| Method No.: |
PV2120 |
| |
| Control No.: |
T-PV2120-01-0305-ACT |
| |
| Matrix: |
Air |
| |
| Procedure: |
A sample is collected by drawing air through an orifice into an evacuated fused
silica-lined stainless steel canister. The canisters are analyzed in the laboratory, where they are first
pressurized with nitrogen. Aliquots of the air sample are withdrawn, cryofocused, and analyzed by gas
chromatography/mass spectrometry to determine the concentrations of compounds collected. |
| |
| Recommended sampling volume and sampling time: |
The canister volume is approximately 400 mL. Short-term sampling
orifices allow the canister to be filled in less than one minute. Long-term area samples or personal air samples may
be collected over periods up to 8 hours or longer. |
| |
| Detection Limits: |
Detection and reliable quantitation limits will vary with analyte
response factor. Detection limits of low ppb levels are possible for most common analytes. |
| |
| Status of method: |
Partially Validated. This method has been subjected to the
established evaluation procedures of the SLTC Methods Development Team. |
| |
| Date: |
May, 2003 |
| |
| Chemist: |
Patrick Hearty
|
Applied IH Chemistry Team
Program Support Division
OSHA Salt Lake Technical Center
Salt Lake City, Utah 84115-1802
|
1. General Discussion
1.1 Background and History
Evacuated stainless steel canisters with electro-polished inner surfaces, called SUMMA canisters, are widely used in
EPA applications when sampling for volatile organic compounds (VOC's) in the environment (1).
Canisters have been evaluated for use with a range of volatile organics, including aliphatic and aromatic
hydrocarbons, and chlorinated compounds (2). This technique has also been applied to a variety of
practical applications, such as indoor air quality problems (3), and other situations involving
low levels of volatile contaminants, including the docking of the Russian Priroda module with Space Station Mir (4).
The evacuated canisters offer a number of advantages, including elimination of the need for a sampling pump,
avoidance of questions concerning sorbent tube collection and recovery, and the ability to make replicate injections
and dilutions during analysis.
The development of a process which coats stainless steel with fused silica has led to important advances in chemical
sampling and analysis. This material offers the structural strength and impermeability of steel combined with the
inertness of fused silica. Entech, Inc., Simi Valley, CA, has combined the fused silica coating with the polished
canister technology in a smaller MiniCan, 400-mL capacity, for use as a personal sampler in industrial hygiene. This
new technique is leading to improved analytical methods for a variety of reactive and labile compounds of interest at
very low levels (5, 6).
Short-term area samples are collected by attaching a sampling orifice to the inlet of the MiniCan. Sampling begins
immediately, and is completed when the pressure inside the canister is equal to the atmospheric pressure on the
outside, or when the sampling orifice is detached from the canister. For personal air samples, a MiniCan is mounted
in a holster attached to a belt fastened around the waist of the worker. A sampling orifice with regulator is
attached to the inlet of the canister, and a length of inert tubing leading from the breathing zone of the worker is
connected to the inlet of the orifice. In the laboratory, the canister is pressurized with nitrogen, and the contents
are analyzed by gas chromatography/mass spectrometry.
The data presented in this method were produced during an evaluation study conducted at OSHA's Salt Lake Technical
Center (7).
1.2 Limit Defining Parameters.
1.2.1 Detection Limits
Standards of n-hexane (5.3 ppb), tetrachloroethylene (5.2 ppb),
toluene (5.3 ppb), and p-xylene (5.3 ppb) were analyzed for determination of detection
limits. Lower limits of detection are estimated to be: 0.2 ppb for n-hexane, 0.4 ppb
for tetrachloroethylene, 1 ppb for toluene, and 4 ppb for p-xylene. (Section 4.1)
1.2.2 Minimum Injection Volume
Using canisters spiked with standards of 50 ppb of the four analytes listed above, injection
volumes from 5 mL to 200 mL were analyzed. Results indicate that a 10 mL injection is the smallest volume which
provides acceptable precision. (Section 4.2)
1.2.3 Storage Stability
Canisters spiked with each of the four test compounds at 25 ppb were stored at room
temperature for up to 14 days. The average of four replicates was as follows: n-hexane
125%; toluene 97.4%; tetrachloroethylene 89.7%; p-xylene 100%. (Section 4.3)
To test recovery at higher levels, a single test was run by spiking one canister with trichloroethane at 100 ppm.
The measured value after 5 days was 102%.
The draft NIOSH method for canister sampling of VOC's indicates 30-day sample stability for most compounds (8).
Some compounds have been reported stable for up to 4 months (2).
1.2.4 Precision
Five canister replicates were spiked with the four test analytes each at a level of 50 ppb.
The coefficients of variation for the four analytes were as follows: n-hexane 13.4%;
toluene 6.7%; tetrachloroethylene 7.3%; p-xylene 6.8%. Data were taken from Table 4.2.
Statistical analyses may be found in Tables 4.4.1 through 4.4.4.
1.3 Advantages
1.3.1 Problems with collection efficiency and analyte recovery, which may be encountered
with sorbents or filters, are avoided. Evacuated canister sampling is a whole-air sampling technique.
1.3.2 When collecting short-term samples, no sampling error is associated with this method.
1.3.3 Sampling pumps are not needed.
2. Sampling Procedure
2.1 Apparatus
2.1.1 Entech Minicans, 400-mL volume (Entech P/N 29-MC400), were used in this study.
Sampling canisters may be obtained from a contract laboratory, and returned to the contract lab after samples are
collected. Canisters obtained from the contract lab will be certified clean (9).

