Butyl Butyrate
Isobutyl Isobutyrate
Related Information: Chemical Sampling -
Butyl Butryrate, Isobutyl
Isobutyrate
|
| Method no.: |
PV2090 |
| |
|
| Control no.: |
T-PV2090-01-9303-CH |
| |
|
| Matrix: |
Air |
| |
|
| Target Concentration: |
100 ppm (590 mg/m<sup>3</sup>) |
| |
|
| Procedure: |
Samples are collected by drawing a known
volume of air through a charcoal tube. Samples are desorbed with 1:99
dimethyl
formamide:carbon disulfide and analyzed by gas chromatography with a flame
ionization detector (GC-FID). |
| |
|
| Air volume and sampling rate studied: |
10 liters at 0.2 Lpm |
| |
|
| Status of method: |
Partially Validated Method. This method has been only
partially evaluated and is presented for information and trial use. |
| |
|
| Date: |
March, 1993
Chemist: Mary E. Eide |
| |
|
| |
|
Organic Service Branch I
OSHA Salt Lake Technical Center
Sandy, Utah |
1. General Discussion
1.1 Background
1.1.1 History of procedure
Butyl butyrate and isobutyl isobutyrate have been proposed as a replacement for the cellosolves in the manufacturing of circuit
boards. Sampling for butyl butyrate and isobutyl isobutyrate using charcoal tubes had good retention and storage efficiencies.
Desorption of both butyl butyrate and isobutyl isobutyrate with carbon disulfide showed nonconstant desorption ranging from 91.5% to
97.6% for loadings of 0.592 to 11.85 mg butyl butyrate and
91.8% to 98.9% for loadings of 0.59 to 11.8 mg isobutyl isobutyrate. The butyl butyrate and isobutyl isobutyrate desorption efficiency
using 1:99 dimethyl formamide:carbon disulfide was constant and averaged 101% for butyl butyrate and 100% for isobutyl
isobutyrate.
1.1.2 Potential workplace exposure (Ref. 5.1)
Butyl butyrate is used in the manufacture of circuit boards, and in the flavoring and fragrance industries.
1.1.3 Toxic Effects (This section is for information purposes and should not be taken as the basis for OSHA
policy.) (Ref. 5.2)
Butyl butyrate and isobutyl isobutyrate are moderate eye, mucous membrane, and respiratory irritants. An exposure of 5000 ppm for 6
hours killed two out of three rats exposed, but there was no exposure symptoms when rats were exposed to 500 ppm for 6 hours. Isobutyl
isobutyrate exposure of 5000 ppm for 6 hours killed half the rats exposed.
1.1.4 Physical properties:
Butyl butyrate (Ref. 5.1):
Compound:
Synonyms:
Molecular weight:
Density:
Freezing point:
Boiling point:
Flash point:
Color:
Molecular formula:
CAS:
IMIS:
RTECS:
DOT: |
CH3(CH2)2COO(CH2)3CH3
butyl butanoate; butanoic acid butyl ester; butyric acid butyl ester;
n-butyl n-butyrate 144.21
0.871
-91.5°C
166°C
53.5°C (128°F) (open cup)
clear liquid
C8H16O2
109-21-7
SLC1
22464; ES8120000
UN 2528 |
Isobutyl isobutyrate (Ref. 5.3):
Compound:
Synonyms:
Molecular weight:
Density:
Freezing point:
Boiling point:
Flash point:
Color:
Molecular formula:
CAS:
IMIS:
RTECS:
DOT: |
(CH3)2CHCOOCH2CH(CH3)2
2-Methylpropyl isobulyrate; 2-Methylpropyl 2-methylpropanoate;
Isobutyric acid, isobuutyl ester; 2-Methylpropyl 2-methypropionate
144.21
0.855
-81°C
148°C
37°C (99°F) (open cup)
clear liquid
C8H16O2
97-85-8
1537
44147; NQ5250000
UN 2528 |
1.2 Limit defining parameters
1.2.1 The detection limit of the analytical procedure for both compounds is 5 ng with a 1
µL injection or 5 µg/mL. This is the smallest amount
that could be detected under normal operating conditions.
1.2.2 The overall detection limit for both compounds are based on a 10 liter air
volume.
