrocellulose, oils and resins; in wood stains, varnish removers, and in products for the treatment of textiles and leathers. (Ref. 5.8)

      4. Physical properties (Refs. 5.5 - 5.8)

         

        analyte 2ME 2MEA 2EE 2EEA
        mol wt: 76.09 118.13 90.11 132.16
        bp (°C): 124.5 145 135.6 156.4
        color: all are colorless liquids
        sp gr: 0.9663 1.005 0.931 0.975
        vp:
        (mm Hg at 20°C)
        6 2 3.7 2
        flash pt.:
        (°C, closed cup)
        43 49 40 49
        odor:
        (Ref. 5.14)
        mild
        non-residual
        mild,
        ether-like
        sweetish mild,
        non-residual
        explosive
        limits, %
        (Ref. 5.14)
          lower:
          upper:



        2.5
        19.8



        1.1
        8.2



        1.8
        14



        1.7
        ?

        synonyms: (Ref. 5.14)

        2ME - Methyl Cellosolve; Glycol monomethyl ether; Ethylene glycol monomethyl ether; Methyl oxitol; Ektasolve; Jeffersol EM

        2MEA - Methyl Cellosolve Acetate; Glycol monomethyl ether acetate; Ethylene glycol monomethyl; ether acetate

        2EE - Cellosolve solvent; Ethylene glycol monoethyl ether

        2EEA - Cellosolve acetate; Glycol monoethyl ether acetate; Ethylene glycol monoethyl ether acetate

        chemical formulae:

        • 2ME- CH3OCH2CH2OH
        • 2MEA- CH3OCH2CH2OOCCH3
        • 2EE- CH3CH2OCH2CH2OH
        • 2EEA- CH3CH2OCH2CH2OOCCH3
    2. Limit Defining Parameters (The analyte air concentrations listed throughout this method are based on an air volume of 10 L and a solvent desorption volume of 1.0 mL. Air concentrations listed in ppm are referenced to 25°C and 760 mm Hg.)
      1. Detection limits of the analytical procedure

        The detection limits of the analytical procedure are 2.6, 0.40, 0.47, and 0.46 ng per injection for 2ME, 2MEA, 2EE, and 2EEA respectively. These are the amounts of analytes which give measurable responses with the amounts of interferences present in standards. (Section 4.1)

      2. Detection limits of the overall procedure

        The detection limits of the overall procedure are 3.1, 0.48, 0.56, and 0.55 µg per sample for 2ME, 2MEA, 2EE, and 2EEA respectively. These are the amounts of analytes spiked on the sampling device which allow recovery of amounts of analytes equivalent to the detection limits of the analytical procedure. These detection limits correspond to air concentrations of 0.1 ppm(0.3 mg/m3), 0.01 ppm(0.05 mg/m3), 0.02 ppm(0.06 mg/m3), and 0.01 ppm(0.05 mg/m3) for 2ME, 2MEA, 2EE, and 2EEA respectively. (Section 4.2)

      3. Reliable quantitation limits

        The reliable quantitation limits are the same as the detection limits of the overall procedure since the desorption efficiencies are essentially 100% at these levels. These are the smallest amounts of analytes which can be quantitated within the requirements of recoveries of at least 75% and precisions (±1.96 SD) of ±25% or better. (Section 4.2)


        The reliable quantitation limits and detection limits reported in the method are based upon optimization of the instrument for the smallest possible amounts of analytes. When the target concentration of an analyte is exceptionally higher than these limits, they may not be attainable at the routine operating parameters.


      4. Sensitivity

        The sensitivities of the analytical procedure over a concentration range representing 0.5 to 2 times the target concentrations based on the recommended air volume are 180, 200, 240, and 230 area units per µg/mL for 2ME, 2MEA, 2EE, and 2EEA respectively. These are determined by the slopes of the calibration curves. (Section 4.3) The sensitivity will vary with the particular instrument used in the analysis.

      5. Recovery

        The recovery of 2ME, 2MEA, 2EE, and 2EEA from samples used in a 15-day storage test remained above 97, 97, 99, and 99% respectively. The 2ME and 2EE samples were stored in a closed drawer at ambient temperatures of 20 to 25°C and the 2MEA and 2EEA samples were stored in a refrigerator at 0°C. (Section 4.4) The recovery of analyte from the collection medium during storage must be 75% or greater.