Figure 2.1.1
Four 400 mL Minicans. |
2.1.2 Samples are collected by filling evacuated canisters through a sampling orifice.
Orifices are available which provide practically instant grab sampling (Entech P/N 39-QFS). Pressure regulated
orifices (Entech P/N CS1200E) offer sampling times as short as 2 minutes, or as long as 8 hours or more. These
sapphire orifices are said to provide superior flow stability, compared to needle valve or frit-regulated
controllers.

Figure 2.1.2.1
Short-term sampling orifice. |

Figure 2.1.2.2
Minican with pressure regulated orifice. |
2.1.3 For personal sampling, a holster and belt (Entech P/N 39-35000) can be used to
attach the canister to the waist of an employee. An inert inlet line (Entech P/N 39-36010) is used to draw air from
the employee's breathing zone.

Figure 2.1.3
Minican with pressure regulated orifice and personal sampling belt and inlet line. |
2.1.4 End caps are removed from the canisters prior to attachment of the sampling
regulators, and replaced when sampling is complete.

Figure 2.1.4
Minican inlet with end cap removed and in place. |
2.2 Reagents
None needed.
2.3 Sampling Technique
2.3.1 Choose the time-release regulator, either short- or long-term, appropriate for the
desired application.
2.3.2 Holding the sampling regulator in one hand, slide back the knurled collar with thumb and index finger.
2.3.3 Hold the canister in the other hand, with protective end cap removed, and with tip
of canister facing sampling regulator.
2.3.4 Insert the canister tip into the regulator, and release the knurled collar. No gap
should be observable between the regulator and the fitting at the end of the canister.
2.3.5 Sampling begins immediately. (Note the time of day.)
2.3.6 Bear in mind that this is a whole air sampling technique. Lack of selectivity is
inherent in this method. If for example, the person performing the sampling, or the person being sampled should be
wearing perfume or cologne, volatile components of these will also be sampled.
2.3.7 When sampling is complete, reverse above steps to disengage the canister from the
regulator. Slide back the knurled collar with thumb and index finger, and separate the canister from the regulator.
Release the knurled collar.
2.3.8 Replace the protective end cap onto the canister, and seal each canister with an OSHA Form 21.
2.3.9 Record sampling time. (Note the time of day when sampling is completed.)
2.3.10 No sample blank is necessary if the canisters were assured to be clean at the
outset of sampling. A sample collected in a control area may be included if desired.
2.4 Safety Precautions (sampling)
2.4.1 Follow all safety procedures which apply in the work area being sampled.
2.4.2 If personal sampling is being conducted, attach sampling equipment to the
employees in such a manner that it will not interfere with work performance or safety.
3. Analytical Procedure
It is possible to conduct sampling using this method even if your laboratory is not equipped with apparatus for
cleaning and analysis of canister samples. Contract laboratories will provide loan of cleaned and evacuated canisters
followed by GC/MS analysis of your samples (9).
3.1 Apparatus
3.1.1 Entech Canister System consisting of a Model 7032L 21-Position Loop Autosampler
and Model 7100 Preconcentrator, connected to a GC/mass spectrometer system. A Hewlett-Packard 5973 GC/mass
spectrometer was used in this evaluation.
3.1.2 Entech Model 4600 Dynamic Dilution System.
3.1.3 Entech Model 3100 Canister Cleaning System.
3.1.4 Summa Canisters 6-liter volume, Silonite coated.
3.1.5 A GC column capable of providing adequate separation of the analytes of interest