1.3 Advantages
1.3.1 The sampling procedure is convenient.
1.3.2 The analytical method is reproducible and sensitive.
1.3.3 Reanalysis of samples is possible.
1.3.4 It may be possible to analyze other compounds at the same time.
1.3.5 Interferences may be avoided by proper selection of column and GC parameters.
1.4 Disadvantages
none known
2. Sampling procedure
2.1 Apparatus
2.1.1 A calibrated personal sampling pump, the flow of which can be determined within ±5% at the recommended flow.
2.1.2 Charcoal tubes, lot 120, containing 100 mg adsorbing section with a 50 mg backup section separated by a 2 mm portion of urethane
foam, with a silanized glass wool plug before the adsorbing section and a 3 mm plug of
urethane foam at the back of the backup section. The ends are flame sealed and the glass tube containing the adsorbent is 7 cm
long, with a 6 mm O.D. and 4 mm I.D., SKC tubes or equivalent.
2.2 Sampling technique
2.2.1 Open the ends of the charcoal tube immediately before sampling.
2.2.2 Connect the charcoal tube to the sampling pump with flexible tubing.
2.2.3 Place the tubes in a vertical position to minimize channeling, with the smaller section towards the pump.
2.2.4 Air being sampled should not pass through any hose or tubing before entering the charcoal tube.
2.2.5 Seal the charcoal tube with plastic caps immediately after sampling. Seal each
sample lengthwise with OSHA Form-21 sealing tape.
2.2.6 with each batch of samples, submit at least one blank tube from the same lot used for samples. This tube should be subjected to
exactly the same handling as the samples (break ends, seal, & transport) except that no air is drawn
through it.
2.2.7 Transport the samples (and corresponding paperwork) to the lab for analysis.
2.2.8 Bulks submitted for analysis must be shipped in a separate mailing container from other samples.
2.3 Desorption efficiency
2.3.1 Six tubes were spiked at each loading of 0.52 mg (10.0 ppm), 2.96 mg (50.1 ppm), 5.92 mg (100 ppm), and 11.9 mg (201 ppm) butyl
butyrate, and 0.59 mg (10.0 ppm), 2.95 mg (50.0 ppm), 5.9 mg (100 ppm), and 11.8 mg (200 ppm) isobutyl Isobutyrate. They were allowed to
equilibrate overnight at room temperature. They were opened, each section
placed into a separate 2 mL vial, desorbed with 1 mL carbon disulfide for 30 minutes with occasional shaking, and analyzed by GC-FID. The
desorption for butyl butyrate was
non-constant ranging from 91.5% to 97.6%. (Table 2.3.1.1)
Table 2.3.1.1
Buytl Butyrate
Carbon Disulfide Desorption Efficiency
|
|
Tube #
|
|
%
|
Recovered
|
|
|
|
0.592mg
|
2.96mg
|
5.92mg
|
11.8mg
|
|
1
2
3
4
5
6
|
92.6
90.7
94.6
90.3
91.0
89.6 |
97.4
93.8
97.3
94.4
96.2
96.1 |
96.5
95.3
96.4
96.3
95.9
96.0 |
97.7
98.7
97.4
97.3
97.0
97.5 |
| average |
91.5 |
95.9 |
96.1 |
97.6 |
|
Table 2.3.1.2
Isobutyl Isobutyrate
Carbon Disulfide Desorption Efficiency
|
|
Tube #
|
|
% Recovered
|
|
|
|
0.59mg
|
2.95mg
|
5.9mg
|
11.8mg
|
|
1
2
3
4
5
6
|
91.3
91.4
91.5
91.2
94.0
91.3 |
96.1
97.3
97.8
95.3
96.9
96.5
|
97.8
97.6
97.1
96.4
97.0
97.4
|
99.5
99.1
98.7
98.9
98.9
98.0
|
| average |
91.8 |
96.7 |
97.2 |
98.9 |
|
2.3.2 Six tubes were spiked at each loading of 0.5,)2 mg (10.0 Ppm), 2.96 mg (50.1 ppm), 5.92 mg (100 ppm), and 11.9 mg (201 ppm)