      6. Precision (analytical method only)

        The pooled coefficients of variation obtained from replicate determinations of analytical standards at 0.5, 1, and 2 times the target concentrations are 0.009, 0.019, 0.021, and 0.019 for 2ME, 2MEA, 2EE, and 2EEA respectively. (Section 4.3)

      7. Precision (overall procedure)

        The precisions at the 95% confidence level for the 15-day storage tests are ±10.5, ±10.1, ±10.5, and ±10.0% for 2ME, 2MEA, 2EE, and 2EEA respectively. (Section 4.4) These include an additional ±5% for sampling error. The overall procedure must provide results at the target concentration that are ±25% or better at the 95% confidence level.

      8. Reproducibility

        Six samples for each analyte collected from controlled test atmospheres and a draft copy of this procedure were given to a chemist unassociated with this evaluation. The samples for 2ME, 2MEA, 2EE, and 2EEA were analyzed after refrigerated storage for 24, 24, 16, and 18 days respectively. The average recoveries were 105% (SD = 7.3%), 98.8% (SD = 0.5%), 99.3% (SD = 1.0%), and 97.6% (SD = 3.0%) for 2ME, 2MEA, 2EE, and 2EEA respectively. (Section 4.5)

    3. Advantages
      1. The solid sorbent tube provides a convenient method for sampling.
      2. The analysis is rapid, sensitive, and precise.
    4. Disadvantages
      1. Samples for 2MEA and 2EEA should be refrigerated until analyzed to minimize hydrolysis.
      2. It may not be possible to analyze co-collected solvents using this method. Most of the other common solvents are desorbed with carbon disulfide.
  2. Sampling Procedure
    1. Apparatus
      1. Samples are collected by use of a personal sampling pump that can be calibrated within ±5% of the recommended flow rate with the sampling tube in line.
      2. Samples are collected on solid sorbent sampling tubes containing coconut shell charcoal. Each tube consists of two sections of charcoal separated by a urethane foam plug. The front section contains 100 mg of sorbent and the back section 50 mg. The sections are held in place with glass wool plugs in a glass tube 4-mm i.d. × 70-mm length. For this evaluation, SKC Inc. sorbent tubes (catalog number 226-01, lot 120) were used.
    2. Reagents