must be chosen. A 30-m DB-1-MS column, 0.32-mm i.d. with df 0.25 microns (J&W Scientific, catalog #1230132) was used
in this study.
3.2 Reagents
3.2.1 Standard gas mixture according to the compounds to be analyzed.
3.2.2 Liquid nitrogen.
3.2.3 Helium (ulta high purity).
3.3 Canister Cleaning
3.3.1 The Entech Model 3100 is used to clean canisters prior to sampling. Canisters are
evacuated to 13 kPa (2 psi), then filled with clean, humidified nitrogen to 172 kPa (25 psi), while heated to 80°C.
3.3.2 This process is repeated until no residual contaminants remain. Canisters are
pressurized to 207 kPa (30 psi) with nitrogen, and an aliquot is withdrawn for analysis (Section 3.6) to ascertain
cleanliness.
3.3.3 Canisters which are to be used to sample relatively high (ppm) concentrations of
analytes are usually adequately cleaned after 3 cleaning cycles. If a canister which has previously been used for
sampling of ppm-level contaminants is to be used to sample low (ppb) concentrations of analytes, more rigorous
cleaning will be required. Up to 100 cleaning cycles may be necessary. An effective and more efficient approach,
however, is to put the contaminated canisters through 3 cleaning cycles, allow the canisters to sit for a couple of
days, then repeat cleaning through three cycles, and check for cleanliness. Repeat this clean, store, and clean
sequence as many times as necessary.
3.3.4 Canisters should be evacuated to a pressure of 6.7 Pa or less prior to sampling.
It is recommended that cleaning and evacuation be conducted as near to the time of use as practical.
3.3.5 Canisters may be checked for leaks by pressurizing with clean nitrogen to 207 kPa,
rechecking pressure after 24 hours. A pressure drop greater than 14 kPa indicates a leak.
3.4 Standard Preparation
3.4.1 Using nitrogen as the diluent gas, standards of the desired analytes plus internal
standards are prepared in 6-liter Summa canisters using the Entech Model 4600 Dynamic Dilution System.
3.4.2 If electropolished Summa canisters are used, be sure to use humidified nitrogen as
diluent gas, especially if polar compounds are being analyzed. When using fused silica-coated canisters, the
requirement for humidity is not critical, since contaminant molecules have lower affinity for the silica-coated
surface than for bare stainless steel surfaces.
3.5 Sample Preparation
Prior to analysis, the pressure in each sample canister is increased to twice its original
value, using zero-grade nitrogen as the diluent gas. After equilibration, this elevated pressure allows measured
aliquots of the sample gas to be easily withdrawn for analysis.
3.6 Analysis
3.6.1 For samples expected to contain relatively high levels of contaminants (e.g., ppm
levels), the Entech Model 7032L Loop Autosampler is used to withdraw approximately 1-mL aliquots of the sample air,
diluted with nitrogen. This aliquot is cryofocused prior to introduction on to the GC/MS column.
3.6.2 For samples expected to contain levels of contaminants less than 1 ppm, the Entech
Model 7032L Loop Autosampler, is used to withdraw 10- to 100-mL aliquots of the sample air, diluted with nitrogen.
These aliquots are drawn into a sampling loop, then concentrated and cryofocused prior to introduction on to the
GC/MS column.
3.6.3 Replicate analyses or subsequent aliquots of a different size may be drawn from a
sample canister.
3.6.4 Mass spec conditions
| GC Column: |
DB-1-MS, 30m x 0.320mm i.d. |
| Initial temperature: |
35°C, hold for 5 minutes |
| Program rate: |
10°C/minute |
| Final temperature: |
280°C |
| |
|
| zone temperatures: |
GC injector: 250°C |
|
Transfer line: 280°C |
|
Source: 230°C |
|
Analyzer: 150°C |
| |
|
| Electron energy: |
70 eV |
| Scan range: |
24-250 AMU |
4. Back-up Data
4.1 Detection Limits
