Hydrogen Cyanide
[324 KB PDF, 22 pages]
Related Information: Chemical Sampling - Hydrogen Cyanide
| Method Number: |
1015 |
| |
|
| Version: |
1.0 |
| |
|
Target concentration: OSHA PEL: ACGIH TLV: |
10ppm (11mg/m3) |
| |
|
| Procedure: |
Diffusive samples are collected by exposing ULTRA II Passive Samplers, field loaded with 600 mg of soda lime, to workplace air.
Samples are extracted with water and analyzed by ion chromotography using an electrochemical detector. |
| |
|
Recommended sampling time: Sampling rate: |
240 min 28.4 mL/min and 25 °C and 760 mmHg |
| |
|
| Reliable quantitation limit: |
0.44 ppm (0.48 mg/m3) |
| |
|
| Standard error of estimate at the target concentration: |
9.8%*
*For diffusive samplers when sampling site atmospheric pressure and temperature are known. when either or both of these values are unknown, see Section 4.5 for applicable standard errors of estimate. |
| |
|
| Special requirements: |
When using an ULTRA II Passive Sampler report the temperature and uncorrected sampling site atmospheric pressure. |
| |
|
| Status of method: |
Fully validated method. This method has been subjected to the established validation procedures of the Methods Development Team. |
| |
|
|
November 2010 |
Michael Simmons |
| |
|
| |
|
Methods
Development Team
Industrial Hygiene Chemistry Division
OSHA Salt Lake Technical Center
Sandy UT 84070-6406 |
1. General Discussion
For assistance with accessibility problems in using figures and illustrations presented in this method, please contact the Salt Lake Technical Center (SLTC) at (801) 233-4900. This procedure was designed and tested for internal use by OSHA personnel. Mention of any company name or commercial product does not constitute endorsement by OSHA.
1.1 Background
1.1.1. History
Previous to this method OSHA used a 37-mm mixed cellulose ester filter, followed by a bubbler containing
10 mL of 0.1 N NaOH, for the collection of hydrogen
cyanide and particulate cyanide with analysis by ion specific electrode (OSHA Method ID-120).1 In addition, OSHA
occasionally used a solid sorbent tube
containing a glass fiber pre-filter, followed by two sections of soda lime, for sampling hydrogen cyanide (NIOSH Method 6010).2 In theory both sampling
methods work by collecting particulate cyanide on the prefilter and hydrogen
cyanide using the bubbler or sorbent. Particulate cyanide, however, will decompose on the prefilter forming hydrogen cyanide,
which then passes through the filter and is collected by the bubbler or sorbent, resulting in inaccurate hydrogen cyanide and particulate cyanide exposure results.
A study by OSHA indicates that as much as 40% of particulate cyanide can be stripped off the pre-filter and collect in the bubbler when using OSHA Method
ID-120.3 Since OSHA has separate PEL's for particulate cyanide and hydrogen cyanide, neither method can
insure accurate hydrogen cyanide exposure results. The use of a passive sampler for monitoring hydrogen cyanide eliminates the potential for particulate cyanide
decomposition within the sampler helping to insure more accurate determinations of personal workplace exposure to hydrogen cyanide.
Soda lime, the same medium used in NIOSH Method 6010, was chosen as the sampling medium. Soda lime is a mixture of chemicals consisting of mostly calcium hydroxide
but also 6-18% water and 5-20% sodium hydroxide.4 The ULTRA II Passive Sampler was selected since the sampling medium can easily be loaded and unloaded
from the sampler in the field. Limiting the time the sampling medium spends in the sampler is desirable because it is caustic and sometimes clumps together.
Samples are desorbed using water and analyzed by ion chromatography using pulsed amperometric detection with disposable silver working electrodes.
The disposable silver working electrode, sold by Dionex Corporation, along with the detection waveforms used in this method were developed and described by Cheng
et al.5
When using OSHA Method ID-120 the presence of sulfide, which causes a positive interference, was determined by placing a drop of the bubbler sample solution on lead
acetate paper. If the lead acetate paper changed red then cadmium carbonate was added to the sample to remove the sulfide interference. However, addition of excess
cadmium carbonate could result in low recoveries by cyanide. In this method sulfide is
separated from cyanide by the analytical column. However, sampling hydrogen
cyanide in the presence of hydrogen sulfide will result in some cyanide reacting with sulfide on the sample medium producing thiocyanate. If cyanide is suspected to
have reacted with sulfide, indicated by the detection of sulfide, the sample should also be analyzed for thiocyanate. The micrograms thiocyanate found is mathematically
converted into equivalent micrograms cyanide and used in the final exposure calculation.
1.1.2. Toxic effects (This section is for information only and should not be taken as the basis of OSHA policy).6
Human physiological response to various air concentrations of hydrogen cyanide are listed below in Table 1.1.2. Workplace exposure can occur from inhalation
and through skin absorption. Reported symptoms of exposure to hydrogen cyanide include "lightheadedness, breathlessness, feeling shaky, headache, and nausea".
Individuals who have become exposed to acute levels of hydrogen cyanide, which did not result in death, have shown a complete and rapid recovery once removed from the
source of exposure. Reported symptoms of chronic exposure to cyanide include "headache, weakness, irritation of the throat, vomiting, and effort dyspnea".
Thyroid gland enlargement along with "lacrimation, abdominal colic, salivation, and nervous instability" has also been reported.
Table 1.1.2 Human Physiological Response to Various Air Concentrations of HCN†
|
| Response |
concn (ppm) |
|
| immediately fatal |
270 |
| fatal after 10 min |
181 |
| fatal after 30 min |
135 |
| fatal after 0.5-1 hr or lager (dangerous to life) |
110-135 |
| tolerated for 0.5-1 hr without immediate or late effects |
45-54 |
| slight symptoms after several hours |
18-36 |
|
| †Adapted from Hartung7 |
1.1.3 Workplace exposure
Hydrogen cyanide is used as a chemical building block in the production of chemicals used in various industries. Estimates of use "for
hydrogen cyanide in the
United States are adiponitrile for nylon, 41%; acetone cyanohydrin for acrylic plastics, 28%; sodium cyanide for gold recovery, 13%; cyanuric chloride for pesticides
and other agriculture products, 9%; chelating agents such as EDTA, 4%; and methionine for animal feed, 2%". Other minor uses of
hydrogen cyanide include the
"manufacturing of ferrocyanides, acrylates, lactic acid, pharmeceuticals, and specialty chemicals". The use of
hydrogen cyanide as a fumigant is decreasing.8
United States production of hydrogen cyanide in 2003 was 2.019 billion pounds.9
1.1.4 Physical properties and descriptive information10,11
| synonyms: |
hydrocyanic acid; prussic acid; formonitrile |
| IMIS number:12: |
1440 |
| CAS number: |
74-90-8 |
| molecular weight: |
27.03 |
| boiling point: |
25.6 °C |
| melting point: |
-13.4 °C |
| molecular formula: |
HCN |
| vapor pressure: |
620 mmHg at 20 °C |
| vapor density: |
0.941 (air = 1.0) |
| appearance: |
colorless gas |
| odor: |
some individuals may detect a bitter almond odor |
| solubility: |
miscible with water, alcohol; slightly soluble in ether |
| flash point: |
-17.78 °C, closed cup |
| explosive limits: |
lower, 5.6%; upper, 40% by volume in air |
This method was validated according to the OSHA SLTC "Validation Guidelines for Air Sampling Methods Utilizing Chromatographic Analysis"
13. The guidelines define analytical parameters, specify required laboratory tests, statistical calculations and acceptance
criteria. The analyte air concentrations throughout this method are based on the recommended sampling and analytical
parameters. Air concentrations listed in ppm are referenced to 25 °C and
760 mmHg (101.3 kPa).
