Accident Summary Nr: 202489696 - Employee Is Injured When Autoclave Door Blows Off

Accident Summary Nr: 202489696 -- Report ID: 0950621 -- Event Date: 06/19/2009
Inspection NrDate OpenedSICNAICSEstablishment Name
31322857907/22/20098071621511Siemens Healthcare Diagnostics

Abstract: At 1:30 p.m. on June 19, 2009, Employee #1 was walking through a room where medical testing equipment was manufactured. An autoclave door (5 ft square) blew off, striking a wall approximately 12 ft away. Employee #1 was struck by glass and burned by steam. He was hospitalized at U.C Davis Medical center for lacerations and first-degree chest and face burns. Employee #1 was taking a short-cut through this area at the time of the incident. The autoclave had been under 30 psi pressure, resulting in the equivalent of approximately 72 tons of pressure on the door. After the incident, the employer implemented new maintenance procedures and installed different door hardware.

Keywords: BURN, CHEST, STEAM, EXPLOSION, LACERATION, FACE, DOOR, PRESSURE RELEASE

Employee Details
Employee # Inspection Nr Age Sex Degree of Injury Nature of Injury Occupation
1 313228579 Hospitalized injury Burn/Scald(Heat) Clinical Laboratory technologists and technicians

Thank You for Visiting Our Website

You are exiting the Department of Labor's Web server.

The Department of Labor does not endorse, takes no responsibility for, and exercises no control over the linked organization or its views, or contents, nor does it vouch for the accuracy or accessibility of the information contained on the destination server. The Department of Labor also cannot authorize the use of copyrighted materials contained in linked Web sites. Users must request such authorization from the sponsor of the linked Web site. Thank you for visiting our site. Please click the button below to continue.

Close