Figure 4.1 shows chromatograms of 20-mL and 10-mL injections of a standard of approximately 5
ppb each of the four test compounds. Based on a 100-mL injected sample volume and a two-fold dilution, the estimated
limits of detection are: n-hexane 0.2 ppb; tetrachloroethylene 0.4 ppb; toluene 1 ppb;
and p-xylene 4 ppb. Detection limits were calculated based on peak heights which are
three times the baseline noise.
Figure 4.1 Chromatograms of 20-mL and 10-mL volumes of standard 2 (5 ppb). 1 = n-hexane
(5.3 ppb), 2 = internal standard (1,1,1-trichloroethane), 3 = toluene (5.3 ppb), 4 = tetrachloroethylene (5.2 ppb),
5 = p-xylene (5.3 ppb)
4.2 Minimum Injection Volume
Table 4.2 shows the results of injections of a standard of approximately 50 ppb of each of the
four test compounds, with injection volumes varying from 5 to 200 mL. Due to lack of acceptable reproducibility of
the results from injections of 5 mL, it was concluded that 10 mL is the minimum injection volume which produces
reliable results.
Table 4.2 Instrument response of 50-ppb standard (area counts)
|
| mL |
hexane |
toluene |
tetrachloro-ethene |
xylene |
ISTD |
|
| 5 |
135213 |
77835 |
244387 |
48399 |
1174510 |
| 5 |
299161 |
191667 |
426274 |
121338 |
1111884 |
| 10 |
4459700 |
4534540 |
6104070 |
4245197 |
1318405 |
| 10 |
3690372 |
3667848 |
5004722 |
3339190 |
1112885 |
| 20 |
12708131 |
14293658 |
18675442 |
13303503 |
789727 |
| 20 |
12661318 |
13970257 |
18132314 |
12841465 |
1073221 |
| 50 |
31717802 |
45084647 |
64610183 |
41904604 |
2195409 |
| 50 |
30678491 |
43439246 |
63863835 |
40392582 |
1716193 |
| 50 |
30201947 |
43346349 |
63159344 |
40179816 |
1569277 |
| 50 |
29278081 |
40377803 |
58893541 |
37513228 |
1636387 |
| 50 |
39971450 |
38043062 |
54059640 |
35324003 |
834744 |
| 100 |
61064562 |
126363870 |
244037259 |
122562739 |
2049239 |
| 100 |
56598966 |
111798212 |
213612977 |
109534835 |
2245114 |
| 150 |
161232524 |
221738723 |
451565957 |
267346044 |
1743758 |
| 150 |
160587701 |
213603130 |
438511070 |
249931430 |
1948079 |
| 200 |
109569656 |
300828355 |
618466534 |
346332656 |
2497457 |
| 200 |
211332078 |
310746877 |
617009392 |
322998984 |
2033987 |
|
4.3 Storage Stability
For storage stability studies, canisters were spiked with standards at approximately 25 ppb of
each compound and stored at ambient temperatures. Two canisters were spiked with hexane and toluene, and two with
tetrachloroethylene and xylene. Aliquots were analyzed on day 0, and subsequently on days 3, 9, and 14. Table 4.3.1
shows the results, in percentages of theoretical amounts, for hexane and toluene. Similar data for
tetrachloroethylene and xylene are shown in Table 4.3.2. These results are represented graphically for hexane in
Figure 4.3.1.1, for toluene in Figure 4.3.1.2, for tetrachloroethylene in Figure 4.3.2.1, and for xylene in Figure
4.3.2.2.
Table 4.3.1. Storage data for n-hexane and toluene at 25 ppb
|
| time (days) |
| |
n-hexane recovery (%) |
| |
toluene recovery (%) |
|
| 0 |
| |
98.4 |
102.1 |
101.6 |
98.0 |
| |
99.7 |
104.3 |
100.4 |
95.5 |
| 3 |
| |
98.5 |
99.5 |
- |
- |
| |
99.8 |
101.3 |
- |
- |
| 9 |
| |
109.6 |
112.6 |
98.7 |
97.8 |
| |
101.1 |
102.4 |
100.9 |
102.7 |
| 14 |
| |
119.8 |
115.3 |
134.3 |
131.2 |
| |
95.3 |
91.3 |
102.7 |
100.3 |
|
Table 4.3.2. Storage data for tetrachloroethene and p-xylene at 25 ppb
|
| time (days) |
| |
tetrachloroethene recovery (%) |
| |
p-xylene recovery (%) |
|
| 0 |
| |
98.9 |
104.9 |
100.8 |
95.4 |
| |
98.9 |
105.1 |
100.4 |
95.6 |
| 3 |
| |
101.8 |
105.3 |
- |
- |
| |
100.4 |
103.2 |
- |
- |
| 9 |
| |
81.9 |
80.0 |
87.3 |
94.0 |
| |
96.3 |
99.4 |
102.7 |
102.4 |
| 14 |
| |
93.4 |
87.5 |
89.9 |
88.1 |
| |
96.5 |
93.1 |
106.1 |
104.2 |
|