2. Sampling procedure
All safety practices that apply to the work area being sampled should be followed. The sampling equipment should be attached to the worker in a manner that will not
interfere with work performance or safety.
2.1 Apparatus
2.1.1 ULTRA II Passive Sampler
Samples are collected with an ULTRA II Passive Sampler field loaded with 600 mg or soda lime from a sorbent vial. OSHA policy is to not reuse the sampler.
For this validation the ULTRA II Passive Samplers were purchased from SKC Inc. (catalog no. 590-259). The soda lime, purchased in bulk from SKC Inc., is the same
medium placed in an active sorbent tube (catalog no. 226-28) also sold by SKC Inc. Six hundred
milligrams of soda lime was placed into 4-mL sorbent vials, with PTFE-lined
cap with threads compatible with the vial adapter on the ULTRA II, and then used to load the sampler just prior to sampling.
2.1.2 A thermometer and barometer are required to determine the sampling site air temperature and
atmospheric pressure.
2.2 Reagents
None Required
2.3 Technique
2.3.1 SKC Inc. ULTRA Passive II Passive Sampler
Remove the sampler from it's re-sealable plastic package.
Remove the white plastic plug from the back of the sampler. Remove the cap from the sorbent vial and carefully screw the sorbent vial onto the sampler vial
adaptor without spilling any soda lime. Invert the sampler and vial assembly and gently tap the sampler until all the soda lime has been transferred into the sampler.
Unscrew the sorbent vial and reinsert the white plastic plug into the back of the sampler. Place the cap and the sorbent vial into the re-sealable plastic package
for storage. Do not throw away the cap and sorbent vial. It is recommended that the sampler be loaded immediately before use. If this is not possible, then load the
sampler on the day of use.
Record the start time on the sampler label and on the form OSHA-91A. Remove the cover when ready to being sampling. CAUTION- The Sampler immediately begins to sample
when the cover is removed.
Attach the sampler to the worker near his/her breathing zone with the perforations in the sampler facing forward. Assure that the area directly in front of the sampler
unobstructed through the sampling period.
Sample for up to 240 min when using ULTRA II Passive Samplers loaded with soda lime to collect
TWA (long-term) samples. The sampling rate for hydrogen cyanide using ULTRA II
Passive Samplers loaded with soda lime is 28.4 mL/min at 25 °C and 760 mmHg.
Sample for 15 min when using ULTRA II Passive Samplers loaded with soda lime to collect short-term samples. The sampling rate for
hydrogen cyanide using ULTRA II
Passive Samplers loaded with soda lime is 28.4 mL/min at 25 ° C and 760 mmHg.
At the end of the sampling period, immediately detach the sampler from the worker and attach the cover. Record the stop time on the sampler label and also on the
Form OSHA-91A.
Hold the sampler with the plastic plug facing upwards and remove the plug. Screw the sorbent vial onto the sampler vial adaptor and invert the sampler and vial
assembly. Gently tap or shake the sampler until all the soda lime has been transferred back into the sorbent vial. Detach the sorbent vial from the sampler and attach
the vial cap. Place the vial and sampler in its re-sealable plastic package, close the package, and seal it with a Form OSHA-21.
Prepare a blank in a low background area by loading and then unloading sampler as described above. Return the vial to its re-sealable plastic package, close the
package, and seal it with a Form OSHA-21.
Verify that the sampling times are properly recorded on the form OSHA 91-A for each sample. Also, identify blank samples on this form.
Record the room temperature and atmospheric pressure at the sampling site on the Form OSHA-91A.
List any compounds that could be considered potential interferences, especially hydrogen sulfide.
Submit the samplers to the laboratory for analysis as soon as possible. If delay is unavoidable, store the samples in a refrigerator as a precaution.
Ship any bulk sample(s) in a container separate from the air samples.
3. Analytical procedure
Adhere to the rules set down in your laboratory's Chemical Hygiene Plan14 (for example OSHA SLTC adheres to: "The OSHA SLTC
Chemical Hygiene Plan"). Avoid skin contact and inhalation of all chemicals and review all appropriate MSDS's before beginning the analytical procedure. Follow all applicable
quality assurance practices established in your laboratory's internal quality system (for example OSHA SLTC follows: "The OSHA SLTC Quality Assurance Manual").
3.1 Apparatus
Ion chromotograph with an electrochemical and conductivity detector. A dionex ISC-3000 Ion Chromatography System equipped with an
Integrated DC Detector/Chromatography
module, an AS Auto-sampler with a 100 place tray, an SP Pump, an EO Eluent
Organizer with pressure regulator, an ED Electrochemical detector and a CD Conductivity
detector were used in this validation.
IC column and guard column that can separate cyanide from potential
interferences. A Dionex IonPac® AS15 analytical column (250-mm × 4-mm i.d.) and a
Dionex IonPac® AG15 guard column (50-mm × 4-mm i.d.) were used in this validation.
IC column and guard column that can separate thiocyanate from potential interferences. A Dionex IonPac® AS16 analytical column (250-mm × 4-mm i.d.)
and a Dionex IonPac® AG16 guard column (50-mm × 4-mm i.d.) were used in this validation.
A means to integrate chromatograms. Dionex Chromeleon software was used in this validation.
A mechanical rotator. A Fisher Roto-Rack that accepts 4-mL vials was used in this validation.
Autosampler vials. National Scientific Company clear glass 2-mL screw thread vials with PTFE- lined septum caps were used in this validation.
Polypropylene vial racks for 2 and 4-mL vials.
Analytical balance capable of weighing at least 0.01 mg. An Ohaus Galaxy 160D balance was used in this validation.
A means to dispense and dilute solutions. A Hamilton Microlan 540B dual syringe diluter/dispenser was used in this validation.
Water purifier. A Barnstead NANOpure Diamond system was used to produce 18.0 Ω-cm DI water in this validation.
Syringe and filters certified for ion chromatography for the filtration of sample solutions prior to analysis. Pall Corporation ion chromatography certified
Acrodisc® 25 mm syringe filters with 0.2 µm Supor® (PES) membranes were used in this validation.
Class A 2-L volumetric flask.
Standard preparation vials. Wheaton 20-mL glass scintillation vials were used in this validation.
3.2 Regents
Sodium Hydroxide (NaOH), [CAS no. 1310-73-2]. The sodium hydroxide used in this validation was a 99.995% trace metal basis grade (lot no. 17130TD) purchased from
Sigma Aldrich (Milwaukee, WI).
Cyanide (CN‾) 1000 mg/L standard solution. The 1000 mg/L (1000 µg/mL)
cyanide standard used in this validation was (lot no. 1013925) purchased from
CPI International (Santa Rosa, CA).
