Platinum in workplace atmospheres
|
|
| Method No.: |
ID-130-SG |
|
| Control no.: |
T-ID130SG-PV-01-8503-M |
|
| Date: |
March 1985 |
|
| Sampling Procedure: |
Collection medium - collected on 0.8 µm cellulose membrane filter
Sampling rate - 1.5-2.0 liters per minute
Recommended air volume - 100 liters |
|
| Analytical Summary: |
The air filter wipe sample or bulk aliquot is extracted with H20, acidified with
HNO3, and then analyzed on an HGA graphite furnace atomic absorption spectrophotometer. |
|
Detection Limit:
|
0.01 µg/mL |
Precision:
|
| Status of Method: |
|
- Introduction
1.1 Scope.
This method describes the collection and analysis of airborne soluble platinum salts. It is applicable for time-weighted average exposure evaluations. The analysis is based on the technique of atomic absorption spectrophotometry using an HGA graphite furnace accessory.
1.2 History.
Flame atomic; absorption analysis was previously used to determine Platinum concentrations. But because the PEL value for soluble platinum salts is so low, it is necessary to use flameless (heated graphite furnace) atomic absorption techniques to obtain the low detection limit necessary.
1.3 Industrial Uses
1.3.1 In electrical, electronic, and telecommunication equipment, pure
platinum and-platinum-rich alloys are used as contacts, and in dies for forming insulators for electric lamp bases.
1.3.2 in the chemical industry the pure metal is used for the anodes in the manufacture of persulphuric acid, in the manufacture of
platinum catalysts, such as hexachloroplatinic acid, for the production of nitric and sulphuric acid, organic and vitamin products and high octane petrol.
1.3.3 In the metallurgical, glass and ceramic industries, platinum and platinum-rhodium alloys are specially suitable for the windings of
high-temperature electric furnaces.
1.3.4 In the aircraft industry, a platinum-tungsten alloy is used for sparking-plug electrodes.
1.3.5 In the manufacture of fiberglass, Platinum alloys, by virtue of their high resistance to corrosion by molten glass, are used in the equipment providing the small orifices through which the molten glass is forced.
1.3.6 In the manufacture of laboratory apparatus.
1.3.7 In the Jewelry industry, usually hardened by copper or iridium.
1.3.8 In electroplating - solutions of complex platinum salts.
1.3.9 In photography (paper sensitized with a mixture of potassium chloroplatinate and ferric oxalate); in X-ray fluorescent
screen (barium platocyanide.)
1.4 Physical and Chemical Properties
The physical and chemical properties of platinum metal are given in Table
I.
2. Range and Detection Limit
2.1 The lower analytical limit for platinum in routine analysis is
2.2 The detection limit for platinum is 0.01 µq/mL.
3. Precision and Accuracy
4. Interferences
There are no noticeable interferences in the analysis of platinum by flameless
AA.
5. Sampling Procedure
Samples collected should be full-shift (at least 7 hrs.) samples. The sample is
collected on a cellulose membrane filter (0.9 µm, 37 mm diameter) at a flow rate of 2 L/min. The minimum air volume is 250 L, maximum 960 L.
The sample cassettes are plugged, sealed with OSHA Form 21's, labeled, and sent to the laboratory for analysis as soon as possible.
6. Analytical Procedure
6.1 Apparatus
6.1.1 Sample collection
Personal sampling pump, 0.8 µm, 37-mm diameter cellulose membrane filters, 2- or 3-piece filter cassettes.
6.1.2 Sample Analysis
Atomic absorption spectrophotometer
HGA graphite furnace
Electrodeless Discharge lamp (with power supply) or hollow cathode lamp for Pt
Eppendorf pipets and tips for sampling, or autosampler
Laboratory glassware
6.2 Reagents
All reagents used should be ACS analyzed reagent grade or better.
6.2.1 HNO3 (concentrated)
6.2.2 Spex 1000 ppm stock Pt solution (or equivalent certified aqueous stock standard)
or
1000 ppm Pt stock solution, prepared as follows:
Dissolve 0.1000 a of platinum metal in a minimum volume of aqua regia and evaporate just to dryness. Add 5 mL of
HCl and 0.1 g of NaCl and again evaporate just to dryness. Dissolve residue in 20 mL of (1:1)
HCl and dilute to 100 mL with deionized water.
6.3 Safety Precautions
Before using any instrument, the operator should read the instruction manual for safe operation of the equipment.
Since metallic elements and other toxic substances are vaporized and vented into the
atmosphere during graphite furnace operation, it is imperative that a fume hood be used. Always turn on the wall switch and ensure that the exhaust is operating.