Figure 4.3.1.1
Storage test for 25 ppb of n-hexane |

Figure 4.3.1.2
Storage test for 25 ppb of toluene |

Figure 4.3.2.1
Storage test for 25 ppb of tetrachloroethane |

Figure 4.3.2.2
Storage test for 25 ppb of p-xylene |
A single canister was spiked with 100 ppm trichloroethane. This was analyzed immediately, and
after storage up to 13 days. Results are tabulated in Table 4.3.3, and shown graphically in Figure 4.3.3.
Table 4.3.3 Storage test for 1,1,1-tricholoroethane at 100 ppm
|
| | |
time (days) |
| |
recovery (%) |
| |
|
| | |
0 |
| |
102.0 |
98.0 |
| |
| | |
3 |
| |
100.4 |
93.3 |
| |
| | |
5 |
| |
107.2 |
96.1 |
| |
| | |
7 |
| |
96.8 |
90.0 |
| |
| | |
12 |
| |
91.4 |
80.3 |
| |
| | |
13 |
| |
91.3 |
80.3 |
| |
| | |
17 |
| |
98.3 |
120.7 |
| |
|

Figure 4.3.3
Storage test for 100 ppm of 1,1,1-trichloroethane. |
4.4 Precision
The data in Section 4.4 were extracted from Table 4.2, and show the results of five replicate
analyses of a canister spiked with 50 ppb of each of the 4 test compounds. Table 4.4.1 shows results for
n-hexane along with statistical analysis. Table 4.4.2 shows similar results for toluene,
Table 4.4.3 for tetrachloroethylene, and Table 4.4.4 for o-xylene.
Table 4.4.1
Precision data for n-hexane, 50-mL injection, 50 ppb
|
| | |
Area Counts X 10-7 |
| |
|
| |
|
| | |
3.1718 |
| |
|
| |
| | |
3.0627 |
| |
mean = 3.2370 |
| |
| | |
3.0202 |
| |
SD = 0.4339 |
| |
| | |
2.9278 |
| |
CV = 13.40 % |
| |
| | |
3.9971 |
| |
|
| |
|
Table 4.4.2
Precision data for toluene, 50-mL injection, 50 ppb
|
| | |
Area Counts X 10-7 |
| |
|
| |
|
| | |
4.5085 |
| |
|
| |
| | |
4.3439 |
| |
mean = 4.2058 |
| |
| | |
4.3346 |
| |
SD = 0.2814 |
| |
| | |
4.0378 |
| |
CV = 6.69 % |
| |
| | |
3.8043 |
| |
|
| |
|
Table 4.4.3
Precision data for tetrachloroethylene, 50-mL injection, 50 ppb
|
| | |
Area Counts X 10-7 |
| |
|
| |
|
| | |
6.4610 |
| |
|
| |
| | |
6.3864 |
| |
mean = 6.0917 |
| |
| | |
6.3159 |
| |
SD = 0.4429 |
| |
| | |
5.8894 |
| |
CV = 7.27 % |
| |
| | |
5.4060 |
| |
|
| |
|
Table 4.4.4
Precision data for xylene, 50-mL injection, 50 ppb
|
| | |
Area Counts X 10-7 |
| |
|
| |
|
| | |
4.1905 |
| |
|
| |
| | |
4.0393 |
| |
mean = 3.9603 |
| |
| | |
4.0180 |
| |
SD = 0.2690 |
| |
| | |
3.7513 |
| |
CV = 6.79 % |
| |
| | |
3.5324 |
| |
|
| |
|
4.5 Canister cleaning.
4.5.1 Figure 4.5.1 shows a chromatogram after 3 cleaning cycles of a canister which had
contained 33 ppm trichloroethane. A 1-mL aliquot was sampled by loop injection without pre-concentration. The
canister is adequately clean for sampling and analysis at the ppm level.