Thiocyanate (SCN‾) 1000 mg/L standard solution. The 1000 mg/L (1000 µg/mL)
cyanide standard used in this validation was (lot no. 1097227) purchased from Ricca
Chemical Company (Arlington, TX).
Eluent [62.5 mM NaOH]: Add 5 g of NaOH to a 2-L volumetric flask, dilute to mark with 18.0 mΩ-cm DI water and mix well. Transfer eluent to appropriate
container(s) and degas. Store eluent under a helium blanket to avoid reaction with carbon dioxide in the air.
Extraction solvent. The extraction solvent used is 18.0 mΩ-cm DI water.
The extraction efficiency is affected by extraction solvent, the sampling
medium, and the technique used to extract the samples. Other reagents and
techniques than those described in this method can be used provide they are tested as specified
in the validation guidelines15.
3.3 Standard preparation
Prepare a concentrated stock standard of 100 µg/mL using a 1000 µg/mL cyanide standard using the eluent as the
diluent. Prepare working analytical standards
by diluting milliliter amounts of the concentrated stock standard with measured amounts of eluent. For example, to prepare a target level of standard of 73.5 µg/sample
cyanide, add 2.1 mL of 100 µg/mL concentrated stock into a 10 mL volumetric flask and solution and dilute to the mark with eluent. If necessary, prepare
thiocyante working analytical standards in the same manner.
Bracket sample concentrations with standard concentrations. If upon analysis, sample concentrations fall outside the range of prepared standards, prepare and analyze
additional standards to confirm instrument response, or dilute high samples with extraction solvent and reanalyze the diluted samples.
3.4 Sample preparation
For each sample break the Form OSHA-21 seal, remove the 4-mL sorbent vial from the re-sealable plastic package and write the sample number on the vial.
Add 3.5 mL of 18.0 MΩ-cm DI water to each vial.
Place the 4-mL vials on a mechanical rotator and rotate at approximately 40 rpm for 60 min.
Remove the vials from the rotator and allow samples to settle for 15 min.
Filter 2 mL of each sample and transfer the filtrate to 2-mL glass auto-sampler vials and seal with PTFE caps.
Analyze as described in Section 3.5
3.5 Analysis
Sampling hydrogen cyanide in the presence of hydrogen sulfide will result in some cyanide reacting with sulfide on the sample medium producing thiocyanate. If cyanide
is suspected to have reacted with sulfide, indicated by the detection of sulfide, the sample should also be analyzed for thiocyanate. The micrograms thiocycnate
found is converted mathematically into equivalent micrograms cyanide and used in the final exposure calculation (see Section 3.7).
Analyze all samples for Cyanide as described in Section 3.5.1. If cyanide and sulfide are both detected analyze the sample for thiocyanate as described in Section
3.5.2
3.5.1 Analytical conditions for analysis of cyanide
| IC conditions |
|
| column: |
Dionex IonPac® AS15 column (250-mm × 4-mm i.d.) and
AG15 guard column (50-mm × 4-mm i.d.) |
| run time: |
20 min |
| injection volume: |
25.0 µL* |
| eluent: |
62.5 mM NaOH |
| flow rate: |
1.0 mL/min |
| pump pressure: |
~ 1000 psi |
| column temperature: |
30 °C |
| retention times: |
~4.8 min (sulfide) ~5.8 min (cyanide) ~17.2 min (thiosulfate) |
| *Injection volume can be increased if lower quantation limits are desired for the analysis of short term samples. |
| Electrochemical detector conditions |
|
| detector mode: |
pulsed amperometric |
| reference electrode: |
Ag/AgCI (3 M KCI) |
| working electrode: |
silver (disposable electrode) |
| counter electrode: |
titanium (detector cell body) |
| waveform: |
| cyanide waveform† |
|
| time(s) |
potential (V) |
gain |
ramp |
integration |
|
| 0.00 |
-0.100 |
Off |
Off |
Off |
| 0.20 |
-0.100 |
On |
Off |
On |
| 0.90 |
-0.100 |
Off |
Off |
Off |
| 0.91 |
-1.000 |
Off |
Off |
Off |
| 0.93 |
-0.300 |
Off |
Off |
Off |
| 1.00 |
-0.300 |
Off |
Off |
Off |
|
| †Use manufacture recommended waveform |
 |
Figure 3.5.1.1.
Chromatogram obtained at the target concentration with the recommended analytical conditions. |
 |
Figure 3.5.1.2.
Chromatogram obtained when a sample was taken
from a test atmosphere containing the target concentration of HCN and 11 ppm of
H2S. |
3.5.2 Analytical conditions for analysis of thiocyanate
Thiocyanate may be analyzed by pulsed amperometric detection with a silver electrode using the same column and eluent as described below.
16 However, using a conductivity detector is preferable due to the larger linear dynamic range of the detector.
| IC conditions |
|
| column: |
Dionex IonPac® AS16 column (250-mm × 4-mm i.d.) and
AG16 guard column (50-mm × 4-mm i.d.) |
| run time: |
11.5 min |
| injection volume: |
25.0 µL* |
| eluent: |
6235 mM NaOH |
| flow rate: |
1.0 mL/min |
| pump pressure: |
~1600 PSI |
| column temperature: |
30 °C |
| retention times: |
~9.8 min (thiocyanate) |
*Injection volume can be increased if lower quantitation limits are desired for the analysis of short term samples |
| Conductivity detector and suppressor conditions |
|
| cell heater temp: |
35 °C |
| temp compensation: |
1.7% / °C |
| suppressor (current): |
4-mm ASRS (155 mA) |
 |
Figure 3.5.2.
Chromatogram obtained at the concentration of 31.5 µg/sample of thiocyanate with the recommended analytical conditions. |
3.5.3 An external standard calibration method is used. A calibration curve can be constructed by plotting response of standard injections
versus micrograms of analyte per sample. Bracket the samples with freshly prepared analytical standards over a range of concentrations.
 |
Figure 3.5.3.1.
Calibration curve for cyanide (y= 4.09× = 3.98). |
 |
Figure 3.5.3.2.
Calibration curve for thiocyanate (y = 7E-05x2 + 0.0292x - 0.009). |
3.6 Interferences (analytical)
3.6.1 Any compound that produces a detector response and has similar retention time as the analyte is a potential interference. If any potential
interferences were reported, they should be considered before samples are extracted.
3.6.2 When necessary, the identity of any analyte peak can be confirmed with additional analytical techniques (Section 4.10).
3.7 Calculations
The amount of hydrogen cyanide per sample is obtained from the appropriate calibration curve in terms of micrograms per sample, uncorrected for extraction efficiency.
