Inspection Detail
Inspection: 1452851.015 - Ge Precision Healthcare Llc
Inspection Information - Office: Monrovia District Office
Site Address:
Ge Precision Healthcare Llc
1500 Duarte Rd.
Duarte, CA 91010
Mailing Address:
9900 West Innovation Dr., Wauwatosa, WI 53226
Union Status: NonUnion
SIC:
NAICS: 334517/Irradiation Apparatus Manufacturing
Inspection Type: Accident
Scope: Partial
Advanced Notice: N
Ownership: Private
Safety/Health: Safety
Close Conference: 05/20/2020
Emphasis:
Case Closed: 11/08/2022
| Type | Activity Nr | Safety | Health |
|---|---|---|---|
| Accident | 1522706 |
| Violations/Penalties | Serious | Willful | Repeat | Other | Unclass | Total |
|---|---|---|---|---|---|---|
| Initial Violations | 1 | 1 | ||||
| Current Violations | 1 | 1 | ||||
| Initial Penalty | $0 | $0 | $0 | $935 | $0 | $935 |
| Current Penalty | $0 | $0 | $0 | $750 | $0 | $750 |
| FTA Penalty | $0 | $0 | $0 | $0 | $0 | $0 |
| # | Citation ID | Citaton Type | Standard Cited | Issuance Date | Abatement Due Date | Current Penalty | Initial Penalty | FTA Penalty | Contest | Latest Event | Note |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 1. | 01001 | Other | 3314(D) | 05/20/2020 | 06/25/2020 | $750 | $935 | $0 | 06/10/2020 | O - Administrative Law Judge Order |
Investigation Summary
At 2:30 p.m. on November 26, 2019, an employee who worked as an industrial engineer for a medical radiological device manufacturer was troubleshooting a Revolution Model XKD X-ray Table. A malfunction was occurring in the vertical function of the table. The issue appeared to be a motor that was preventing the table from going up and down. The employee removed the cover to the motor compartment and attempted to engage the drive chain. The employee was operating the foot pedal and reached into the compartment with his gloved right hand to touch the chain to see if there was tension. As he reached inside the compartment, the motor engaged causing the chain to move. The chain caught his pinky finger at the base of the finger. The employee sustained an open fracture and amputation to his finger, requiring hospitalization.
Keywords: Amputated, Amputation, Caught In, Chain, Finger, Foot Control, Foot pedal, Fracture, Glove, Instantaneous amputation, Lockout, Lockout/Tagout, Machine operator, Malfunction, Mech Malfunction, Misjudgment, Misjudgment of Hazardous Situation, Open Fracture, Partial Amputation, Pinch Point, Pinched, Pulled In, Reaching, Traumatic Amputation, Troubleshooting
| # | Inspection | Age | Sex | Degree of Injury | Nature of Injury | Occupation |
|---|---|---|---|---|---|---|
| 1 | 1452851.015 | 57 | M | Hospitalized injury | Industrial engineers |
Translate