Powered by GoogleTranslate

Inspection Detail

Inspection: 115995862 - Varian Oncology Systems

Inspection Information - Office: Kansas City Area Office

 

Inspection Nr: 115995862
Report ID: 0728500
Date Opened: 10/16/1995

Site Address:
Varian Oncology Systems
3030 Baltimore
Kansas City, MO 64111

Mailing Address:
3130 Hansen Way M/S F-056, Palo Alto, CA 94304

Union Status: NonUnion

SIC:3844

NAICS: 0 


Inspection Type: Accident

Scope: Partial

Advanced Notice: N

Ownership: Private

Safety/Health: Safety

Close Conference: 04/09/1996

Emphasis:

Case Closed: 05/30/1997


Related Activity
Type Activity Nr Safety Health
Accident 361886682
Violation Summary
Violations/Penalties Serious Willful Repeat Other Unclass Total
Initial Violations 2 2
Current Violations 1 1
Initial Penalty $18,900 $0 $0 $0 $0 $18,900
Current Penalty $0 $0 $0 $250 $0 $250
FTA Penalty $0 $0 $0 $0 $0 $0

Violation Items
# Citation ID Citaton Type Standard Cited Issuance Date Abatement Due Date Current Penalty Initial Penalty FTA Penalty Contest Latest Event Note
1. 01001A Serious 19100147 C01 04/09/1996 05/12/1996 $6,300 $6,300 $0 04/29/1996 F - Formal Settlement Citation has been deleted.
2. 01001B Serious 19100147 C02 I 04/09/1996 05/12/1996 $0 $0 $0 04/29/1996 F - Formal Settlement Citation has been deleted.
3. 01001C Serious 19100147 C04 I 04/09/1996 05/12/1996 $0 $0 $0 04/29/1996 F - Formal Settlement Citation has been deleted.
4. 01001D Serious 19100147 C04 II 04/09/1996 05/12/1996 $0 $0 $0 04/29/1996 F - Formal Settlement Citation has been deleted.
5. 01001E Serious 19100147 C05 I 04/09/1996 05/12/1996 $0 $0 $0 04/29/1996 F - Formal Settlement Citation has been deleted.
6. 01001F Serious 19100147 C06 I 04/09/1996 05/12/1996 $0 $0 $0 04/29/1996 F - Formal Settlement Citation has been deleted.
7. 01001G Serious 19100147 C07 I 04/09/1996 05/12/1996 $0 $0 $0 04/29/1996 F - Formal Settlement Citation has been deleted.
8. 01001H Other 19100147 C07 IV 04/09/1996 01/08/1997 $250 $6,300 $0 04/29/1996 F - Formal Settlement  
9. 01002A Serious 19100147 F02 I 04/09/1996 05/12/1996 $6,300 $6,300 $0 04/29/1996 F - Formal Settlement Citation has been deleted.
10. 01002B Serious 19100147 F04 04/09/1996 05/12/1996 $0 $0 $0 04/29/1996 F - Formal Settlement Citation has been deleted.

Investigation Summary

Investigation Nr: 170789135
Event: 10/14/1995
Employee killed when crushed by counterbalance

On October 14, 1995, Employee #1, of Varion Oncology Systems, and three employees of Unique Electronics were dismantling a linear accelerator. The counterbalance weight, weighing 4,000 lb, had been disconnected, leaving the head of the unit, weighing 4,000 lb, in the overhead position. In normal operation the head and the counterbalance rotate around a central axis. The four employees were moving the counterbalance weight when the head of the unit rotated downward, striking, crushing, and killing Employee #1.

Keywords: HEAD, COUNTERBALANCE, UNSECURED, WORK RULES, CONCUSSION, DISMANTLING, CONSTRUCTION, CRUSHED, STRUCK BY

Investigated Inspection
# Inspection Age Sex Degree of Injury Nature of Injury Occupation
1 115995862 Fatality Concussion Occupation not reported
Back to Top

Thank You for Visiting Our Website

You are exiting the Department of Labor's Web server.

The Department of Labor does not endorse, takes no responsibility for, and exercises no control over the linked organization or its views, or contents, nor does it vouch for the accuracy or accessibility of the information contained on the destination server. The Department of Labor also cannot authorize the use of copyrighted materials contained in linked Web sites. Users must request such authorization from the sponsor of the linked Web site. Thank you for visiting our site. Please click the button below to continue.