This total amount is then corrected by subtracting the total amount (if any) found in the blank. The air concentration is calculated using the following formulas.
|
RSS= |
RNTP |
( |
TSS
TNTP |
) |
3 2 |
( |
PNTP
PSS |
) |
where |
RSS is the sampling rate at the sampling site (mL/min)
RNTP is the sampling rate at NTP (28.4 mL/min)
TSS is the temp at the sampling site (K)
TNTP is 298.2 K
PSS is the sampling site pressure (mmHg)
PNTP is 760 mmHg |
|
|
CM= |
[( |
MCNGFCN
EE(CN) |
) + ( |
MSCNGFSCN
EE(SCN) |
)] 1000 |
where |
CM is HCN concn by weight in air (mg/m3)
MCN is micrograms per sample CN-
MSCN is micrograms per sample SCN-
GFCN is the CN- gravimetric factor (1.039 HCN/CN-)
GFCN is the SCN- gravimetric factor (0.465 HCN/SCN-)
EE(SCN) is the SCN- extraction efficiencey (decimal form)
RSS is the sampling rate at the sampling site (mL/min)
t is the sampling time (min) |
RSSt |
|
| CV = |
CMVM
Mr |
where |
CV is concn of HCN by volume (ppm)
VM is 24.46 (molar volume at NTP)
CM is concn by weight (mg/m3)
Mr is molecular weight of HCN (27.03) |
If the sampling site temperature is not provided, assume that it is 22.2 °C. If the sampling site atmospheric pressure is not given, calculate
an approximate value based on the sampling site elevation from the following equation.
| PSS=AE2-BE + 760 |
where |
PSS is the approximate atmospheric pressure (mmHg) |
| |
|
E is the sampling site elevation (ft) |
| |
|
A is 3.768 × 10-7 mmHg/ft2 |
| |
|
B is mmHg/ft |
4. Method Validation
General instruction for the laboratory validation of OSHA sampling and analytical methods that employ chromatographic analysis is presented in "Validation Guidelines
for Air Sampling Methods Utilizing Chromatographic Analysis"17. These
Guidelines detail required validation tests, show
examples of statistical calculations, list validation acceptance criteria, and define analytical parameters. Air concentrations listed in ppm are referenced to 25 °C and 760 mmHg (101.3 kPa).
4.1 Detection limit of the analytical procedure (DLAP)
The DLAP is measured as the mass of analyte introduced onto the chromatographic column. Ten analytical standards were prepared with equally descending increments of
cyanide with the highest standard containing 1.2 µg/mL. This is the concentration that would produce a peak approximately 10 times the response of a reagent
blank at or near the chromatographic retention time of the analyte. These standards and the reagent blank at or near the chromatographic retention time of the analyte.
These standards and the reagent blank were analyzed with the recommended analytical parameters (25-µL injection volume). The data obtained were used to
determine the required parameters (standard error of estimate and slope) for the calculation of the DLAP. Values of 0.46 and 0.80 were obtained for the slope and
standard error of estimate respectively. The calculated DLAP of 5.2 ng did not produce a detector response so the DLAP was set to 7.5 ng cyanide.
Table 4.1 Detection Limit of the Analytical Procedure |
|
concentration (µg/mL cyanide) |
mass on column (ng) |
area counts (nC*min) |
|
| 0.0 |
0.0 |
0.0 |
| 0.3 |
7.5 |
1.06 |
| 0.4 |
10.0 |
2.34 |
| 0.5 |
12.5 |
3.76 |
| 0.6 |
15.0 |
4.84 |
| 0.7 |
17.5 |
6.27 |
| 0.8 |
20.0 |
7.32 |
| 0.9 |
22.5 |
8.53 |
| 1.0 |
25.0 |
11.1 |
| 1.1 |
27.5 |
11.4 |
| 1.2 |
30.0 |
12.6 |
|
 |
Figure 4.1.
Plot of data to determine DLAP (y = 0.46 × - 1.63) |
4.2 Detection limit of the overall procedure (DLOP) and reliable
quantitation limit (RQL)
The DLOP is measured as mass per sample and expressed as equivalent air concentrations based on the recommended sampling parameters. Ten samplers were spiked with equally
descending increments of the analyte, such that the highest sampler loading was 4.55 µg/sample. This is the amount spiked on a sampler that would produce a peak
approximately 10 times the response of a sample blank at or near the chromatographic retention time of the analyte. These spiked samplers and the sample blank were
analyzed with the recommended analytical parameters, and the data obtained used to calculate the required parameters (standard error of estimate and the slope) for
the calculation of the DLOP. Values of 2.98 and 0.94 were obtained for the slope and standard error of estimate respectively. The calculated DLOP of 0.95 µg
cyanide did not produce a detector response so the DLOP was set to 1.4
µg/sample cyanide or 1.45 µg/sample hydrogen cyanide (0.19 ppm or 0.21 mg/m3).
Table 4.2 Detection Limit of Overall Procedure |
|
mass per sample (µg cyanide) |
area counts (nC/*min) |
|
| 0.00 |
0.00 |
| 1.40 |
0.92 |
| 1.75 |
2.33 |
| 2.10 |
3.72 |
| 2.45 |
4.83 |
| 2.80 |
5.99 |
| 3.15 |
7.18 |
| 3.50 |
8.41 |
| 3.85 |
9.70 |
| 4.20 |
10.95 |
| 4.55 |
12.32 |
|
 |
Figure 4.2.1.
Plot of data to determine the DLOP/RQL (y = 2.98x - 1.98).
|
The RQL is considered the lower limit for precise quantitative measurements. It is determined from the regression line parameters that were obtained
for the calculation of DLOP providing 75% to 125% of the analyte is recovered. The RQL is 3.2 µg/sample cyanide or 3.3 µg/sample
hydrogen cyanide (0.44 ppm
or 0.48 mg/m3). Recovery at this concentration is 97.8%.
When short term samples are collected, the air concentration equivalent to the reliable
quantitation limit becomes larger. For example, the reliable quantitation limit is 7.0
ppm (7.7 mg/m3) for hydrogen cyanide when 0.426 L is sampled (15 min sample). |
 |
Figure 4.2.2.
Chromatogram of the RQL for Cyanide. |
4.3 Precision of the analytical method
The precision of the analytical method measured as the mass equivalent to the standard error of estimate determined from the linear regression of data points from
standards over a range that covers 0.1 to 2 times the target concentration. A calibration curve was constructed and shown in Section 3.5.3 from the three injections
of five standards. The standard error of estimate was 1.5 µg/sample cyanide.
Table 4.3 Instrument Calibration |
|
| × target concn |
0.1× |
0.5× |
1.0× |
1.5× |
2.0× |
| (µg/sample cyanide) |
7.35 |
35.0 |
73.5 |
108.5 |
143.5 |
|
| area counts |
24.14 |
138.45 |
299.31 |
440.93 |
581.75 |
| (nC*min) |
26.03 |
139.67 |
297.04 |
444.68 |
583.25 |
| |
26.08 |
138.62 |
296.86 |
439.07 |
579.59 |
|
4.4 Storage stability test
Storage samples for hydrogen cyanide were prepared by collecting samples from a dynamically generated controlled test atmosphere using the recommended sampling time.
The concentration of hydrogen cyanide in the test atmosphere was the target concentration (9.8 ppm of 10.8 mg/m3), and the relative humidity was 80% at 21
°C. Eighteen storage samples were prepared. Three samples were analyzed on the day of generation. The other fifteen samples were stored in a closed drawer at ambient
temperature (about 20 °C). At 3-4 day intervals three samples were selected and analyzed. Sample results are not corrected for extraction efficiency. Results
were calculated using the sampling rate determined in Section 4.7 converted to its equivalent sampling rate at sampling site temperature and pressure.
