Powered by GoogleTranslate

Napthalene

OSHA Method ORG-35 | April 1982

For problems with accessibility in using figures and illustrations, please contact the Salt Lake Technical Center at 801-233-4900.
These procedures were designed and tested for internal use by OSHA personnel. Mention of any company name or commercial product does not constitute endorsement by OSHA.

The tables on this page are best viewed on a desktop, notebook, or tablet computer.

Method no.: 35
Matrix: Air

Target concentration: 10 ppm (OSHA PEL)

Procedure: Samples are collected on Chromosorb 106 tubes, desorbed with carbon disulfide and analyzed by gas chromatography with a flame ionization detector (FID).

Recommended air volume
and sampling rate:

10 L at 0.2 L/min

Reliable quantitation limit: 0.08 ppm (0.4 mg/m3)

Standard error of estimate
at the target concentration:
(Figure 4.6.2.)
6.4%

Status of method: Evaluated method. This method has been subjected to the established evaluation procedures of the Organic Methods Evaluation Branch.

Date: April 1982

Chemist: Mary Eide

Organic Methods Evaluation Branch
OSHA Salt Lake Technical Center
Salt Lake City, UT 84165-0200

  1. General Discussion
    1. Background
      1. History

        NIOSH method S292 recommends collection of naphthalene on charcoal tubes and desorption with carbon disulfide (Ref. 5.1.). Naphthalene desorption efficiency in this method (S292) is dependent upon the mass of naphthalene on the charcoal tube. NIOSH requires a 75% or better desorption efficiency for its methods. A loading of 10 mg of naphthalene on the charcoal tube gives a 76.2% desorption efficiency. To obtain a loading of 10 mg of naphthalene from an atmosphere at the PEL of 10 ppm, 200 L of air must be sampled. This is the air volume recommended by NIOSH. With lower loading the desorption efficiency decreases, for example at 0.5 mg, equivalent to the PEL of 10 ppm with a 10-L air volume, the desorption efficiency is 61%.

        Other solvents were considered for desorbing charcoal tubes, but no solvent better than carbon disulfide was found. Other collection tubes tried were XAD-4, Coated Resin (SKC, cat. no. 226-23), and Chromosorb 106. Chromosorb 106 had a 100% desorption efficiency over the range of 0.004 mg to 1.02 mg loading, when desorbed with carbon disulfide, and samples were stable over a 17-day storage time.

      2. Toxic effects (This section is for information only and should not be taken as the basis for OSHA policy).

        Naphthalene is readily absorbed through the skin and can cause skin irritation, corneal ulcerations, and cataracts (Refs. 5.2. and 5.3.). Naphthalene inhalation and ingestion can cause headaches, nausea, extensive sweating, and confusion. Symptoms of severe poisoning are vomiting, then coma, and hematuria (Ref. 5.4.). Sensitive individuals can have a "hemolytic crisis" (rupture of red blood cells) when exposed to naphthalene (Ref. 5.5.). Naphthalene has been shown to be a teratogen in rats (Ref. 5.6.).

      3. Potential workplace exposure

        The major uses of naphthalene in 1978 were: phthalic anhydride production (141,000 metric tons), insecticide production (35,000 metric tons), 2-naphthol production (16,000 metric tons), synthetic tanning agents (12,000 metric tons), moth repellents (4,000 metric tons), surfactant production (2,000 metric tons), and miscellaneous (2,000 metric tons) (Ref. 5.7.). The total consumption in 1979 was 212,000 metric tons, while the production in 1979 was 226,000 metric tons. The major sources of naphthalene are from coal tars and petroleum refineries (Ref. 5.7.).

      4. Physical properties (Ref. 5.8.).
        molecular weight: 128.2
        boiling point: 218°C
        melting point: 80.2°C
        color: white crystals
        vapor pressure: 0.05 mm Hg at 20°C
        flash point: 79°C (174°F) (closed cup)
        odor: moth balls
        specific gravity: 1.14
        lower explosive limit: 0.9%
        odor threshold: 0.3 ppm
        structural formula: Figure 1.1.4.
    2. Limit defining parameters
      1. Detection limit of the analytical procedure

        The detection limit of the analytical procedure is 4 ng per injection. This is the amount of analyte that gives a peak area similar to the peak area of trace contaminants in the desorbing solution. (Section 4.1.)

      2. Detection limit of the overall procedure

        The detection limit of the overall procedure is 4 µg per sample (0.08 ppm or 0.4 mg/m3, assuming the recommended air volume). This is the amount of analyte spiked on the sampling tube which allows recovery of an amount of analyte equivalent to the detection limit of the analytical procedure. (Section 4.2.)

