Standard Cited: 19100303 B02 General.
Inspection Nr: 1399494.015
Citation: 01005C
Citation Type: Serious
Abatement Status: Abatement Completed
Initial Penalty: $0.00
Current Penalty: $3,500.00
Issuance Date: 10/23/2019
Nr Instances: 1
Nr Exposed: 1
Abatement Date: 02/28/2020
Gravity: 5
Report ID: 0213900
Contest Date: 11/19/2019
Final Order: 02/24/2020
Related Event Code (REC):
Emphasis:
Type | Latest Event | Event Date | Penalty | Abatement Due Date | Citation Type | Failure to Abate Inspection |
---|---|---|---|---|---|---|
Penalty | J: ALJ Decision | 02/24/2020 | $3,500.00 | 02/28/2020 | Serious | |
Penalty | C: Contested | 11/19/2019 | $0.00 | 11/19/2019 | Serious | |
Penalty | Z: Issued | 10/23/2019 | $0.00 | 11/19/2019 | Serious |
Text For Citation: 01 Item/Group: 005C Hazard:
29 CFR 1910.303(b)(2): Listed or labeled electrical equipment was not used or installed in accordance with instructions included in the listing or labeling: a) Wilmad LabGlass Inc. DBA SP Scienceware - Grinding Area: Metallic knockout boxes were used as pendant drop down outlets to connect equipment including the concentric machine model #102T5K-132, Occurred on or about 05/08/2019. b) Wilmad LabGlass Inc. DBA SP Scienceware - Grinding Area: A power strip was used to connect multiple equipment around the grinding area, observed on or about 05/08/2019. c) Wilmad LabGlass Inc. DBA SP Scienceware - Quartz Glass Area: A power strip was used to connect multiple equipment including a sum pump and the Shopcraft Drill Press Model #T7000-20P, S/N 00936, observed on or about 05/08/2019.