Accident Report Detail
Accident Summary Nr: 201503661 - Employee's eye injured by exposure to laser radiation
| Inspection Nr | Date Opened | SIC | NAICS | Establishment Name |
|---|---|---|---|---|
| 125519223 | 03/28/2003 | 8221 | 611310 | Uc Berkeley |
Abstract: On March 14, 2003, an employee of U C Berkeley, was working with a class 4 Nd:YAG near infrared open beam laser. He wore laser safety eyewear during beam alignment, but afterwards he removed safety eyewear. While taking a meter measurement for output energy, his eye caught a flash from a hidden optic in the beam's path. His exposure exceeded current limits for lasers. The exposure to laser damaged the macula in his retina, resulting in irreversible retinal damage and loss in visual acuity.
| Employee # | Inspection Nr | Age | Sex | Degree of Injury | Nature of Injury | Occupation |
|---|---|---|---|---|---|---|
| 1 | 125519223 | Non Hospitalized injury | Other | Chemistry teachers |