      3. Reliable quantitation limit

        The reliable quantitation limit is 4 µg per sample (0.08 ppm or 0.4 mg/m3, assuming the recommended air volume). This is the smallest amount of analyte which can be quantitated within the criteria of 75% recovery and 95% confidence limits of ±25%. (Section 4.3.)

      4. Sensitivity

        The sensitivity of the analytical procedure over a concentration range representing 0.5 to 2 times the target concentration based on the recommended air volume is 684 area units per µg/mL. This is determined by the slope of the calibration curve. (Section 4.4.) The sensitivity may vary with the particular instrument used in the analysis. The instrument used in this evaluation was a Hewlett-Packard 5840A.

      5. Recovery

        The recovery of analyte from the collection medium during storage must be 75% or greater. The recovery of naphthalene from samples used in a 17-day storage test remained above 94% when the samples were stored at both 23°C and -5°C. (Section 4.6.)

      6. Precision (analytical method)

        The pooled coefficient of variation obtained from replicate determinations of analytical standards at 0.5, 1, and 2 times the target concentration is 0.0058. (Section 4.4.)

      7. Precision (overall procedure)

        The precision at the 95% confidence level for the 17-day storage test is ±12.5%. (Figure 4.6.2.) This includes an additional ±5% for sampling error.

    3. Advantages
      1. The sampling procedure is convenient.
      2. The analytical method is reproducible and sensitive.
      3. Re-analysis of samples is possible.
      4. Samples are stable, even at room temperature.
      5. It may be possible to analyze other compounds at the same time.
      6. Interferences may be avoided by proper selection of column and GC parameters.
      7. The desorption efficiency remained high over a range of 0.004 to 1.02 mg per tube.
    4. Disadvantages
      1. The precision, in part, is limited by the reproducibility of the pressure drop across the tubes. This applies, in general, to all sampling procedures using sampling tubes.
      2. There are trace amounts of naphthalene in the Chromosorb 106 packing. This was confirmed by GC/MS. A blank from the same lot as the tubes used in sampling must be sent to the laboratory and analyzed with the samples.
      3. The carbon disulfide contained trace contaminants that may interfere with the naphthalene analysis. Instrument parameters must be chosen to avoid this.
  2. Sampling Procedure
    1. Apparatus
      1. Personal sampling pump whose flow can be determined within ±5% at the recommended flow.
      2. Chromosorb 106 tubes containing a 100-mg adsorbing section with a 50-mg backup section separated by a 2-mm portion of urethane foam, with silanized glass wool before the adsorbing section and a 3-mm plug of urethane foam at the back of the backup section. The ends are flame sealed and the glass tube containing the adsorbent is 70 mm × 6-mm o.d. × 4-mm i.d. SKC tubes or equivalent.
    2. Reagents

      None required

    3. Sampling technique
      1. The ends of the Chromosorb 106 tubes are broken immediately before sampling.
      2. Connect the Chromosorb 106 tube to the sampling pump with flexible tubing.
      3. Tubes should be placed in a vertical position to minimize channeling, with the smaller section towards the pump.
      4. Air being sampled should not pass through any hose or tubing before entering the Chromosorb 106 tubes.
      5. Seal the Chromosorb 106 tubes with plastic caps immediately after sampling. Seal each sample lengthwise with OSHA Form 21.
      6. With each batch of samples, submit at least one blank tube from the same lot used for samples. This tube should be subjected to exactly the same handling as the samples (break ends, seal, transport) except no air is drawn through it.
      7. Ship the samples (and corresponding paperwork) to the lab for analysis.
      8. Bulks submitted for analysis must be shipped in a separate container from the samples.
    4. Sampler capacity
      1. Breakthrough was determined two ways. First, a standard naphthalene concentration was prepared in a gas bag filled with humid air (80% relative humidity). When samples were drawn from the gas bag atmosphere, no breakthrough was observed with a loading of 3.72 mg on the sampling tube. The total sampling time was 283 min at a flow rate of 0.192 L/min. This concentration is equivalent to an air concentration of 13 ppm. (Section 4.7.1.).
      2. Breakthrough was also studied with a tube packed with silanized glass wool in series with the sampling and backup tubes. A solution of naphthalene was injected onto the glass wool and 33 L of humid air were drawn through. No breakthrough was observed with up to 998 µg of naphthalene on the sampling tube. No breakthrough was observed at three times the recommended air volume in either breakthrough study (Section 4.7.2.).
    5. Desorption efficiency

      The desorption efficiency was essentially 100% for 0.5 to 2 times the PEL for a 10-L air sample (Section 4.5. and 4.2.).

