Inspection Detail
Inspection: 305355349 - Medtronic Inc
Inspection Information - Office: Santa Ana District Office
Site Address:
Medtronic Inc
1851 E Deere Ave
Santa Ana, CA 92705
Mailing Address:
, , 00000
Union Status: NonUnion
SIC:3999
NAICS: 339999/All Other Miscellaneous Manufacturing
Inspection Type: Accident
Scope: Partial
Advanced Notice: N
Ownership: Private
Safety/Health: Safety
Close Conference: 04/22/2005
Planning Guide: Safety-Manufacturing
Emphasis:
Case Closed: 09/22/2006
Type | Activity Nr | Safety | Health |
---|---|---|---|
Accident | 362470312 |
Violations/Penalties | Serious | Willful | Repeat | Other | Unclass | Total |
---|---|---|---|---|---|---|
Initial Violations | 1 | 2 | 3 | |||
Current Violations | 1 | 2 | 3 | |||
Initial Penalty | $6,750 | $0 | $0 | $1,125 | $0 | $7,875 |
Current Penalty | $6,750 | $0 | $0 | $1,125 | $0 | $7,875 |
FTA Penalty | $0 | $0 | $0 | $0 | $0 | $0 |
# | Citation ID | Citaton Type | Standard Cited | Issuance Date | Abatement Due Date | Current Penalty | Initial Penalty | FTA Penalty | Contest | Latest Event | Note |
---|---|---|---|---|---|---|---|---|---|---|---|
1. | 01001 | Other | 3314 B | 04/22/2005 | 05/15/2005 | $750 | $750 | $0 | - | ||
2. | 01002 | Other | 3314 H | 04/22/2005 | 05/15/2005 | $375 | $375 | $0 | - | ||
3. | 02001 | Serious | 4184 B | 04/22/2005 | 05/15/2005 | $6,750 | $6,750 | $0 | - |
Investigation Summary
On December 2, 2004, Employee #1 was operating a Leica Microsystems, microtome machine, Model Number 2155, Serial Number 050231619. Her employer manufactures human heart valves and other related heart repair products. The microtome machine slices tissue samples that are embedded in a paraffin block, similar to a punch press, which then allows the samples to be viewed under a microscope. An adjustable knife is located on a stationary fixture below a specimen-holder clamp and moves up and down like a ram on a punch press. The stroke length of the knife is approximately 2.75 inches. The clamp holds the paraffin block and moves horizontally outwards in very small increments for each new slice, while cycling up and down. During each cycle the chuck scrapes the paraffin block against the fixed knife mounted horizontally at the base of the machine. As Employee #1 was setting up a paraffin block in the specimen-holder clamp, she inadvertently activated the machine through a foot pedal. Her left hand came into contact with the knife and the tips of her thumb and index finger on that hand were amputated. Employee #1 was hospitalized and treated for her injury, which included bone loss. The employer later modified the machine to operate through a hand control.
Keywords: KNIFE, AMPUTATED, SLICER, FINGER, FOOT CONTROL, MACHINE OPERATOR, CAUGHT BETWEEN, FOOT-POWERED PRESS, PUNCH PRESS, PUNCTURE
# | Inspection | Age | Sex | Degree of Injury | Nature of Injury | Occupation |
---|---|---|---|---|---|---|
1 | 305355349 | Hospitalized injury | Amputation | Science technicians, n.e.c. |