Inspection Detail
Inspection: 103827747 - Finlay Testing Laboratories Inc
Inspection Information - Office: Hi-E-Honolulu
Site Address:
Finlay Testing Laboratories Inc
99-940 Iwaena St
Aiea, HI 96701
Mailing Address:
, , 00000
Union Status: NonUnion
SIC:8734
NAICS: 0
Inspection Type: Accident
Scope: Complete
Advanced Notice: N
Ownership: Private
Safety/Health: Health
Close Conference: 03/06/1991
Planning Guide: Health-Manufacturing
Emphasis:
Case Closed: 05/24/1991
| Type | Activity Nr | Safety | Health |
|---|---|---|---|
| Accident | 360195267 |
| Violations/Penalties | Serious | Willful | Repeat | Other | Unclass | Total |
|---|---|---|---|---|---|---|
| Initial Violations | 4 | 5 | 9 | |||
| Current Violations | 4 | 5 | 9 | |||
| Initial Penalty | $600 | $0 | $0 | $40 | $0 | $640 |
| Current Penalty | $600 | $0 | $0 | $40 | $0 | $640 |
| FTA Penalty | $0 | $0 | $0 | $0 | $0 | $0 |
| # | Citation ID | Citaton Type | Standard Cited | Issuance Date | Abatement Due Date | Current Penalty | Initial Penalty | FTA Penalty | Contest | Latest Event | Note |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 1. | 01001 | Other | 510002 A | 03/11/1991 | 02/27/1991 | $40 | $40 | $0 | - | ||
| 2. | 02001 | Serious | 2030005 A | 03/11/1991 | 04/15/1991 | $120 | $120 | $0 | - | ||
| 3. | 02002 | Serious | 2030007 M | 03/11/1991 | 04/15/1991 | $120 | $120 | $0 | - | ||
| 4. | 02003 | Serious | 2030008 A | 03/11/1991 | 04/15/1991 | $120 | $120 | $0 | - | ||
| 5. | 03001 | Other | 630004 C01 | 03/11/1991 | 02/27/1991 | $0 | $0 | $0 | - | ||
| 6. | 03002 | Other | 670004 D | 03/11/1991 | 02/27/1991 | $0 | $0 | $0 | - | ||
| 7. | 03003 | Other | 2020001 E | 03/11/1991 | 04/15/1991 | $0 | $0 | $0 | - | ||
| 8. | 04001A | Serious | 2010001 | 05/03/1991 | 01/28/1991 | $240 | $240 | $0 | - | ||
| 9. | 04001B | Serious | 2010002 I02 | 05/03/1991 | 05/20/1991 | $0 | $0 | $0 | - | ||
| 10. | 05002 | Other | 2010002 I03 | 05/03/1991 | 02/27/1991 | $0 | $0 | $0 | - |
Investigation Summary
On January 28, 1990, Employees #1 and #2 were using a Magnoflux X-ray machine to radiographically inspect a Beech 18 aircraft. At approximately 1:30 p.m., they prepared to take a picture by placing the X-ray tube above and the film below the wing. The timer, located on the control panel 75 ft away from the X-ray tube, was set for 6.0 minutes. After 6.0 minutes had elapsed, the X-ray tube appeared to turn off. Employees #1 and #2 then repositioned the X-ray tube and laid down new film while preparing to take another picture. A humming noise was heard coming from the X-ray tube and, when the employees went back to the control panel, the elapsed time indicator now read 11.4. The employee's fingers had been in the X-ray beam only a few seconds out of the 5.4 minutes that the tube had remained on without their knowledge. Employees #1 and #2 were hospitalized.
Keywords: WORK RULES, OVEREXPOSURE
| # | Inspection | Age | Sex | Degree of Injury | Nature of Injury | Occupation |
|---|---|---|---|---|---|---|
| 1 | 103827747 | Hospitalized injury | Radiation Effects | Inspectors, testers and graders | ||
| 2 | 103827747 | Hospitalized injury | Radiation Effects | Inspectors, testers and graders |
Translate