butyl butyrate, and 0.19 mg (10.0 ppm), 2.95 mg (50.0 ppm), 5.9 mg (100 lipm), and 11.8 mg (200 ppm) isobutyl isobutyrate. They were
allowed to equilibrate overnight at
room temperature. They were opened, each section placed into a separate 2 mL vial, desorbed with 1
mL, 1:99 dimethyl formamide:carbon disulfide, for 30
minutes with occasional shaking, and were analyzed by GC-FID. The desorption efficiency averaged
101% for butyl butyrate and 100% for isobutyl
isobutylrate. (Tables 2.3.2.1 and 2.3.2.2)
Table 2.3.2.1 Butyl Butyrate
1:99 DMF:Carbon Disulfide Desorption Efficiency
|
|
Tube #
|
|
%
|
Recovered
|
|
|
|
0.592mg
|
2.96mg
|
5.92mg
|
11.8mg
|
1
2
3
4
5
6
|
97.4
100
100
97.5
100
101 |
101
100
101
102
102
101
|
102
102
102
102
102
101
|
102
102
102
103
102
102 |
| average |
99.3 |
101 |
102 |
102 |
| overall average: 101% |
| standard deviation: ±1.41 |
|
Table 2.3.2.2 Isobutyl Isobutyrate
1:99 DMF:Carbon Disulfide Desorption Efficiency
|
|
Tube #
|
|
%
|
Recovered
|
|
|
|
0.59mg
|
2.95mg
|
5.9mg
|
11.8mg
|
|
1
2
3
4
5
6
|
101
101
100
99.2
100
100 |
102
101
100
101
99.2
99.9
|
100
100
99.7
100
100
100
|
101
101
101
100
101
100 |
| average |
100 |
100 |
99.9 |
101 |
| overall average: 100% |
| standard deviation: ±0.763 |
|
2.4 Retention efficiency
Six tubes were spiked with 5.92 mg (100 ppm) butyl butyrate and 5.9 mg (100 ppm) isobutyl isobutyrate, allowed to equilibrate overnight,
and had 10 liters humid air (90% RH) pulled through them. They were opened, desorbed, and analyzed by GC-FID. The retention efficiency
averaged 101% for butyl butyrate and 99.5% isobutyl isobutyrate. There was no butyl butyrate
or isobutyl isobutyrate found on the backup portions of the tubes. (Tables
2.4.1 and 2.4.2)
Table 2.4.1
Butyl Butyrate Retention Efficiency
|
| Tube # |
% Recovered
'A' |
% Recovered
'B' |
Total |
|
1
2
3
4
5
6
|
101
101
101
102
101
101 |
0.0
0.0
0.0
0.0
0.0
0.0 |
98.5
99.0
99.5
99.8
99.7
101 |
|
|
|
average 99.6 |
|
Table 2.4.2
Isobutyl Isobutyrate Retention Efficiency
|
| Tube # |
% Recovered
'A' |
% Recovered
'B' |
Total |
|
1
2
3
4
5
6
|
98.5
99.0
99.5
99.8
99.7
101 |
0.0
0.0
0.0
0.0
0.0
0.0 |
98.5
99.0
99.5
99.8
99.7
101 |
|
|
|
average 99.6 |
|
2.5 Storage
Tubes were spiked with 5.92 mg (100 ppm) butyl butyrate (BUBU) and 5.9 mg (100 ppm) isobutyl isobutyrate (IBUIBU), and stored at
room temperature until opened and analyzed. The recoveries averaged 100% for the 14 days stored for both
compounds. (Table 2.5)
Table 2.5
Storage Study
|
| Day |
% Recovered BUBU |
% Recovered IBUIBU |
|
7
7
7
14
14
14 |
99.2
101
99.5
101
100
101
|
100
101
101
101
100
99.1 |
| average |
100 |
100 |
|
2.6 Precision
The precision was calculated using the area counts from six injections of each standard at concentrations of 0.592, 2.96, 5.92, and 11.8
mg/mL butyl butyrate
acid
0.59, 2.95, 5.9, and 11.8 mg/mL isobutyl isobutyrate in the desorbing solution. The pooled coefficient of variation for butyl butyrate was
0.000499 and for isobutyl isobutyrate was 0.00473. (Tables
2.6.1 and 2.6.2)
Table 2.6.1 Butyl Butyrate Precision Study
|
| Injection Number |
0.592mg/mL |
2.96mg/mL |
5.92mg/mL |
11.8mg/mL |
|
1
2
3
4
5
6 |
197949
197906
197924
197635
197791
197627 |
971557
971669
971238
970749
970969
970573 |
1970639
1971073
1972333
1971981
1970218
1970264 |
3690291
3961489