    6. Recommended air volume and sampling rate
      1. The recommended air volume is 10 L.
      2. The recommended sampling rate is 0.2 L/min.
    7. Interferences

      Suspected interferences should be listed on sample data sheets.

    8. Safety precautions
      1. Sampling equipment should be placed on an employee in a manner that does not interfere with work performance or safety.
      2. Safety glasses should be worn when breaking the ends of the adsorbent tubes.
      3. Follow all safety practices that apply to the workplace being sampled.
  3. Analytical Procedure
    1. Apparatus
      1. Gas chromatograph equipped with a flame ionization detector.
      2. GC column capable of separating naphthalene and an internal standard from any interferences. The column used for validation studies was 10 ft × 1/8 in., stainless steel, packed with 3% SP-2100 on 100/120 Supelcoport.
      3. An electronic integrator or some other suitable method of measuring peak areas.
      4. Two-milliliter vials with Teflon-lined caps.
      5. A 1-µL syringe or other convenient size for sample injection.
      6. Pipets for dispensing desorbing reagent. The Glenco 1-mL dispenser was used in this method.
      7. Volumetric flasks, 25-mL and other convenient sizes for preparing standards.
      8. A balance capable of weighing microgram quantities.
      9. Pipets of convenient sizes for standard preparation.
    2. Reagents
      1. Naphthalene, reagent grade.
      2. Carbon disulfide, chromatographic grade.
      3. An internal standard, such as n-hexylbenzene, reagent grade.
      4. Desorbing reagent - 1 µL internal standard/mL CS2.
      5. Purified GC grade nitrogen or helium, hydrogen, and air.
    3. Standard preparation
      1. Standards are prepared by diluting a known quantity of naphthalene in desorbing reagent. Fresh standards should be prepared daily.
      2. A concentration of 524 µg of naphthalene per mL desorbing reagent is equivalent to 10 ppm for a 10-L air sample desorbed with 1 mL of desorbing reagent.
      3. At least two separate stock standards should be prepared.
    4. Sample preparation
      1. Sample tubes are opened and the adsorbent from each section of each tube is placed in separate 2-mL vials. Foam and glass wool plugs are discarded.
      2. Each section is desorbed with 1 mL of desorbing reagent.
      3. The vials are sealed immediately and allowed to desorb for 30 min with occasional shaking.
    5. Analysis
      1. GC conditions
        flow rates (mL/min) temperature (°C)
          nitrogen:   28   injector: 200
          hydrogen:   22   detector: 250
          air: 240   column: 120

        injection size: 1 µL
        elution time: 4.62 min
        chromatogram: Figure 4.8.
      2. Peak areas are measured by an integrator or other means.
      3. An internal standard procedure is used. The integrator is calibrated to report results in parts per million based on a 10-L air sample and a desorption efficiency of 100% (Section 3.7.).
    6. Interferences
      1. Any compound having a retention time similar to that of naphthalene or the internal standard is an interference. Possible interferences should be listed on the sample data sheet. GC parameters should be adjusted if necessary so these interferences will pose no problems.
      2. Retention time data on a single column is not considered proof of chemical identity. Samples over the PEL should be confirmed by GC/MS or other suitable means.
      3. There are trace amounts of naphthalene present in the Chromosorb 106 tubes. A blank of the same lot as the samples must be analyzed at the same time, and a blank correction must be made.
      4. Phthalic anhydride, 1-naphthol, and 2-naphthol do not interfere with naphthalene when the recommended column and conditions of this method are used.
    7. Calculations
      1. To calculate the equivalent ppm air concentration of analytical standards, assuming a 10-L air sample, use the following equation:

        ppm = (1 mL)(A)(24.46)/(128.2)(10 L)

        where            A = µg/mL of naphthalene in standard
        128.2 = molecular weight of naphthalene
        1 mL = desorption volume
        24.46 = molar volume (L/mole) at 25°C and 760 mm Hg

      2. Since the integrator was calibrated to report results in ppm based on a 10-L air sample, the following calculation is used:

        ppm naphthalene = (A-C)/(B)(0.1)

        where            A = ppm calculated by integrator
        B = air volume (L)
        C = blank amount (ppm)

      3. This calculation is done for each section of the sampling tube and the results added together.
    8. Safety precautions
      1. All handling of solvents should be done in a hood.
      2. Avoid skin contact with all solvents.
      3. Wear safety glasses at all times.
  4. Backup Data
    1. Detection limit data

      The detection limit of the analytical procedure was determined by injecting 1 µL of a 4.0 µg/mL standard prepared in carbon disulfide containing 1 µL/mL n-hexylbenzene internal standard. The detection limit of the analytical procedure is 4 ng per injection. Chromatograms of such a determination are shown in Figure 4.1.