Table 4.4 Storage Test for Hydrogen Cyanide |
|
time (days) |
ambient storage recovery (%) |
|
| 0 |
100.8 |
100.0 |
105.1 |
| 4 |
96.0 |
101.0 |
95.3 |
| 7 |
99.3 |
95.0 |
107.5 |
| 11 |
93.5 |
105.1 |
LIA |
| 14 |
102.1 |
103.7 |
97.3 |
| 18 |
93.4 |
97.3 |
104.1 |
|
| LIA = Lost In Analysis |
 |
Figure 4.4.
Ambient storage tests for hydrogen cyanide. |
4.5 Precision (overall procedure)
The precision of the overall procedure at the 95% confidence level is obtained by multiplying the standard error of estimate by 1.96 (the z-statistic from
the standard normal distribution at the 95% confidence level). Ninety-five percent confidence intervals are drawn about the regression line in the storage stability
figure shows in Section 4.4.
4.5.1 ULTRA II Passive Sampler precision
The precisions of the overall procedure at the 95% confidence level for the ambient
temperature (20 °C) 18-day storage test (at the target concentration) are given
in Table 4.5.1. They each include an additional 8.7% for sampling rate variation18,19. There are different values given, depending on whether both, either, or neither temperature (T) or atmospheric pressure (P) are known at the sampling site. If the sampling site
temperature is unknown, it is assumed to be 22.2 ± 15 °C (72 ± 27 °F) and variability of ±7.7% is included. If the atmospheric pressure is not known,
it is estimated from the sampling site elevation and a variability of ±3% is not included.
Table 4.5.1
Standard Error of Estimate and Precision of the Overall Procedure |
|
| known condition |
error (%) |
precision (±%) |
|
| both T and P |
9.79 |
19.2 |
| only T
| 10.2 |
20.0 |
| only P
| 12.5 |
24.5 |
| neither T nor P
| 12.8 |
25.1 |
|
4.5.2 Recovery
The recovery of hydrogen cyanide, in the form of cyanide, from samples used in an 18-day storage test remained above 99.1% when the samples were stored at 20 °C.
4.6 Reproducibility
Six samples were prepared by sampling a dynamically generated controlled test atmosphere similar to that used in the collection of the storage samples. The
concentration of hydrogen cyanide in the test atmosphere was the target concentration (9.86 ppm of 10.9 mg/m3), and the relative humidity was 80% at 21 °C.
The samples were submitted to the OSHA Salt Lake Technical Center for analysis. The samples were analyzed after being stored for 13 days at ambient temperature
(about 20 °C). Sample results were corrected for extraction efficiency. No sample result for hydrogen cyanide had a deviation greater than the precision of
the overall procedure determined in Section 4.5. Theoretical masses for hydrogen cyanide were determined using the sampling rate determined in Section 4.7 converted
to its equivalent sampling rate at sampling site temperature and atmospheric pressure.
Table 4.6.1 Reproducibility Data for Hydrogen Cyanide Collected on ULTRA II Passive Samplers Loaded with Soda Lime |
|
theoretical (µg/sample) |
recovered (µg/sample) |
recovery (%) |
deviation (%) |
|
| 86.1 |
83.5 |
97.0 |
-3.0 |
| 86.1 |
82.9 |
96.3 |
-3.7 |
| 86.1 |
83.3 |
96.7 |
-3.3 |
| 86.1 |
87.8 |
102.0 |
2.0 |
| 86.1 |
82.9 |
96.3 |
-3.7 |
| 86.1 |
84.4 |
98.0 |
-2.0 |
|
4.7 Sampling rate and capacity
The sampling rate and sampler capacity were determined with samples collected for increasing time intervals from a dynamically generated controlled test
atmosphere. The concentration of the test atmosphere was two times the target concentration (19.7 ppm or 21.8 mg/m3) at 80% relative humidity and 21 °C. The preliminary
sampling rate was determined by averaging the nine values for the 0.5, 1 and 2 hour samples. Horizontal lines were placed 10% above and 10% below the preliminary
sampling rate. The sampling rate is 28.4 mL/min at 760 mmHg and 25 °C and represents the average of all values between the lines. The standard deviation and RSD are 1.25 mL/min and 4.40% respectively. The data obtained are shown in Table 4.7 and Figure 4.7. Mass collected is corrected for extraction efficiency. The
recommended sampling time is 240 min (4 hours).
Sampler capacity is exceeded when the plotted sampling rate decreases rapidly as the sampler becomes saturated. Sampler capacity was not exceeded even after
10 hours of sampling at two times the target concentration.
Table 4.7 Determination of Sampling Rate and Time |
|
| sampling rate (mL/min) |
| time(h) |
first |
second |
third |
|
| 0.083 |
28.55 |
29.65 |
28.19 |
| 0.167 |
28.44 |
29.71 |
28.00 |
| 0.5 |
28.57 |
28.85 |
28.76 |
| 1 |
31.09 |
29.77 |
31.24 |
| 2 |
28.24 |
27.30 |
26.05 |
| 3 |
27.64 |
26.95 |
26.63 |
| 4 |
28.74 |
26.71 |
28.25 |
| 6 |
27.94 |
27.08 |
28.40 |
| 8 |
28.80 |
27.53 |
27.43 |
| 10 |
27.76 |
28.08 |
30.43 |
|
 |
Figure 4.7.
Plotted data to determine the sampling rate and recommended sampling time. |
4.8 Extraction efficiency and stability of extracted samples
The extraction efficiency is affected by the extraction solvent, the internal standard, the sampling medium, and the technique used to extract the samples.
Other reagents and techniques than described in this method can be used provided that they are tested as specified in the validation guidelines.20
Extraction Efficiency
The extraction efficiency of cyanide was determined by liquid spiking four vials, each containing 600 mg of soda lime, at each concentration level. These samples
were stored overnight at ambient temperature and then extracted and analyzed. The overall mean extraction efficiency over the range of 0.1 to 2 times the target
concentration was 97.9%. The extraction efficiency at the RQL was 91.2%. The presence of water had no significant effect on extraction efficiency. The extraction
efficiencies for the RQL and for the wet samplers are not included in the overall mean. Wet soda lime was prepared by exposing the samplers to humid air (80%
relative humidity at 21 °C) for 240 min. The data obtained are shown in Table 4.8.1.
Table 4.8.1 Extraction Efficiency of Cyanide from Soda Lime |
|
| Level |
Sample Number |
× target concn |
µg per sample |
1 |
2 |
3 |
4 |
mean |
|
| 0.1 |
7.48 |
96.0 |
94.8 |
93.6 |
93.6 |
94.5 |
| 0.25 |
18.7 |
101.2 |
96.2 |
97.6 |
97.1 |
98.0 |
| 0.5 |
37.4 |
98.4 |
99.2 |
98.3 |
99.2 |
98.8 |
| 1.0 |
74.8 |
97.7 |
97.9 |
97.4 |
97.3 |
97.6 |
| 1.5 |
112.2 |
100.2 |
99.7 |
100.1 |
100.0 |
100.0 |
| 2.0 |
150.6 |
100.4 |
99.7 |
97.0 |
96.7 |
98.4 |
| |
| RQL |
3.15 |
93.1 |
91.6 |
90.5 |
89.7 |
91.2 |
| 1.0(wet) |
74.8 |
97.0 |
97.9 |
96.9 |
97.1 |
97.1 |
|
Stability of extracted samples
The stability of extracted samples was examined by reanalyzing the target concentration samples 24, 48, and 72 hours after the initial analysis. After the original
analysis was performed two vials were recapped with new septa which were replaced after each reanalysis. The remaining two vials retained their punctured septa
throughout the test. All samples were allowed to stand in the autosampler tray at room temperature (20 °C). The samples were reanalyzed with freshly prepared
standards. Diff is the difference between the initial analysis and the subsequent analysis. Each septum was punctured one time for each injection. The data obtained
are shown in Table 4.8.2.