3962383
3961327
3961751
3960444
|
Average
|
197805 |
971126 |
1971085 |
3961280 |
| Standard Deviation |
±146 |
±440 |
±893 |
±795 |
| CV |
0.000738 |
0.000453 |
0.000453 |
0.000201 |
Pooled CV |
0.000499 |
|
|
|
|
Table 2.6.2
Isobutyl Isobutyrate Precision Study
|
| Injection Number |
0.59mg/mL |
2.95mg/mL |
5.9mg/mL |
11.8mg/mL |
|
1
2
3
4
5
6 |
211436
213291
211528
214111
212498
211493 |
987164
999762
987648
998123
996116
1005349 |
1938231
1934835
1931985
1937375
1923764
1937680 |
3933722
3922986
3920749
3914866
3914166
3929574 |
Average
|
212393 |
995694 |
1933979 |
3922677 |
| Standard Deviation |
±1117 |
±7118 |
±5292 |
±6952 |
| CV |
0.000526 |
0.00715 |
0.00274 |
0.00177 |
Pooled CV |
0.00473 |
|
|
|
|
where:
A(1), A(2), A(3), A(4) = # of Injections at each level
CVl, CV2, CV3, CV4 = Coefficients at each level
2.7 Air volume and sampling rate studied
2.7.1 The air volume studied was 10 liters.
2.7.2 The sampling rate studied was 0.2 liters per minute.
2.8 Interferences
Suspected interferences should be listed on sample data sheets.
2.9 Safety precautions
2.9.1 Sampling equipment should be placed on an employee in a manner that does not interfere with work performance or safety.
2.9.2 Safety glasses should be worn at all times In designated areas.
2.9.3 Follow all safety practices that apply to the workplace being sampled.
3. Analytical method
3.1 Apparatus
3.1.1 Gas chromatograph equipped with a flame ionization detector. A HP5890 gas chromatograph was used in this study.
3.1.2 GC column capable of separating the analyte and an internal standard from any interferences The column used in this study
was a 15 meter
DB-WAX capillary column 0.32 mm I.D. with a O.25 µm
film thickness. Other columns that can be used are a longer DB-WAX column, such as a 60 meter
column, or a 60 meter DB-1 capillary column 0.32 mm I.D. with a 1.0 µm film thickness.
3.1.3 An electronic integrator or some other suitable method of measuring peak areas.
3.1.4 Two milliliter vials with Teflon-lined caps.
3.1.5. A 10 µL syringe or other convenient size for sample injection.
3.1.6. Pipets for dispensing the desorbing solution. The Glenco 1 mL dispenser was used in this method.
3.1.7 Volumetric flasks - 5 mL and other convenient sizes for preparing standards.
3.2 Reagents
3.2.1 Purified GC grade nitrogen, hydrogen, and air.
3.2.2 Butyl butyrate, Reagent grade
3.2.3 Isobutyl isobutyrate, Reagent grade
3.2.4 Dimethyl formamide, Reagent grade
3.2.5 Carbon disulfide, Reagent grade
3.2.6 n-Hexylbenzene Reagent grade
3.2.7 The desorbing solution is prepared by adding 250 mL n-hexylbenzene to 1 liter of a 1:99 dimethyl formamide:carbon
disulfide solution to obtain a concentration of 0.25
µL/mL n-hexylbenzene in the solution. The n-hexylbenzene is used as the internal standard.
3.3 Sample preparation
3.3.1 Sample tubes are opened and the front and back section of each tube are placed in
separate 2 mL vials.
3.3.2 Each section is desorbed with 1 mL of the desorbing solution.
3.3.3 The vials are sealed immediately and allowed to desorb for 30 minutes with occasional shaking.
3.4 Standard preparation
3.4.1 At least two separate standards are prepared by diluting a known quantity of butyl butyrate and isobutyl isobutyrate with the
desorbing
solution.