      None required

    3. Technique
      1. Immediately before sampling, break off the ends of the charcoal tube. All tubes should be from the same lot.
      2. Connect the sampling tube to the sampling pump with flexible tubing. Position the tube so that sampled air first passes through the 100-mg section.
      3. Air being sampled should not pass through any hose or tubing before entering the sampling tube.
      4. Place the sampling tube vertically (to avoid channeling) in the employee's breathing zone.
      5. After sampling, seal the tubes immediately with plastic caps and wrap lengthwise with OSHA Form 21.
      6. Submit at least one blank sampling tube with each sample set. Blanks should be handled in the same manner as samples, except no air is drawn through them.
      7. Record sample volumes (in liters of air) for each sample, along with any potential interferences.
      8. Ship any bulk sample(s) in a container separate from the air samples.
    4. Breakthrough
      1. Individual breakthrough studies were performed on each of the four analytes by monitoring the effluent from sampling tubes containing only the 100-mg section of charcoal while sampling at 0.2 L/min from atmospheres containing 10 ppm analyte. The atmospheres were at approximately 80% relative humidity and 20-25°C. No breakthrough was detected in any of the studies after sampling for at least 6 h (>70 L).
      2. A similar study as in 2.4.1. was done while sampling an atmosphere containing 10 ppm of all four analytes. The atmosphere was sampled for more than 5 h (>60 L) with no breakthrough detected.
    5. Desorption efficiency
      1. The average desorption efficiencies of 2ME, 2MEA, 2EE, and 2EEA are 97.9, 100.2, 99.0, and 101.6% respectively over the range of 0.5 to 2 times the target concentrations. (Section 4.6)
      2. Desorption efficiencies must be determined for the lot of charcoal used for samples.
    6. Recommended air volume and sampling rate
      1. The recommended air volume is 10 L.
      2. The recommended sampling rate is 0.1 L/min.
    7. Interferences (sampling)
      1. It is not known if any compound(s) will severely interfere with the collection of any of the four analytes on charcoal. In general, the presence of any other contaminant vapors in the air will reduce the capacity of charcoal to collect the analytes.
      2. Suspected interferences should be reported to the laboratory with submitted samples.
    8. Safety precautions (sampling)
      1. Attach the sampling equipment to the employee so that it will not interfere with work performance or safety.
      2. Wear eye protection when breaking the ends of the charcoal tubes.
      3. Follow all safety procedures that apply to the work area being sampled.
  3. Analytical Procedure
    1. Apparatus
      1. A GC equipped with a flame ionization detector. For this evaluation, a Hewlett-Packard 5830A GC was used with a 7671A Automatic Sampler.
      2. A GC column capable of separating the analyte of interest from the desorption solvent and any interferences. A 6-ft × 2-mm i.d. glass column packed with 0.2% Carbowax 1500 on GP 80/100 Carbopack C was used in this evaluation.
      3. An electronic integrator or some other suitable method of measuring peak areas or heights.
      4. Small vials with Teflon-lined caps capable of holding 2 mL.
      5. A dispenser capable of delivering 1.0 mL to prepare standards and samples. If a dispenser is not available, a 1-mL pipet may be used.
      6. Syringes such as 5-µL for preparation of standards and 1-µL for injection of samples and standards into a GC.
      7. Volumetric flasks and pipets to dilute the pure analytes in preparation of standards.
    2. Reagents
      1. 2-Methoxyethanol, 2-methoxyethyl acetate, 2-ethoxyethanol, and 2-ethoxyethyl acetate, reagent grade.
      2. Methylene chloride, chromatographic grade.
      3. Methanol, chromatographic grade.
      4. A suitable internal standard such as butanol or hexanol, reagent grade.
      5. The desorption solvent consists of methylene chloride/methanol, 95/5 (v/v) containing an internal standard at a concentration of approximately 0.2 µL per mL.
      6. GC grade nitrogen, air, and hydrogen.
    3. Standard preparation
      1. Concentrated stock standards are prepared by diluting the pure analytes with methanol. Working standards are prepared by injecting microliter amounts of concentrated stock standards into vials containing 1.0 mL of desorption solvent delivered from a dispenser.
      2. Working standard concentrations should bracket sample concentrations. If samples fall outside of the concentration range of prepared standards, additional standards should be prepared and analyzed to ascertain the linearity of response.
    4. Sample preparation
      1. Transfer each section of the samples to separate vials. The glass tubes and plugs are discarded.
      2. Add 1.0 mL of desorption solvent to each vial using the same dispenser as used for preparation of standards.
      3. The vials are immediately capped and shaken periodically for 30 min before analysis.
    5. Analysis
      1. GC conditions
        zone temperatures (°C): column - 80 for 2ME
        150 for 2MEA
        100 for 2EE
        150 for 2EEA
        injector - 200
        detector - 250
        gas flows (mL/min): nitrogen - 18
        hydrogen - 28
        air - 240
        injection volume: 0.84 µL
        column: 6-ft × 2-mm i.d. glass, 0.2% Carbowax 1500 on GP 80/100 Carbopack C
        retention times (min):
        2ME - 5.2
        2MEA - 3.8
        2EE - 6.6
        2EEA - 8.8
        chromatograms: Section 4.7
      2. Peak areas (or heights) are measured by an integrator or other suitable means.
      3. A calibration curve is constructed by plotting response (areas or heights) of standard injections versus microliters of analyte per sample. Sample concentrations must be bracketed by standards.
    6. Interferences (analytical)
      1. Any compound that responds on a flame ionization detector and has the same general retention time of the analyte or internal standard is a potential interference. Possible interferences should be reported to the laboratory with submitted samples by the industrial hygienist. These interferences should be considered before samples are desorbed.
      2. GC parameters (i.e. column and column temperature) may be changed to possibly circumvent interferences.
      3. Retention time on a single column is not considered proof of chemical identity. Analyte identity should be confirmed by GC/mass spectrometer if possible.
    7. Calculations