Table 4.8.2 Stability of Extracted Samples for Cyanide
| punctured septa replaced |
punctured septa retained |
|
initial (%) |
24 h (%) |
diff (%) |
48 h (%) |
diff %) |
72 h (%) |
diff (%) |
|
|
initial (%) |
24 h (%) |
diff (%) |
48 h (%) |
diff (%) |
72 h (%) |
diff (%) |
| 97.7 |
98.8 |
1.1 |
97.8 |
0.1 |
98.8 |
1.1 |
|
|
97.4 |
98.7 |
1.3 |
98.9 |
1.5 |
98.6 |
1.2 |
| 97.9 |
98.9 |
1.0 |
98.6 |
0.7 |
99.4 |
1.5 |
|
|
97.3 |
98.6 |
1.3 |
99.0 |
1.7 |
99.1 |
1.8 |
| mean |
|
|
mean |
| 97.8 |
98.8 |
1.0 |
98.2 |
0.4 |
99.1 |
1.3 |
|
|
97.4 |
98.6 |
1.3 |
99.0 |
1.6 |
98.8 |
1.5 |
|
4.9 Sampling interferences
The tested sampling interferences, with the exception of hydrogen sulfide, had no significant effect on the ability of ULTRA II Passive Samplers loaded
with soda lime to collect and retain hydrogen cyanide. If cyanide is suspected to have reacted with sulfide, indicated by the detection of sulfide, the sample should
also be analyzed for thiocyanate. The micrograms thiocyanate found is mathematically converted into equivalent micrograms cyanide and used in the final exposure
calculation as described in Sections 3.5 and 3.7.
Reverse diffusion
Reverse diffusion was tested by sampling a dynamically generated controlled test atmosphere containing two times the target concentration (19.3 ppm of 21.3 mg/m3)
of hydrogen cyanide at 80% relative humidity and 21 °C. Six samplers were exposed to the test atmosphere for 60 min. Sampling was discontinued and the samplers
were separated into two sets of 3 samplers each. The generation system was flushed with contaminant-free air. Sampling was resumed with a set of three samples
exposed to contaminant-free air at 80% relative humidity and 21 °C for 180 min and then all six samplers were analyzed. The data obtained are shown in Table
4.9.1.
Table 4.9.1 Reverse Diffusion of Hydrogen Cyanide from ULTRA II Passive Samplers
Loaded with Soda Lime |
|
| |
mass (µg) |
|
| set |
1 |
2 |
3 |
mean |
|
| first |
42.6 |
42.3 |
39.9 |
41.6 |
| second |
40.6 |
43.6 |
41.0 |
41.7 |
| second/first |
|
|
|
100.2 |
|
Low humidity
The effect of low humidity was tested by sampling a dynamically generated controlled test atmosphere containing two times the target concentration (19.7 ppm or
21.7 mg/m3) of hydrogen cyanide at 20% relative humidity and 21 °C. Three samplers were exposed to the test atmosphere for 240 min. All of the samplers
were immediately analyzed. Results were calculated using the average sampling rate determined in Section 4.7 converted to its equivalent sampling rate at sampling
site temperature and pressure. The results were 92.4%, 93.0%, and 100.2% of theoretical.
Low concentration
The effect of low concentration was tested by sampling a dynamically generated controlled test atmosphere containing 0.1 times the target concentration (0.99 ppm
or 1.09 mg/m3) of hydrogen cyanide at 80% relative humidity and 21 °C. Three samplers were exposed to the test atmosphere for 240 min. All of the
samples were immediately analyzed. Results were calculated using the average sampling rate determined in
Section 4.7 converted to its equivalent sampling rate
at sampling site temperature and pressure. The results were 96.2%, 97.0%, and 104.1% of theoretical.
Chemical sampling interference
The effect of a potential chemical sampling interference was tested by sampling a dynamically generated controlled test atmosphere containing one times the target
concentration (9.4 ppm or 10.4 mg/m3) of hydrogen cyanide at 80% relative humidity and 21 °C and the interference. The interference was hydrogen
sulfide at approximately its permissible exposure level (10.7 ppm or 12.9 mg/m3). Three samplers were exposed to the test atmosphere for 240 min. All
of the samples were immediately analyzed. Results were calculated using the average sampling rate determined in Section 4.7 converted to its equivalent sampling rate
at sampling site temperature and pressure. The results were 76.1%, 75.8%, and 76.9% of theoretical indicating that hydrogen sulfide is an interferent. Stability
of the sample solutions was demonstrated by analyzing the sample solutions 19 days later for hydrogen cyanide. The results were 74.2%, 77.9%, and 79.6% of theoretical.
When the samples were also analyzed for thiocyanate, as described in Section 3.5 and results calculated as shown in Section 3.7, the results were 94.6%, 97.1%, and
96.0% of theoretical.
The interference study with hydrogen sulfide was repeated with results similar to those reported above demonstrating the appropriateness of the thiocyanate correction.
Thiocyanate tests were performed, using a target concentration of 31.5 µg/sample, which is approximately the amount found on a sampler when a test atmosphere
of 10 ppm hydrogen cyanide and 11 ppm of hydrogen sulfide was sampled. These tests were performed to support the analysis of thiocyanate and are shown below.
DLOP and RQL for thiocyanate
Ten samplers were spiked with equally descending increments of analyte, such that the highest sampler loading was 0.210 µg/sample. This is the amount spiked
on sampler that would produce a peak approximately 10 times the response of a sample blank at or near the chromatographic retention time of the analyte. These spiked
samplers and the sample blank were analyzed with the recommended analytical
parameters for thiocyanate, and the data obtained used to calculate the required
parameters (standard error of estimate and the slope) for the calculation of the DLOP. Values
of 0.025 and 0.00017 were obtained for the slope and standard error of estimate respectively. The DLOP was calculated to be 0.020 µg/sample.
Table 4.9.2 Detection Limit of the Overall Procedure |
|
mass per sample (µg thiocyanate) |
area counts (uS*min) |
|
| 0.000 |
0.00000 |
| 0.028 |
0.00067 |
| 0.049 |
0.00122 |
| 0.070 |
0.00227 |
| 0.091 |
0.00239 |
| 0.112 |
0.00283 |
| 0.133 |
0.00348 |
| 0.154 |
0.00385 |
| 0.175 |
0.00452 |
| 0.196 |
0.00488 |
| 0.210 |
0.00532 |
|
 |
Figure 4.9.1.
Plot of data to determine to DLOP/RQL (y = 0.0249× + 0.000095). |
| The RQL is considered the lower limit for precise quantitative measurements. It is determined from the regression line parameters that were
obtained for the calculation of DLOP providing 75% to 125% of the analyte is recovered. The calculated value of the RQL was 0.067 µg/sample thiocyanate.
