Powered by GoogleTranslate

Occupational Safety and Health Administration OSHA

Inspection Detail

Inspection: 1340036.015 - Microvention

Inspection Information - Office: Ca Santa Ana District Office

Nr: 1340036.015Report ID: 0950631Open Date: 08/20/2018

Microvention
35 Enterprise, Floor 2null
Aliso Viejo, CA 92656
Union Status: NonUnion
SIC:
NAICS: 339112/Surgical and Medical Instrument Manufacturing
Mailing: 35 Enterprise, Floor 2null, Aliso Viejo, CA 92656

Inspection Type:Accident
Scope:Partial Advanced Notice:N
Ownership:Private
Safety/Health:Safety Close Conference:01/24/2019
Close Case:10/01/2019

Related Activity:TypeIDSafetyHealth
 Accident1371635    

Violation Summary
Serious Willful Repeat Other Unclass Total
Initial Violations 3 3
Current Violations 3 3
Initial Penalty $72,900 $0 $0 $0 $0 $72,900
Current Penalty $24,300 $0 $0 $0 $0 $24,300
FTA Amount $0 $0 $0 $0 $0 $0

Violation Items
# ID Type Standard Issuance Abate Curr$ Init$ Fta$ Contest LastEvent
  1. 01001 Serious 3314(D) 01/24/2019 $20,250 $24,300 $0 02/08/2019 O - Administrative Law Judge Order
  2. 02001 Serious 3314(G) 01/24/2019 $2,025 $24,300 $0 02/08/2019 O - Administrative Law Judge Order
  3. 03001 Serious 3314(L)(1) 01/24/2019 $2,025 $24,300 $0 02/08/2019 O - Administrative Law Judge Order

Accident Investigation Summary
Summary Nr: 112777.015Event: 08/14/2018Employee Crushes And Burns Finger While Testing Temperature
On August 14, 2018 at 17:52 p.m., Employee # 1 was testing the temperature of a heat sealing machine (SencorpWhite 24 thermal sealer Serial Number 18-10915), in cluding a data logging device to log temperatures of the heat bar. The machine w as in the set up phase of new equipment in the maintenance area, outside of the cleaning room. He was placing thermocouples on a heat sealing bar and the equipm ent was energized. He removed the guards and placed his finger in between two he at sealing bars of the sealing device. He then activated the foot pedal crushing and burning his finger. He was hospitalized for more than 24 hours. The employ er called the case in after hours on August 20, 2018 at 9:00 a.m. The employer w as compliant with reporting regulations. The accident happened due to a lack of training and lockout tagout program and controls and serious violations of stand ards.
Keywords: burn, crushed, finger, foot pedal, heat, lockout/tagout, seal, sealing bar, sealing machine, unguarded
Inspection Degree Nature Occupation
1 1340036.015 Hospitalized injury Engineering technicians, n.e.c.

Thank You for Visiting Our Website

You are exiting the Department of Labor's Web server.

The Department of Labor does not endorse, takes no responsibility for, and exercises no control over the linked organization or its views, or contents, nor does it vouch for the accuracy or accessibility of the information contained on the destination server. The Department of Labor also cannot authorize the use of copyrighted materials contained in linked Web sites. Users must request such authorization from the sponsor of the linked Web site. Thank you for visiting our site. Please click the button below to continue.

Close