      The analyte concentration for samples is obtained from the appropriate calibration curve in terms of micrograms per sample, uncorrected for desorption efficiency. The air concentration is calculated using the following formulae. If any analyte is found on the backup section, it is added to the amount found on the front section. This total amount is then corrected by subtracting the total amount (if any) found on the blank.

      mg/m3 =

      (micrograms of analyte per sample)


      (liters of air sampled) (desorption efficiency)

      ppm =

      (mg/m3) (24.46)


      (molecular weight of analyte)

      where 24.46 = molar volume (liters) at 760 mm Hg and 25°C

      molecular weight of analyte = 76.09 for 2ME, 118.13 for 2MEA,
      90.11 for 2EE, 132.16 for 2EEA
    8. Safety precautions (analytical)
      1. Avoid skin contact and inhalation of all chemicals.
      2. Restrict the use of all chemicals to a fume hood if possible.
      3. Wear safety glasses and a lab coat at all times while in the lab area.
  4. Backup Data
    1. Detection limits of the analytical procedure

      The injection size recommended in the analytical procedure (0.84 µL) was used in the determination of the detection limits of the analytical procedure. The detection limits of 2.6 ng of 2ME, 0.40 ng of 2MEA, 0.47 ng of 2EE, and 0.46 ng of 2EEA were determined by making 0.84-µL injections of 3.09 ng/µL, 0.48 ng/µL, 0.56 ng/µL, and 0.55 ng/µL standards respectively. Chromatograms of such injections are shown in Figures 4.1.1 - 4.1.4.

    2. Detection limits of the overall procedure and reliable quantitation limits

      Six samples for each analyte were prepared by injecting 3.09 µg of 2ME, 0.48 µg of 2MEA, 0.56 µg of 2EE, and 0.55 µg of 2EEA into separate charcoal tubes. The samples were refrigerated and analyzed the next day to determine the amount recovered. Since the amounts recovered were high and approximately equal to the detection limits of the analytical procedure, the detection limits of the overall procedure and the reliable quantitation limits are taken to be 3.1, 0.48, 0.56, and 0.55 µg per sample for 2ME, 2MEA, 2EE, and 2EEA respectively. These limits correspond to air concentrations of 0.1 ppm (0.3 mg/m3), 0.01 ppm (0.05 mg/m3), 0.02 ppm (0.06 mg/m3), and 0.01 ppm (0.05 mg/m3) for 2ME, 2MEA, 2EE, and 2EEA respectively.

      Table 4.2.
      Detection Limit Data

      analyte
      µg/sample
      2ME
      3.1
      2MEA
      0.48
      2EE
      0.56
      2EEA
      0.55
      %
      recovered





      X
      SD
      1.96 SD
      96.1
      94.8
      97.1
      97.1
      93.5
      95.5

      95.7
        1.4
        2.7
      102
        95.9
        97.1
        90.5
        94.8
      107

        97.9
          5.8
        11.4
        96.4
      100
        98.2
      105
        98.2
        98.2

        99.4
          3.2
          6.3
      105
        97.1
      103
        ---
        95.4
        95.4

        99.2
          4.5
          8.8
    3. Sensitivity and Precision (analytical method only)

      The sensitivity and precision of the analytical procedure were determined from multiple injections of analytical standards. The sensitivities are the slopes of the calibration curves. These data are given in Tables 4.3.1. and 4.3.2. and in Figures 4.3.1 - 4.3.4.

