Powered by GoogleTranslate

Accident Report Detail

Accident Summary Nr: 200851624 - Employee dies of asphyxia from cryogenic nitrogen

Accident Summary Nr: 200851624 -- Report ID: 0215000 -- Event Date: 09/20/2000
Inspection NrDate OpenedSICNAICSEstablishment Name
30294259409/20/200038450Ge Medical Systems

Abstract: Employee #1 was working alone to install an MRI. The process required cryogenic nitrogen to cool the magnet. While the nitrogen was being applied to the magnet, it apparently leaked into the work area. Employee #1 died of asphyxia.

Keywords: LEAK, ASPHYXIATED, INSTALLING, VAPOR, INHALATION, GAS, NITROGEN, CHEMICAL VAPOR, CHEMICAL

Employee Details
Employee # Inspection Nr Age Sex Degree of Injury Nature of Injury Occupation
1 302942594 Fatality Asphyxia Occupation not reported

Thank You for Visiting Our Website

You are exiting the Department of Labor's Web server.

The Department of Labor does not endorse, takes no responsibility for, and exercises no control over the linked organization or its views, or contents, nor does it vouch for the accuracy or accessibility of the information contained on the destination server. The Department of Labor also cannot authorize the use of copyrighted materials contained in linked Web sites. Users must request such authorization from the sponsor of the linked Web site. Thank you for visiting our site. Please click the button below to continue.