    2. Detection limit of the overall procedure

      The detection limit of the overall procedure was determined from the following desorption data. This data is presented graphically in Figures 4.2.1. and 4.2.2.

      Table 4.2.1.
      Detection Limit Data

      µg/sample 1020.8 500.2 268.6 131.6 26.3 4.4
      µg recovered



      1041.5
      1037.8
      1044.4
      1040.4
      1038.5
      1032.3
      513.9
      514.6
      507.9
      513.1
      514.5
      511.8
      277.8
      273.8
      270.6
      267.7
      270.7
      269.3
      131.8
      131.1
      131.6
      133.2
      130.7
      133.3
      28.0
      26.6
      25.9
      23.8

      4.6
      4.7
      4.5
      4.3

    3. Reliable quantitation limit

      The recovery of naphthalene from three samples spiked with an amount equivalent to the detection limit of the overall procedure, was 103.6%, 102.03% and 99.25%. The average of these values is 101.7% and the standard deviation is 2.20%. Therefore, the 95% confidence limits (±1.96 SD) is ±4.32%.

    4. Sensitivity and precision data (analytical method)

      Four analytical standards whose concentrations are indicated below were each analyzed six times. The data listed in Table 4.4 were used to determine the calibration curve shown in Figure 4.4. The precision of the analytical method was determined with the 0.5×, 0.9×, and 2× data.

      Table 4.4.
      Sensitivity and Precision

      × target conc.
      µg/mL
        0.5×
      286.8
        0.9×
      451.4
        2×
      1063
        3.28×
      1718.4
      area counts






      X
      SD
      CV

      CV = 0.0058
      209900
      210950
      211200
      209950
      210400
      210300

      210450
      1179.0
      0.00560
      306900
      304200
      309050
      305550
      308300
      307850

      306975
      1820.1
      0.00593
      727800
      740800
      734100
      737900
      734200
      735200

      735000
      4367.6
      0.00594
      1188800
      1185000
      1180500
      1176500
      1181500
      1184000

    5. Desorption efficiency

      Desorption efficiencies were determined by injecting a solution of naphthalene in o-xylene onto Chromosorb 106 tubes and analyzing them a day later. The desorption efficiencies were determined with sample loadings equivalent to 0.5, 1, and 2 times the target concentration based on the recommended air volume of 10 L. Six samples were prepared and analyzed at each concentration.

      Table 4.5.
      Desorption Efficiency

      × target conc. 0.5×  1×   2×  
      desorption
      efficiency,
      %




      X

      X = 101.8
      100.3
      103.4
      101.9
      100.7
      99.6
      100.8

      101.2
      102.7
      102.9
      101.5
      102.6
      102.9
      102.3

      102.5
      102.0
      101.7
      102.3
      101.9
      101.7
      101.1

      101.8
    6. Storage

      Storage samples were prepared by vapor spiking Chromosorb 106 tubes with naphthalene. Tubes containing silanized glass wool were liquid spiked with 5.0 µL of 100.12 mg/mL naphthalene in o-xylene. These tubes were placed in series ahead of Chromosorb 106 tubes and humid air was drawn through them. The naphthalene was leached from the glass wool and adsorbed on the Chromosorb 106. Six samples were analyzed the day after generation. They had been stored in a refrigerator at -5°C. Fifteen samples were stored at -5°C and 15 samples were stored at room temperature (23°C). Three samples from each set were analyzed at intervals of a few days over a 17-day period. Results are shown graphically in Figures 4.6.1. and 4.6.2.

      Table 4.6.
      Storage Tests

      storage time
      (days)
                            % recovery
                            (refrigerated)
                       % recovery
                        (ambient)
        1
        1
        6
        9
      13
      15
      17
      91.8
      98.8
      99.2
      96.0
      93.9
      96.0
      100.0
      98.3
      100.3
      100.2
      97.0
      96.9
      101.6
      98.8
      100.2
      99.4
      99.1
      96.2
      96.5
      100.9
      97.7


      99.6
      97.2
      94.7
      96.6
      97.9


      99.5
      97.0
      85.3
      89.7
      96.5


      94.7
      90.7
      94.7
      97.6
      96.4
    7. Sampler capacity
      1. Breakthrough study using a Teflon gas bag

