Powered by GoogleTranslate

Inspection Detail

Case Status: CLOSED

Inspection: 1452851.015 - Ge Precision Healthcare Llc

Inspection Information - Office: Monrovia District Office

 

Inspection Nr: 1452851.015
Report ID: 0950644
Date Opened: 12/18/2019

Site Address:
Ge Precision Healthcare Llc
1500 Duarte Rd.
Duarte, CA 91010

Mailing Address:
9900 West Innovation Dr., Wauwatosa, WI 53226

Union Status: NonUnion

SIC:

NAICS: 334517/Irradiation Apparatus Manufacturing


Inspection Type: Accident

Scope: Partial

Advanced Notice: N

Ownership: Private

Safety/Health: Safety

Close Conference: 05/20/2020

Emphasis:

Case Closed: 11/08/2022


Related Activity
Type Activity Nr Safety Health
Accident 1522706
Case Status: CLOSED
Violation Summary
Violations/Penalties Serious Willful Repeat Other Unclass Total
Initial Violations 1 1
Current Violations 1 1
Initial Penalty $0 $0 $0 $935 $0 $935
Current Penalty $0 $0 $0 $750 $0 $750
FTA Penalty $0 $0 $0 $0 $0 $0

Violation Items
# Citation ID Citaton Type Standard Cited Issuance Date Abatement Due Date Current Penalty Initial Penalty FTA Penalty Contest Latest Event Note
1. 01001 Other 3314(D) 05/20/2020 06/25/2020 $750 $935 $0 06/10/2020 O - Administrative Law Judge Order  

Investigation Summary

Investigation Nr: 125283.015
Event: 11/26/2019
Employee amputates finger when pulled into drive chain

At 2:30 p.m. on November 26, 2019, an employee who worked as an industrial engineer for a medical radiological device manufacturer was troubleshooting a Revolution Model XKD X-ray Table. A malfunction was occurring in the vertical function of the table. The issue appeared to be a motor that was preventing the table from going up and down. The employee removed the cover to the motor compartment and attempted to engage the drive chain. The employee was operating the foot pedal and reached into the compartment with his gloved right hand to touch the chain to see if there was tension. As he reached inside the compartment, the motor engaged causing the chain to move. The chain caught his pinky finger at the base of the finger. The employee sustained an open fracture and amputation to his finger, requiring hospitalization.

Keywords: Amputated, Amputation, Caught In, Chain, Finger, Foot Control, Foot pedal, Fracture, Glove, Instantaneous amputation, Lockout, Lockout/Tagout, Machine operator, Malfunction, Mech Malfunction, Misjudgment, Misjudgment of Hazardous Situation, Open Fracture, Partial Amputation, Pinch Point, Pinched, Pulled In, Reaching, Traumatic Amputation, Troubleshooting

Investigated Inspection
# Inspection Age Sex Degree of Injury Nature of Injury Occupation
1 1452851.015 57 M Hospitalized injury Industrial engineers
Back to Top

Thank You for Visiting Our Website

You are exiting the Department of Labor's Web server.

The Department of Labor does not endorse, takes no responsibility for, and exercises no control over the linked organization or its views, or contents, nor does it vouch for the accuracy or accessibility of the information contained on the destination server. The Department of Labor also cannot authorize the use of copyrighted materials contained in linked Web sites. Users must request such authorization from the sponsor of the linked Web site. Thank you for visiting our site. Please click the button below to continue.