Powered by GoogleTranslate

Inspection Detail

Inspection: 125872085 - Siemens Medical Solutions Usa

Inspection Information - Office: San Diego District Office

 

Inspection Nr: 125872085
Report ID: 0950632
Date Opened: 06/24/2004

Site Address:
Siemens Medical Solutions Usa
477 No El Camino Real
Encinitas, CA 92024

Mailing Address:
110 Macalyson Court, Cary, NC 27511

Union Status: NonUnion

SIC:5074

NAICS: 423720/Plumbing and Heating Equipment and Supplies (Hydronics) Merchant Wholesalers


Inspection Type: Accident

Scope: Partial

Advanced Notice: N

Ownership: Private

Safety/Health: Safety

Close Conference: 11/30/2004

Emphasis:

Case Closed: 10/10/2006


Related Activity
Type Activity Nr Safety Health
Accident 362417909
Violation Summary
Violations/Penalties Serious Willful Repeat Other Unclass Total
Initial Violations 1 1
Current Violations 1 1
Initial Penalty $0 $0 $0 $750 $0 $750
Current Penalty $0 $0 $0 $750 $0 $750
FTA Penalty $0 $0 $0 $0 $0 $0

Violation Items
# Citation ID Citaton Type Standard Cited Issuance Date Abatement Due Date Current Penalty Initial Penalty FTA Penalty Contest Latest Event Note
1. 01001 Other 3203 A 11/30/2004 01/02/2005 $750 $750 $0 -  

Investigation Summary

Investigation Nr: 201067972
Event: 06/05/2004
Employee's Arm Is Injured by Machine

At 7:00 p.m. on June 5, 2004, Employee #1 was working for his employer, which provides its customers equipment upgrades, preventative maintenance, and trouble shooting of problems. Employee #1 was upgrading a Siemens MAGNETOM Symphony MRI (magnetic resonance imaging) machine by installing a shim filter in the cabinet above and off to the side of the machine. Employee #1 had taken the filter out of the cabinet and was carrying the 22-kilogram part. The magnet in the bore of the MRI machine drew the shim filter into and towards the bore of the machine, which was not deenergized. The shim filter, which Employee #1 was holding, turned him in a clockwise direction and smashed his right forearm, pinning it against the bore of the machine. Employee #1 sustained torn ligaments, tendons, and muscles. He also sustained fractures to his right pinky and ring fingers. Employee #1 transported himself to Scripps Encinitas Hospital. He was hospitalized for more than 24 hours for surgery and treatment. The employer reported the accident to the division on June 7, 2004.

Keywords: MAGNET, FRACTURE, TENDON, FINGER, LIGAMENT, PINNED, STRUCK AGAINST, LOCKOUT, ARM

Investigated Inspection
# Inspection Age Sex Degree of Injury Nature of Injury Occupation
1 125872085 Hospitalized injury Other Misc. electrical & electronic equipment repairers
Back to Top

Thank You for Visiting Our Website

You are exiting the Department of Labor's Web server.

The Department of Labor does not endorse, takes no responsibility for, and exercises no control over the linked organization or its views, or contents, nor does it vouch for the accuracy or accessibility of the information contained on the destination server. The Department of Labor also cannot authorize the use of copyrighted materials contained in linked Web sites. Users must request such authorization from the sponsor of the linked Web site. Thank you for visiting our site. Please click the button below to continue.

Close