3.4.2 A third analytical standard should be prepared at a higher concentration to check the
linearity of the detector. For this study two standard at 1 µL/mL (0.871 mg/mL butyl butyrate and
0.855 mg/mL isobutyrate), one standard at 4 µL/mL (3.48 mg/mL butyl butyrate and 3.42 mg/mL isobutyl isobutyrate), one standard at
8
µL/mL (6.97 mg/mL butyl butyrate and 6.84 mg/mL isobutyl isobutyrate), and one standard at 14
µL/mL (12.2 mg/mL butyl butyrate and 11.97 mg/mL isobutyl isobutyrate) were used.
3.5 Analysis
3.5.1 Gas chromatograph conditions.
|
Flow rates (mL/min.)
|
Temperature (°C)
|
Nitrogen (make-up): 30
Hydrogen (carrier): 2
Hydrogen (detector): 30
Air: 350 |
Injector: 181
Detector: 221
Column: 60
10°/min to 130 |
Injection size: 1µL
Chromatogram:see Figure 1
3.5.2 Gas chromatograph conditions for 60 meter DB-1 capillary column.
|
Flow rates (mL/min.)
|
Temperature (°C)
|
Nitrogen (make-up): 30
Hydrogen (carrier): 2
Hydrogen (detector): 60
Air: 420 |
Injector: 180
Detector: 220
Column: 50
10°/min to 180 |
Injection size: 1 µL
Chromatogram: see Figure 2
3.5.3 Peak areas are measured by an integrator or other suitable means.
3.6 Interferences (analytical)
3.6.1 Any compound having the general retention time of the analyte or the internal standard used is an interference. Possible
interferences should be listed on the sample data sheet. GC parameters should be adjusted if necessary so these interferences will pose
no problems.
3.6.2 Retention time data on a single column is rot considered proof of chemical identity.
Samples over the target concentration should be confirmed by GC/Mass Spec or other suitable means.
3.7 Calculations
3.7.1 The instrument is calibrated with a standard of 0.871 mg/mL (1 µL/mL) butyl butyrate and 0.855 mg/mL
(1 µL/mL) isobutyl isobutyrate in the desorbing solution. The linearity of the calibration is checked with a standard of 3.48 mg/mL (4
µL/mL) butyl butyrate and 3.42 mg/mL(4 µL/mL) isobutyl isobutyrate in the
desorbing solution.
3.7.2 If the calibration is non-linear, two more standards must be analyzed so a
calibration curve
can be plotted and sample values obtained.
3.7.3 To calculate the concentration of analyte in the air sample the following formulas are
used:
3.7.4 The above equations can be consolidated to form the following formula. To calculate the ppm of analyte in the sample based on a 10 liter
sample:
µg/mL
24.45
MW
DV
10 L
DE |
=
=
=
=
=
= |
concentration of analyte in sample or standard
Molar volume (liters/mole) at 25°C and 760
mm Hg.
Molecular weight (g/mole)
Desorption volume
10 liter air sample
Desorption efficiency |
3.7.5 This calculation is done for each section of the sampling tube and the results added together.
3.8 Safety precautions
3.8.1 All handling of solvents should be done in a hood.
3.8.2 Avoid skin contact with all chemicals.
3.8.3 Wear safety glasses, gloves and a lab coat at all
times.
4. Recommendations for further study
Collection studies need to be performed.

Figure 1. An analytical standard containing 871 µg/mL butyl butyrate and
855
µg/mL isobutyl isobutyrate in 1:99 dimethylformamide:carbon disulfide with
0.25 µL/mL n-hexylbenzene internal standard, analyzed on a 15 meter DB-WAX
capillary column. |

Figure 2. An analytical standard containing 871 µg/mL butyl butyrate and
855 µg/mL isobutyl isobutyrate in 1:99 dimethyl formamide:carbon disulfide
with 0.25 µL/mL n-hexylbenzene internal standard, analyzed on a 60 meter
DB-1 capillary column. |
5. References
5.1 Windholz, M., "The Merck Index", Eleventh Edition, Merck Co., Rahway N.J., 1989, p. 239.
5.2 Clayton, G.D.t Clayton, F.E., "Patty's Industrial Hygiene and Toxicology", Third Edition, Volume 2A, John Wiley & Sons,
New York N.Y., 1981, p. 2286.
5.3 Sax, N., Lewis, R., "Hawley's Condensed Chemical Dictionary", Eleventh Edition, Van Nostrand Reinhcld Co., New York, 1987,
p. 654.
|
| |
|