        A Teflon gas bag was filled with 60 L of humid air at 80% relative humidity. A solution containing 24 mg of naphthalene in xylene was injected into the gas bag. The gas bag was allowed to equilibrate for 25 h. Naphthalene crystals were observed in the gas bag after this period of time. Samples were taken from the gas bag using a Chromosorb 106 tube containing only the 100-mg section of adsorbent connected in series with a normal Chromosorb 106 tube. The flow was from the gas bag, through the tube with the main section only, then through the second Chromosorb 106 tube. This second Chromosorb 106 tube was changed at timed intervals and analyzed for the presence of naphthalene. The air was drawn at 0.192 L/min.

        Table 4.7.1.
        Breakthrough Study Using a Gas Bag

        tube no.


        total time
        (min)

        total air drawn
        (L)

        µg found

        1
        2
        3
        4
        main tube

        110
        210
        260
        283
        283

        21.08
        40.24
        49.82
        54.23
        54.23

              0
              0
              0
              0
        3720

      2. A second breakthrough study was performed using a similar sampling train to that used for storage sample generation. A tube packed with silanized glass wool was placed in series with a Chromosorb 106 tube with the backup portion removed. These tubes were followed by another Chromosorb 106 tube which was changed at timed intervals. The silanized glass wool packed tube was injected with 20 µL of a 49.9 mg/mL naphthalene in o-xylene solution. The total loading was 998 of µg naphthalene. The air drawn was at 23.5°C and relative humidity of 72.34%. The flow rate was 0.193 L/min.

        Table 4.7.2.
        Breakthrough Study Using Spiked Glass Wool

        tube no.
        total time
        (min)
        total air drawn
        (L)
        µg found
            1
            2
            3
            4
            5
        glass wool tube
        main tube
          42
          62
          96
        126
        171
        171
        171
          8.11
        11.97
        18.53
        24.32
        33.00
        33.00
        33.00
            0
            0
            0
            0
            0
            0
        976
        Structural formula of naphthalene

        Figure 1.1.4. Structural formula of naphthalene.

        Detection limit chromatograms

        Figure 4.1. Detection limit chromatograms.

        Desorption data

        Figure 4.2.1. Desorption data.

        Desorption data

        Figure 4.2.2. Desorption data.

        Calibration curve

        Figure 4.4. Calibration curve.

        Refrigerated storage

        Figure 4.6.1. Refrigerated storage.

        Ambient storage

        Figure 4.6.2. Ambient storage.

        Chromatogram of a 369 µg/mL naphthalene standard in CS2 with 1 µl/mL n-hexylbenzene as an internal standard

        Figure 4.8. Chromatogram of a 369 µg/mL naphthalene standard in CS2 with 1 µl/mL n-hexylbenzene as an internal standard.

  5. References
    1. "NIOSH Manual of Analytical Methods", U.S. Department of Health, Education and Welfare, Public Health Service, Center for Disease Control, National Institute for Occupational Safety and Health, Second Edition, Vol. 3, Method S292.
    2. Meyer, R.T. N. Engl. J. Med. 1955, 252, 622-628.
    3. Adams, D.R. BR. J. Ophthanol. 1930, 14, 545-576.
    4. Patty, F.A. "Industrial Hygiene and Toxicology", John Wiley and Sons, New York, Second Edition, Vol. 2, p. 1238.
    5. Athreya, B.K.; Swain, A.K.; and Dickstein, B. Indian J. Child. Health 1961, 10, 305-308.
    6. Harris, S.J.; Bond, G.P.; and Riemeir, R.W. Toxicology Appl. Pharmacol. 1979, 48-1 part 2, A35.
    7. "Encyclopedia of Chemical Technology." John Wiley and Sons, New York, Vol. 15, p. 710, 714.
    8. "Occupational Health Guidelines for Chemical Hazards", U.S. Department of Health and Human Services, Public Health Service, Center for Disease Control, National Institute for Occupational Safety and Health, U.S. Department of Labor, Occupational Safety and Health Administration, Vol. 2, Naphthalene 1-5.
Back to Top

Thank You for Visiting Our Website

You are exiting the Department of Labor's Web server.

The Department of Labor does not endorse, takes no responsibility for, and exercises no control over the linked organization or its views, or contents, nor does it vouch for the accuracy or accessibility of the information contained on the destination server. The Department of Labor also cannot authorize the use of copyrighted materials contained in linked Web sites. Users must request such authorization from the sponsor of the linked Web site. Thank you for visiting our site. Please click the button below to continue.

Close