Figure 4.5.1 Total ion chromatogram of a MiniCan that had contained 33 ppm of 1,1,1-trichloroethane and
then was cleaned 3 cycles. Analysis by loop injection. Arrow shows the retention time of
1,1,1-trichloroethane.
4.5.2 Figure 4.5.2 shows a chromatogram after 100 cleaning cycles of a canister which
had contained 100 ppm of trichloroethane. A 100-mL aliquot was sampled for pre-concentration prior to injection. The
canister is adequately clean for sampling and analysis at the ppb level. It is recommended that highly contaminated
canisters be cleaned for 3 cycles, allowed to sit for several days, then cleaned for 3 more cycles and tested for
cleanliness. Repeat this sequence until acceptable cleanliness is achieved.

Figure 4.5.2 Total ion chromatogram of a MiniCan that had previously contained 100 ppm of 1,1,1-Trichloroethane and
had been cleaned 100 cycles. 1,1,1-trichloroethane elutes at 11.4 min.
- Methods TO14A and TO15, Compendium of Methods for the Determination of Toxic Compounds in Air, Second Edition.
Center for Environmental Research Information - Office of Research and Development, U.S. Environmental Protection
Agency, 1999.
- Wai-mei Sin, D., Wong, Y-C., Sham, W-C., and Wang, D., Development of an Analytical Technique and Stability
Evaluation of 143 C3-C12 Volatile Organic Compounds in SUMMA Canisters by Gas Chromatography-Mass Spectrometry, The
Analyst, 126, pp 310-21, 2001.
- Farant, J-P., and Simon, P. An Approach to the Resolution of Sporadic Gaseous Indoor Air Pollution Events,
Proceedings of the 2nd European Conference on Energy Performance and Indoor Climate in Buildings and the 3rd
International Conference on Indoor Air Quality, Ventilation and Energy Conservation in Buildings, pp. 952-56, 1998.
- Cole, H., Ward, S., Manuel, S., Rather, D., Simon, P., Farant, J-P., Krasnec, J., Gouzenberg, A., Moukhamedieva,
L., and Mikos, K. The Application of Grab and Time Integrated Sampling to the Characterization of Trace Contaminants
Contributed by the Docking and Integration of the Priroda Module to Space Station Mir, Proceedings of the 28th
International Conference on Environmental Systems, 1998.
- Formaldehyde and VOC's in Indoor Air Quality Determinations by GC/MS, Entech Instruments Applications Note 101,
Entech Instruments, Inc., Simi Valley, CA.
- Ochiai, N., Takino, M., Daishima, S., and Cardin, D. Analysis of Volatile Sulfur Compounds in Breath by Gas
Chromatography-Mass Spectrometry using a Three-Stage Cryogenic Trapping Preconcentration System, Journal of
Chromatography B, 762, pp. 67-75, 2001.
- Chan, Y., and Hearty, P. Evaluation of the Entech Canister System, 2000. Unpublished.
- Draft method, "Volatile Organic Compounds in Air, GC/MS." Steven P. Sanders, 3M Corporation, 2001.
- Certified laboratories which will analyze Entech Canisters on a fee-for-analysis basis include Galson
Laboratories, East Syracuse, NY, and Aerotech Laboratories, Inc., Phoenix, AZ.
|