Powered by GoogleTranslate

Accident Report Detail

Accident Summary Nr: 170208946 - Employee exposed to ethylene oxide

Accident Summary Nr: 170208946 -- Report ID: 0950614 -- Event Date: 05/02/1991
Inspection NrDate OpenedSICNAICSEstablishment Name
11183100405/20/199180620Humana Hospital

Abstract: A central supply worker loaded the ethylene oxide (EtO) sterilizer and started the cycle. A sensor alarm light came on indicating empty tanks, and the conditioning cycle did not start. The employee pressed the RESET button, but the sterilizer still did not switch back on. Employee #1 then turned on the exhaust fan, waited 15 minutes, opened the sterilizer door, and left the area. Alarms subsequently sounded in two areas. Employee #1 went to the emergency assembly area and was sent to the emergency room for medical surveillance. He showed no symptoms and was released. There were probably trace amounts of EtO in the piping to the sterilizer that triggered the alarm to go off. The employee was an experienced worker , but he failed to notify the supervisor when the sterilizer did not work. The employee has since received retraining.

Keywords: ETHYLENE OXIDE, WORK RULES, INHALATION, GAS LEAK, CHEMICAL VAPOR, ACCIDENTAL DISCHARGE, MECH MALFUNCTION, CHEMICAL

Employee Details
Employee # Inspection Nr Age Sex Degree of Injury Nature of Injury
1 111831004

Thank You for Visiting Our Website

You are exiting the Department of Labor's Web server.

The Department of Labor does not endorse, takes no responsibility for, and exercises no control over the linked organization or its views, or contents, nor does it vouch for the accuracy or accessibility of the information contained on the destination server. The Department of Labor also cannot authorize the use of copyrighted materials contained in linked Web sites. Users must request such authorization from the sponsor of the linked Web site. Thank you for visiting our site. Please click the button below to continue.

Close