Electrical power is provided to the furnace through two flexible copper power cables. Since this is a very high current for
heating the graphite tube to incandescence, avoid coming into contact with the cables, or any spills near the instrument. Always unplug the line power before working on the instrument, and be sure only qualified
personnel do so.
OSHA-approved safety glasses should be worn at all times while using atomic absorption instruments. Improper cooling or seating of the graphite tube could cause it to explode on atomization. Never look directly at the furnace during atomization; even in normal operation the intense light/ is harmful to the eye. Also, never look at the glow of an EDL; always use a mirror and view it indirectly.
For safe operation with respect to harming equipment components, certain precautions should be taken. Do not exceed
2750° for an atomization temperature; to do so would harm the tube. Keep the atomization time below 15 seconds. Of course, be sure cooling water is circulating around the furnace before heating it. Never operate an EDL below its recommended
wattage. Be sure the air is circulating around the D 2 components (purge
2 air) before lighting the D2 lamp. The contact rings in the furnace
should be clean. But most important, read the instruction manual for the equipment.
6.4 Standards Preparation
6.4.1 Standards are prepared to match the matrix of the samples, which are 5%
v/v HNO3.
6.4.2 The stock solutions are made by serial dilution of the 1000 ppm Ft standard with deionized water as shown in
Table II.
6.4.3 The working standards are prepared from these stock solutions as indicated in
Table III.
6.5 Sample Preparation
6.5.1 All glassware must be rinsed with 1:1 HNO3 and deionized water prior to use.
6.5.2
The soluble platinum salts are extracted from the filter by placing the filter in a clean 125 ml Phillips
beaker, adding ~5 mL deionized water, and sonicating for 15 minutes in an ultrasonic bath.
The sample is then filtered using an AA filter and a Millipore filtration apparatus (the transfer from the beaker to the
filtration device must be quantitative). Transfer the filtrate to a clean volumetric flask, add
enough HNO3 to result in an acid concentration of 5% v/v, and dilute to volume. To determine the appropriate dilution volume, use the following equation:


In most cases, a final volume of 10 mL gives the required sensitivity for analysis of Pt.
6.6 Analysis
6.6.1 Working standards are prepared as previously indicated.
6.6.2 The analysis is performed using the graphite furnace AA technique. The instrumental parameters are as follows:
Atomic absorption unit:
Pt wavelength = 266 nm
UV mode
Slit setting = 4
Absorbance function
Repeat mode
D2 background corrector ON
Chart recorder:
SmV ranqe (may he varied)
20 mm/min speed
SERVO function
Furnace parameters:
|
Step |
Ramp time |
Hold time |
Temp. |
Internal
flow*
|
Inj. Vol. |
|
|
Dry |
50 sec |
40 sec |
100°C |
150 |
40 mL |
|
Char |
50 sec |
30 sec |
1200°C |
150 |
|
|
Atomize |
0 sec |
9 sec |
2800°C |
20 |
|
*Note: The internal flow above is for the HGA500 furnace
6.6.3 The 1.0 ppm standard should give a near full-scale deflection on the chart recording. If this is not the case, adjust the internal flow or chart parameters accordingly. The entire series of standards should be run at the beginning and at the end of the analysis. A standard should also be run every fourth or fifth sample in the sample range. A set of quality control samples should be run with the samples.
6.7 Calculations
A linear regression of standard ppm vs. standard absorbance is performed using
the OSHA Automatic Colorimetric program. The sample results are calculated based on sample absorbance values. The following
equation is used:

*blank corrected
Table II
|
|
Stock soln. |
Soln. used |
mL used |
final vol (mL) |
|
|
100 ppm
10 ppm
5 ppm
2 ppm
1 ppm
0.1 ppm |
1000 ppm stock
100 ppm stock
100 ppm stock
100 ppm stock
10 ppm stock
10 ppm stock |
10
10
10
5
10
1 |
100
100
100
100
100
100 |
|
Table III
|
|
Standard |
Stock soln. used |
mL stock |
mL final vol. |
|
|
0.01 ppm
0.02 ppm
0.05 ppm
0.1 ppm
0.2 ppm
0.5 ppm
1.0 ppm |
0.1 ppm stock
0.1 ppm stock
1.0 ppm stock
1.0 ppm stock
2.0 ppm stock
5.0 ppm stock
10.0 ppm stock |
5.0
10
2.5
5
5
5
5 |
50
50
50
50
50
50
50 |
|
|