      Table 4.3.1.
      Sensitivity and Precision Data for 2ME and 2MEA

      analyte

      2 ME

       

      2 MEA

      × target conc.
      µg/mL
      0.5×
      77.3

      154.6

      309.2
      0.5×
      120.6

      241.2

      482.4
      area counts






      X
      SD
      CV

      CV
      13050
      13080
      13330
      13390
      13330
      13330

      13252
      146.8
      0.011
      27290
      27410
      27080
      27130
      26950
      27150

      27168
      161.5
      0.006

      0.009
      55740
      55570
      55550
      54490
      55090
      55370

      55302
      454.8
      0.008
      23410
      22910
      23210
      23270
      23090
      23370

      23210
      186.3
      0.008
      46560
      46850
      47160
      46890
      46460
      46630

      46758
      257.6
      0.006

      0.019
        93860
        94140
      101000
        95000
        93960
        92600

        95093
        2994
        0.031

      Table 4.3.2.
      Sensitivity and Precision Data for 2EE and 2EEA

      analyte

      2 EE

      2 EEA

      × target conc.
      µg/mL
      0.5×
      93.1

      186.2

      372.4
      0.5×
      136.5

      273.0

      546.0
      area counts






      X
      SD
      CV

      CV
      21140
      20310
      21800
      21420
      21310
      21870

      21308
       564.7
       0.027
      44390
      44150
      43780
      43790
      44720
      44580

      44235
       397.6
       0.009

       0.021
      94640
      91580
      90160
      90160
      89500
      89460

      90917
        1979
       0.022
      30470
      29890
      30280
      30350
      30120
      30510

      30270
       232.8
       0.008
      60830
      31100
      61480
      61070
      60600
      60770

      60975
       310.9
       0.005

       0.019
      123000
      123300
      132000
      124100
      122800
      121300

      124417
          3826
         0.031
    4. Recovery data (storage)

      Storage samples were generated from test atmospheres (air at approximately 80% relative humidity and ambient pressures and temperatures) containing the analytes near the target concentration of 5 ppm. All four analytes were done separately. The samples were generated by drawing approximately 10 L of test atmospheres through charcoal tubes at about 0.2 L/min for 50 min. For each set of 36 samples generated for each analyte, six samples were analyzed immediately after generation, fifteen were stored in a refrigerator at 0°C and fifteen were stored in a closed drawer at ambient temperatures of 20-25°C. Six samples, three from refrigerated and three from ambient storage, were analyzed in three-day intervals over a period of fifteen days. The results are given in Tables 4.4.1.-4.4.4. and shown graphically in Figures 4.4.1 - 4.4.8.

      Table 4.4.1.
      Storage Data for 2ME


      storage time
      (days)

      % recovery

      (refrigerated)

      (ambient)

        0
        0
        3
        6
        9
      12
      15

      97.2
      98.2
      90.7
      100.4  
      94.9
      97.2
      98.4

      96.8
      99.1
      90.7
      93.7
      95.4
      98.0
      96.7

      99.0
      98.0
      91.0
      94.7
      93.4
      96.9
      98.5

      97.2
      98.2
      93.2
      95.9
      97.8
      99.3
      100.7  

      96.8
      99.1
      94.4
      99.9
      99.2
      98.8
      98.8

      99.0
      98.0
      93.1
      98.9
      98.4
      101.0  
      98.0

      Table 4.4.2.
      Storage Data for 2MEA


      storage time
      (days)

      % recovery

      (refrigerated)

      (ambient)

        0
        0
        3
        6
        9
      12
      15

      98.0
      100.8  
      97.5
      95.7
      97.0
      95.2
      97.2

      98.4
      102.1  
      99.6
      98.3
      98.6
      95.9
      97.9

      100.2  
      100.7  
      96.6
      98.2
      97.9
      96.9
      98.4

      98.0
      100.8  
      96.0
      88.6
      88.4
      86.4
      89.2

      98.4
      102.1  
      96.3
      89.8
      92.2
      86.4
      89.5

      100.2  
      100.7  
      95.8
      88.9
      90.1
      85.5
      88.2

      Table 4.4.3.
      Storage Data for 2EE


      storage time
      (days)

      % recovery

      (refrigerated)

      (ambient)

        0
        0
        3
        6
        9
      12
      15

      98.2
      100.5  
      94.4
      96.1
      93.1
      102.3  
      98.4

      98.6
      101.9  
      97.8
      96.5
      95.4
      102.3  
      100.4  

      99.1
      101.4  
      96.2
      96.6
      97.4
      103.0  
      101.5  

      98.2
      100.5  
      98.4
      97.8
      98.2
      103.4  
      100.5  

      98.6
      101.9  
      101.0  
      99.3
      99.0
      105.5  
      102.9  

      99.1
      101.4  
      102.1  
      98.8
      98.7
      104.1  
      101.5  

      Table 4.4.4.
      Storage Data for 2EEA


      storage time
      (days)