The nearest spiking amount to the calculated RQL, 0.070 µg/sample thiocyanate, had a recovery of 125%; therefore, the RQL was set to 0.091 µg/sample.
Recovery at this concentration is 102%. |
 |
Figure 4.9.2.
Chromatogram of the RQL (0.091 µg/sample) for thiocyanate. |
Precision of the analytical method for thiocyanate
The precision of the analytical method measured as the mass equivalent to the standard error of estimate determined from the linear regression of data points from
standards over a range that covers 0.1 to 2 times the target concentration. A calibration curve was constructed and shown in Section 3.5.3 from the three injections
of five standards. The standard error of estimate was 0.81 µg/sample thiocyanate.
Table 4.9.3 Instrument Calibration |
|
× target concn (µg/sample thiocyanate) |
0.1× 3.15 |
0.5× 15.75 |
1.0× 31.50 |
1.5× 47.25 |
2.0× 63.00 |
|
| area counts |
0.087 |
0.456 |
0.969 |
1.530 |
2.097 |
| (µS*min) |
0.088 |
0.464 |
0.986 |
1.550 |
2.124 |
|
0.089 |
0.470 |
0.991 |
1.562 |
2.145 |
|
Extraction efficiency
The extraction efficiency of thiocyanate was determined by liquid-spiking four vials, each containing 600 mg of soda lime, at each concentration level. These
samples were stored overnight at ambient temperature and then extracted and analyzed. The overall mean extraction efficiency over the range of 0.1 to 2 times
the target concentration was 96.9%. The extraction efficiency at the RQL was 99.3%. The presence of water had no significant effect on extraction efficiency.
The extraction efficiencies for the RQL and for the wet samplers are not included in the overall mean. Wet soda lime was prepared by exposing the samplers to humid
air (80% relative humidity at 21 °C) for 240 min. The data obtained are shown in Table 4.9.4.
Table 4.9.4 Extraction Efficiency of Thiocyanate from Soda Lime |
|
| level |
sample number |
× target concn |
µg per sample |
1 |
2 |
3 |
4 |
mean |
|
| 0.1 |
3.15 |
97.5 |
97.8 |
97.2 |
98.2 |
97.7 |
| 0.25 |
7.88 |
96.9 |
97.0 |
97.4 |
96.5 |
97.0 |
| 0.5 |
15.7 |
96.6 |
96.8 |
97.3 |
96.6 |
96.8 |
| 1.0 |
31.5 |
95.9 |
97.6 |
96.7 |
97.3 |
96.9 |
| 1.5 |
47.2 |
96.6 |
96.5 |
97.0 |
96.4 |
96.6. |
| 2.0 |
63.0 |
95.7 |
96.1 |
96.8 |
96.2 |
96.2 |
| |
| RQL |
0.091 |
99.1 |
99.5 |
98.9 |
99.7 |
99.3 |
| 1.0 (wet) |
31.5 |
94.1 |
92.6 |
94.2 |
93.6 |
93.6 |
|
Stability of extracted samples
The stability of extracted samples was examined by reanalyzing the target concentration samples 24, 48, and 72 hours after the initial analysis. After the original
analysis was performed two vials were recapped with new septa which were replaced after each reanalysis. The two remaining vials retained their puncture septa
throughout the test. All samples were allowed to stand in the autosampler tray at room temperature (20 °C). The samples were reanalyzed with freshly prepared
standards. Diff is the difference between the initial analysis and the subsequent analysis. Each septum was punctured one time for each injection. The data
obtained are shown in Table 4.9.5.
Table 4.9.5 Stability of
Extracted Samples for Thiocyanate
| punctured septa replaced |
punctured septa retained |
|
initial
(%) |
24 h
(%) |
diff
(%) |
48 h
(%) |
diff
(%) |
72 h
(%) |
diff
(%) |
|
|
initial (%) |
24 h
(%) |
diff
(%) |
48 h
(%) |
diff
(%) |
72 h
(%) |
diff
(%) |
| 95.9 |
96.8 |
0.9 |
96.0 |
0.1 |
96.3 |
0.4 |
|
|
96.7 |
97.0 |
0.3 |
96.8 |
0.1 |
96.0 |
-0.7 |
| 97.6 |
97.5 |
-0.1 |
97.1 |
-0.5 |
98.8 |
1.2 |
|
|
97.3 |
96.8 |
-0.5 |
96.3 |
-1.0 |
97.1 |
-0.2 |
| mean |
mean |
| 96.8 |
97.2 |
0.4 |
96.6 |
-0.2 |
97.6 |
0.8 |
|
|
97.0 |
96.9 |
-0.1 |
96.6 |
-0.4 |
96.6 |
-0.4 |
|
4.10 Qualitative Analysis
When necessary, the identity of an analyte peak can be confirmed by an additional analytical technique or alternate column. Several examples are briefly described:
Cyanide can be analyzed using a Dionex Ionpac® AS7 column using pulsed amperometric detection with a 75 mM NaOH/100 mM NaAc eluent (see Dionex Product Manual
for IonPac® AS7).21
Cyanide can also be analyzed by ion chromatograhy using ion-exclusion chromatography with pulsed amperometric detection (see Dionex Application Note 227).22
The Cyanide is converted to hydrogen cyanide using an eluent of methanesulfonic acid and detected with a platinum electrode. This analytical technique also has
advantages if large amounts of sulfide are present since the cyanide elutes before the sulfide. (Caution: this technique will generate hydrogen cyanide gas. Pump
column waste stream into a concentrated sodium hydroxide solution to prevent off-gassing of hydrogen cyanide.)
Another technique is analysis of hydrogen cyanide by headspace GC-NPD: Place 200µL aliquots of working standard solutions and filtered sample solutions into
separate 2-mL GC autosampler vials and seal. Using a 200-µL syringe inject 200 µL of 1:1 phosphoric acid into each vial. Heat vials in a 60 °C water
bath for 20 min; this will allow the acid to react with the cyanide releasing hydrogen cyanide into the headspace of the vial. Allow vials to cool to room temperature
before analysis. Analyze hydrogen cyanide by injecting 10 µL of the headspace, using a 25-µL gas tight syringe, and separating with an HP-PLOT/Q column.
An example of analytical conditions is given by Calafat and Stanfill.23 Hydrogen cyanide results will be low when
sulfide is present to the sample.
 |
Figure 4.11.
Diagram of apparatus used to generate test HCN atmospheres. |
4.11 Generation of test atmospheres
A dynamically generated test atmosphere, as diagramed in Figure 4.11, was set up in a walk-in hood. House air was humidified and regulated using a Miller Nelson
Model 401 Flow-Temperature-Humidity Control system. A measured flow (typically 43 or 83 µL per min) of a 1.0 M NaCN water solution was pumped through a
0.53-mm uncoated fused silica capillary tube into a single port septum inlet adaptor connected to a 500-mL 4 neck round bottom flask, using a Series D ISCO Syringe
pump with controller. Through a second neck of the flask, using an inlet adaptor, was pumped 1 L/min of nitrogen. The third neck of the flask was connected to the
inlet manifold and the fourth was internally threaded to accept a nylon bushing with o-ring that held a thermometer. In the flask was placed 20 mL of 1:1
sulfuric acid and a magnetic stir bar. The sulfuric acid solution was constantly stirred using a magnetic stir bar and maintained at a constant temperature of 30
°C (HCN has a boiling point of 25.6 °C) using a round bottom flask heater.
