Powered by GoogleTranslate

Inspection Detail

Inspection: 112063516 - Usc School Of Pharmacy

Inspection Information - Office: Long Beach District Office

 

Inspection Nr: 112063516
Report ID: 0950642
Date Opened: 10/31/1990

Site Address:
Usc School Of Pharmacy
1985 Zonal Avenue
Los Angeles, CA 90033

Mailing Address:
, , 00000

Union Status: NonUnion

SIC:8733

NAICS: 0 


Inspection Type: Accident

Scope: Complete

Advanced Notice: N

Ownership: Private

Safety/Health: Safety

Close Conference: 12/19/1990

Emphasis:

Case Closed: 02/22/1991


Related Activity
Type Activity Nr Safety Health
Accident 361061716
Violation Summary
Violations/Penalties Serious Willful Repeat Other Unclass Total
Initial Violations 1 8 9
Current Violations 1 8 9
Initial Penalty $2,000 $0 $0 $2,500 $0 $4,500
Current Penalty $2,000 $0 $0 $2,500 $0 $4,500
FTA Penalty $0 $0 $0 $0 $0 $0

Violation Items
# Citation ID Citaton Type Standard Cited Issuance Date Abatement Due Date Current Penalty Initial Penalty FTA Penalty Contest Latest Event Note
1. 01002 Other 3220 E03 12/19/1990 12/19/1990 $0 $0 $0 -  
2. 01003 Other 3221 A 12/19/1990 01/20/1991 $0 $0 $0 -  
3. 01004 Other 3221 D02 12/19/1990 12/19/1990 $0 $0 $0 -  
4. 01005 Other 5162 C 12/19/1990 01/20/1991 $0 $0 $0 -  
5. 01006 Other 5162 D 12/19/1990 01/20/1991 $0 $0 $0 -  
6. 01007 Other 6151 C04 12/19/1990 12/19/1990 $0 $0 $0 -  
7. 01008 Other 6151 E02 12/19/1990 01/20/1991 $0 $0 $0 -  
8. 02001 Other 5162 E 12/19/1990 12/27/1990 $2,500 $2,500 $0 -  
9. 03001 Serious 5167 A 12/19/1990 12/27/1990 $2,000 $2,000 $0 -  

Investigation Summary

Investigation Nr: 170180061
Event: 10/31/1990
Two employees burned when ethanol ignites

Employees #1 and #2 were carrying out a chemical reaction on an open lab countertop. This was a hydrogenation reaction starting with material dissolved in ethanol, using hydrogen, and with palladium as a catalyst. The mixture ignited in the beaker, either from contact with oxygen, or from the heat of the reaction with the hydrogen. The beaker exploded with a small "pop" and ethanol was propelled onto both employees. The ethanol ignited, and Employees #1 and #2 ran to the emergency shower in the hallway, only to have the chain come off in their hands. They finally were able to extinguish the flames in the second emergency shower, but not before suffering third-degree burns on their arms, torsos, and thighs. The employees had not carried out the reaction before with this amount of material.

Keywords: BURN, HYDROGEN, FLAMMABLE LIQUID, TORSO, EQUIPMENT FAILURE, ETHANOL, CHEMICAL REACTION, ARM, LEG

Investigated Inspection
# Inspection Age Sex Degree of Injury Nature of Injury Occupation
1 112063516 Hospitalized injury Burn/Scald(Heat) Chemists, except biochemists
2 112063516 Hospitalized injury Burn/Scald(Heat) Clinical Laboratory technologists and technicians
Back to Top

Thank You for Visiting Our Website

You are exiting the Department of Labor's Web server.

The Department of Labor does not endorse, takes no responsibility for, and exercises no control over the linked organization or its views, or contents, nor does it vouch for the accuracy or accessibility of the information contained on the destination server. The Department of Labor also cannot authorize the use of copyrighted materials contained in linked Web sites. Users must request such authorization from the sponsor of the linked Web site. Thank you for visiting our site. Please click the button below to continue.