      % recovery

      (refrigerated)

      (ambient)

        0
        0
        3
        6
        9
      12
      15

      99.1
      100.2  
      99.9
      100.1  
      99.5
      99.3
      99.4

      99.4
      101.7  
      98.2
      100.0  
      97.4
      98.5
      96.8

      99.5
      99.9
      100.1  
      101.8  
      101.8  
      98.9
      97.4

      99.1
      100.2  
      96.0
      90.1
      89.6
      87.7
      86.3

      99.4
      101.7  
      99.3
      92.4
      91.9
      90.2
      87.4

      99.5
      99.9
      96.4
      90.2
      90.7
      88.3
      86.0

    5. Reproducibility data

      Six samples for each analyte, collected from controlled test atmospheres (about 80% R.H., 20-26°C, 645-660 mm Hg) containing the analyte near the target concentration, were analyzed by a chemist unassociated with this evaluation. The samples were generated by drawing the test atmosphere through the sampling tubes for 50 min at approximately 0.2 L/min. The samples were stored in a refrigerator for 24, 24, 16, and 18 days for 2ME, 2MEA, 2EE, and 2EEA respectively before being analyzed.

      Table 4.5.1.
      Reproducibility Data for 2ME and 2MEA



      sample
      no.

      2ME

      2MEA

      ppm
      found

      ppm
      theoretical

      %
      found

      ppm
      found

      ppm
      theoretical

      %
      found

         1
         2
         3
         4
         5
         6

      X
      SD

      6.53
      5.67
      5.66
      5.45
      5.69
      5.47

      5.46
      5.46
      5.46
      5.46
      5.46
      5.46

      119.6
      103.8
      103.7
        99.8
      104.2
      100.2

      105.0
          7.3

      5.84
      5.78
      5.84
      5.81
      5.85
      5.80

      5.89
      5.89
      5.89
      5.89
      5.89
      5.89

      99.2
      98.1
      99.2
      98.6
      99.3
      98.5

      98.8
        0.5

      Table 4.5.2.
      Reproducibility Data for 2EE and 2EEA



      sample
      no.

      2ME

      2MEA

      ppm
      found

      ppm
      theoretical

      %
      found

      ppm
      found

      ppm
      theoretical

      %
      found

         1
         2
         3
         4
         5
         6

      X
      SD

      6.29
      6.29
      6.44
      6.39
      6.30
      6.38

      6.39
      6.39
      6.39
      6.39
      6.39
      6.39

      98.4
      98.4
      100.8  
      100.0  
      98.6
      99.8

      99.3
        1.0

      6.13
      6.06
      6.13
      5.71
      6.10
      6.25

      6.21
      6.21
      6.21
      6.21
      6.21
      6.21

      98.7
      97.6
      98.7
      91.9
      98.2
      100.6  

      97.6
        3.0

    6. Desorption efficiency

      The desorption efficiency for each analyte was determined by injecting a known amount of analyte onto the front section of a charcoal tube. Eighteen samples were prepared for each analyte, six samples for each concentration listed below. The samples were stored in a refrigerator and analyzed the next day.