The 1 M NaCN solution was continuously pumped onto the inside wall of the round bottom flask, allowed to flow down into the sulfuric acid solution, where it reacted
with the acid and formed hydrogen cyanide gas. The hydrogen cyanide gas was swept out of the flask at 1 L/min with nitrogen into the inlet manifold and mixed with
100 L/min of dilution air. The gas mixture then flowed continuously into the mixing chamber (76-cm × 15-cm) and then into the sampling chamber (67-cm × 8-cm).
Samples were collected in the sampling chamber. Temperature and humidity were measured near the exit of the sampling chamber using an Omega Digital Thermo-hygrometer
model RH411. With the exception of low humidity tests OSHA generates test atmospheres at 80% relative humidity and 22 °C resulting in an absolute humidity of 15.5
mg/L H2O. The air velocity in the sampling chamber was 0.4 m/s.
The exact concentration of the 1 M NaCN solution used to generate the hydrogen cyanide test atmosphere was determined by analyzing a small aliquot using a NIST
traceable standard to calibrate the analytical instrument. The theoretical concentration of the test atmosphere was used for calculating sample recoveries.
Bubblers containing 0.1 N NaOH were initially used, at a sampling rate of 50 mL/min, to verify the theoretical concentration of the test atmosphere generated inside
the sampling chamber. A PAC III Dräger monitor, calibrated using an independent source of hydrogen cyanide, was also
occasionally used as an additional
check on theoretical test atmosphere concentrations. The monitor was connected to the sampling chamber by placing a two port calibration adaptor over the monitor
sensor cap and attaching one port via a sampling tube to the sampling chamber and the second port to a vacuum pump. The flow rate past the sensor was ~0.05 L/min.
References
1. Cyanide (OSHA Method ID-120), 1998. U.S. Department of Labor, Occupational
Safety and Health Administration Website.
2. Williamson, G. Hydrogen Cyanide [27 KB PDF, 5 pages] (NIOSH Method 6010), 1994. Centers for Disease Control and Prevention, National
Institute for Occupational Safety and Health
web site. (accessed January 2008).
3. Simmons, M. Effect of Humidity on the Collection of Particulate Cyanide & Use
of Coated Filters for the Collection of Hydrogen Cyanide and Particulate Cyanide (Study); U.S. Department of Labor, Occupational Safety & Health Administration.
Unpublished work, March 2010.
4. The Merck Index; 12th ed.; Budavari, S., Ed.; Merck & Co. Inc.: Whitehouse Station, NJ, 1996; p 1469.
5. Cheng, J.; Jandick, P.; Avdalovic, N. Pulsed amperometric detection of sulfide, cyanide, iodide, thiosulfate, bromide and thiocyanate with microfabricated disposable
silver working electrodes in ion chromatography. Anal. Chim. Acta 2005, 536, pp 267-274.
6. American Conference of Governmental Industrial Hygienists, Inc. Documentation of the Threshold Limit Values and Biological Exposure Indices
, 7th ed.; Cincinnati, OH, 2001; Vol. 2, pp. Hydrogen Cyanide - 1 through Hydrogen Cyanide - 6.
7. Hartung, R.; Cyanides and Nitriles. In Patty's Industrial Hygiene and Toxicology, 4th ed; Clayton, G.D., Claytong, F.E.,
eds.; John Wiley & Sons, Inc.: New York, 1994: Vol. II, Part D, pp 3119-3136.
8. Kirk-Othmer Encyclopedia of Chemical Technology, 4th ed.; Kroschwitz, J.I., Ex. Ed.; John Wiley & Sons, Inc.: New York, 1993;
Vol 7, pp 753-765.
9. Toxicological Profile for Cyanide,
July 2006. Centers for Disease Control and Prevention, Agency for Toxic
Substances & Disease Registry web site (accessed May 2010).
10. The Merck Index; 12th ed.; Budavari, S;, Ed,; Merck & Co. Inc.: Whitehouse Station, NJ, 1996; p 882.
11. American Conference of Governmental Industrial Hygienists, Inc. Documentation of the Threshold Limit Values and Biological Exposure Indices
, 7th ed.; Cincinnati, OH, 2001; Vol. 2, p Hydrogen Cyanide - 1.
12. Hydrogen Cyanide (Chemical Sampling Information),
2007. U.S. Department of Labor, Occupational Safety and Health Administration
Web site (accessed January 2010).
13. Eide, M.; Simmons, M.; Hendricks, W.
Validation Guidelines for Air Sampling Methods Utilizing Chromatographic Analysis [248 KB PDF*, 47 pages]; OSHA Salt
Lake Technical Center, U.S. Department of Labor: Sandy, UT, 2009 (accessed May 2010).
14. Occupational Exposure to Hazardous Chemicals in Laboratories. Code of Federal Regulations, Part 1910. 1450, Title, 29, 2003.
15. Eide, M.; Simmons, M.; Hendricks, W.
Validation Guidelines for Air Sampling Methods Utilizing Chromatographic Analysis [248 KB PDF*, 47 pages]; OSHA
Salt Lake Technical Center, U.S. Department of Labor; Sandy, UT, 2009 (accessed May 2010).
16. Cheng, J., Jandik, P.; Avdalovic, N. Pulsed amperometric detection of sulfide, cyanide, iodide, thiosulfate, bromide, and thiocyanate with microfabricated
disposable silver working electrodes in ion chromatography. Anal. Chim. Acta 2005, 536,
pp 267-274
17. Eide, M.; Simmons, M.; Hendricks, W.
Validation Guidelines for Air Sampling Methods Utilizing Chromatographic Analysis [248 KB PDF*, 47 pages]; OSHA
Salt Lake Technical Center, U.S. Department of Labor: Sandy, UT, 2009 (accessed May 2010).
18. Hendricks, W. Development of a Protocol for Laboratory Testing of Diffusive Samplers, 1996. U.S. Department of Labor, Occupational Safety and Health Administration
Web Site (accessed 2010).
19. Hendricks, W. Determined of the Sampling Rate Variation for SKC 575 Series Passive Samplers, 1998. U.S. Department of Labor, Occupational Safety and Health
Administration Web Site. (accessed January 2010).
20. Eide, M.; Simmons, M.; Hendricks, W. Validation Guidelines for Air Sampling Methods Utilizing Chromatographic Analysis [248 KB PDF *, 47 pages]; OSHA Salt
Lake Technical Center, U.S. Department of Labor: Sandy, UT, 2009 (accessed June, 2009)
21. Product Manual for IonPac® AS7 & IonPac® AG7, 2008.
Dionex. (accessed March 2010), Document No. 031299.
22. Application Note 227: Determination of Total Cyanide in Municipal Wastewater and Drinking Water Using Ion-Exclusion Chromatography with Pulsed Amperometric
Detection (ICE-PAD). Dionex. (accessed March 2010).
23. Calafat, A. M.: Stanfill, S.B. Rapid quantitation of cyanide in whole blood by automated headspace gas chromatography. J Chromatogr B
2005, 772, pp 131-137.
*These files are provided for downloading.
|