      Table 4.6.1.
      Desorption Efficiency Data for 2ME and 2MEAA

      analyte

      2 ME

      2 MEA

      µg/sample
      ppm

      77.3
      2.48

      154.6
        4.97

      309.2
        9.94

      120.6
        2.50

      241.2
        4.99

      482.4
        9.98

      desorption
      efficiency, %





      X

      X

      97.2
      96.8
      97.3
      96.1
      97.2
      99.5

      97.4

        96.5
        98.1
        95.8
        97.6
        96.7
        97.5

        97.0

        97.9

        98.7
        98.6
        99.1
        98.4
      103.0
        98.8

        99.4

        99.7
      103.7
      100.8
      103.7
      102.6
      101.5

      102.0

        99.3
      100.3
      100.2
        99.5
      100.3
        99.5

        99.9

      100.2

        98.6
        97.4
        99.6
        98.8
      101.4
        97.5

        98.9

      Table 4.6.2.
      Desorption Efficiency Data for 2EE and 2EEA

      analyte

      2 EE

      2 EEA

      µg/sample
      ppm

      93.1
      2.53

      186.2
        5.05

      372.4
        10.1

      136.5
        2.53

      273.0
        5.05

      546.0
        10.1

      desorption
      efficiency, %





      X

      X

        98.6
        99.1
        98.7
        98.4
        98.6
        98.6

        98.7

        99.0
        99.4
        99.0
        98.2
        99.7
        99.0

        99.0

        99.0

        98.2
        99.2
        99.9
        99.7
        99.7
        99.7

        99.4

      101.0
      104.5
      103.1
      101.2
      104.2
      102.9

      102.8

      101.2
      102.5
      102.1
      102.0
        99.7
      103.8

      101.9

      101.6

        99.6
      100.2
      101.2
      100.5
        99.5
        99.6

      100.1

    7. Chromatograms of the four analytes are shown in Figures 4.8.1 - 4.8.3. The chromatograms represent 0.84-µL injections of standards equivalent to 5-ppm air samples.

       

      Figure 4.1.1. Detection limit chromatogram for 2-methoxyethanol.

      Figure 4.1.1. Detection limit chromatogram for 2-methoxyethanol.

      Figure 4.1.2. Detection limit chromatogram for 2-methoxyethyl acetate.

      Figure 4.1.2. Detection limit chromatogram for 2-methoxyethyl acetate.

      Figure 4.1.3. Detection limit chromatogram for 2-ethoxyethanol.

      Figure 4.1.3. Detection limit chromatogram for 2-ethoxyethanol.

      Figure 4.1.4. Detection limit chromatogram for 2-ethoxyethyl acetate.

      Figure 4.1.4. Detection limit chromatogram for 2-ethoxyethyl acetate.

      Figure 4.3.1. Calibration curve for 2-methoxyethanol.

      Figure 4.3.1. Calibration curve for 2-methoxyethanol.

      Figure 4.3.2. Calibration curve for 2-methoxyethyl acetate.

      Figure 4.3.2. Calibration curve for 2-methoxyethyl acetate.

      Figure 4.3.3. Calibration curve for 2-ethoxyethanol.

      Figure 4.3.3. Calibration curve for 2-ethoxyethanol.

      Figure 4.3.4. Calibration curve for 2-ethoxyethyl acetate.

      Figure 4.3.4. Calibration curve for 2-ethoxyethyl acetate.

      Figure 4.4.1. 2-Methoxyethanol refrigerated storage samples.

      Figure 4.4.1. 2-Methoxyethanol refrigerated storage samples.

      Figure 4.4.2. 2-Methoxyethanol ambient storage samples.

      Figure 4.4.2. 2-Methoxyethanol ambient storage samples.

      Figure 4.4.3. 2-Methoxyethyl acetate refrigerated storage samples.

      Figure 4.4.3. 2-Methoxyethyl acetate refrigerated storage samples.

      Figure 4.4.4. 2-Methoxyethyl acetate ambient storage samples.

      Figure 4.4.4. 2-Methoxyethyl acetate ambient storage samples.

      Figure 4.4.5. 2-Ethoxyethanol refrigerated storage samples.

      Figure 4.4.5. 2-Ethoxyethanol refrigerated storage samples.

      Figure 4.4.6. 2-Ethoxyethanol ambient storage samples.

      Figure 4.4.6. 2-Ethoxyethanol ambient storage samples.

      Figure 4.4.7. 2-Ethoxyethyl acetate refrigerated storage samples.

      Figure 4.4.7. 2-Ethoxyethyl acetate refrigerated storage samples.

      Figure 4.4.8. 2-Ethoxyethyl acetate ambient storage samples.

      Figure 4.4.8. 2-Ethoxyethyl acetate ambient storage samples.

      Figure 4.8.1. Chromatogram of a 2-methoxyethanol standard.

      Figure 4.8.1. Chromatogram of a 2-methoxyethanol standard.

      Figure 4.8.2. Chromatogram of a 2-ethaoxyethanol standard.

      Figure 4.8.2. Chromatogram of a 2-ethaoxyethanol standard.

      Figure 4.8.3. Chromatogram of 2-methoxyethyl acetate and 2-ethoxyethyl acetate standards.

      Figure 4.8.3. Chromatogram of 2-methoxyethyl acetate and 2-ethoxyethyl acetate standards.

  5. References
    1. "NIOSH Manual of Analytical Methods", 2nd ed.; Department of Health, Education and Welfare, National Institute for Occupational Safety and Health: Cincinnati, OH, 1977; Vol. 2, Method No. S79; DHEW(NIOSH) Publ.(U.S.), No. 77-157-B.
    2. "NIOSH Manual of Analytical Methods", 2nd ed.; Department of Health, Education and Welfare, National Institute for Occupational Safety and Health: Cincinnati, OH, 1977; Vol. 2, Method No. S39; DHEW(NIOSH) Publ.(U.S.), No. 77-157-B.
    3. "NIOSH Manual of Analytical Methods", 2nd ed.; Department of Health, Education and Welfare, National Institute for Occupational Safety and Health: Cincinnati, OH, 1977; Vol. 5, Method No. S361; DHEW(NIOSH) Publ.(U.S.), No. 79-141.
    4. "NIOSH Manual of Analytical Methods", 2nd ed.; Department of Health, Education and Welfare, National Institute for Occupational Safety and Health: Cincinnati, OH, 1977; Vol. 2, Method No. S41; DHEW(NIOSH) Publ.(U.S.), No. 77-157-B.
    5. "American Conference of Governmental Industrial Hygienists: Documentation of the Threshold Limit Values, Supplemental Documentation for 1982", pp. 259-260, Cincinnati, OH (1982).
    6. "American Conference of Governmental Industrial Hygienists: Documentation of the Threshold Limit Values, Supplemental Documentation for 1982", p. 260, Cincinnati, OH (1982).
    7. "American Conference of Governmental Industrial Hygienists: Documentation of the Threshold Limit Values, Supplemental Documentation for 1982", p. 171, Cincinnati, OH (1982).
    8. "American Conference of Governmental Industrial Hygienists: Documentation of the Threshold Limit Values, Supplemental Documentation for 1982", p. 172, Cincinnati, OH (1982).
    9. "Documentation of the NIOSH Validation Tests"; Department of Health, Education and Welfare, National Institute for Occupational Safety and Health: Cincinnati, OH, 1977; Method No. S39; DHEW(NIOSH) Publ.(U.S.), No. 77-185.
    10. "Documentation of the NIOSH Validation Tests"; Department of Health, Education and Welfare, National Institute for Occupational Safety and Health: Cincinnati, OH, 1977; Method No. S41; DHEW(NIOSH) Publ.(U.S.), No. 77-185.
    11. "Documentation of the NIOSH Validation Tests"; Department of Health, Education and Welfare, National Institute for Occupational Safety and Health: Cincinnati, OH, 1977; Method No. S79; DHEW(NIOSH) Publ.(U.S.), No. 77-185.
    12. Nagano, K.; Nakayama, E.; Koyano, M.; Oobayaski, H.; Adachi, H.; Yamada, T. Jap. J. Ind. Health 1979, 21, 29-35.
    13. "Current Intelligence Bulletin 39, Glycol Ethers"; May 2, 1983, U.S. Department of Health and Human Services, Public Health Service, Center for Disease Control, NIOSH.
    14. "Pocket Guide to Chemical Hazards", NIOSH/OSHA, Sept. 1978, DHEW (NIOSH) Publ. No. 78-210.
Back to Top

Thank You for Visiting Our Website

You are exiting the Department of Labor's Web server.

The Department of Labor does not endorse, takes no responsibility for, and exercises no control over the linked organization or its views, or contents, nor does it vouch for the accuracy or accessibility of the information contained on the destination server. The Department of Labor also cannot authorize the use of copyrighted materials contained in linked Web sites. Users must request such authorization from the sponsor of the linked Web site. Thank you for visiting our site. Please click the button below to continue.