|
ENFLURANE AND HALOTHANE
| Method no.: |
29 |
|
|
| Matrix: |
Air |
|
|
| Target concentration: |
Halothane - 1 ppm (8.1 mg/m3)
Enflurane - 1 ppm (7.6 mg/m3) |
|
|
| Procedure: |
Two standard size charcoal tubes are connected in series
for each sample. Samples are desorbed with carbon disulfide (CS2)
and analyzed by gas chromatography (GC) using a flame ionization detector
(FID). |
|
|
Recommended air volume and sampling rate: |
10 L at 0.1 L/min |
|
|
| Reliable quantitation limit: |
Halothane - 0.023 ppm (0.19 mg/m3)
Enflurane - 0.040 ppm (0.30 mg/m3) |
|
|
Standard error of estimate at the target concentration: |
|
(Figure 4.7.4.) (Figure 4.7.2.) |
Halothane - 6.43% Enflurane - 8.01% |
|
|
| Status of method: |
A sampling and analytical method which has been
subjected to the established evaluation procedures of the Organic
Methods Evaluation Branch. |
|
|
|
|
|
|
| Date: May 1981 |
Chemist: Mike Shulsky |
Organic Methods Evaluation Branch
OSHA Analytical Laboratory
Salt Lake City, Utah
1. General Discussion
1.1. Background
1.1.1. History
Halothane has been collected in glass sampling bulbs with
subsequent GC analysis using an electron capture detector.
(Ref. 5.1.) Charcoal passive dosimeters have been used
for collection of both enflurane and halothane with GC/FID
analysis (Ref. 5.2.).
Charcoal tubes were used in this evaluation because of
their widespread use for other organics and their efficiency
for collecting both analytes. Also, with charcoal
high desorptions were obtained for both analytes.
The analytical method is sensitive and analytical parameters
can be adjusted to eliminate interferences.
NIOSH recommends a 1 h ceiling value of 2 ppm for both
halothane and enflurane (Ref. 5.4.). Since ceiling values
in the Federal Register are usually twice the 8-h TWA PEL,
a target concentration of 1 ppm was chosen.
1.1.2. Toxic effects (This section is for information only and
should not be taken as the basis of OSHA policy).
There are conflicting reports in the literature dealing
with the effects of exposure of humans to low levels of
anesthetics. One study indicates that chronic low-level
exposure of female operating room personnel to anesthetics
in general can cause spontaneous abortion (Ref. 5.3.).
Animal studies have shown that chronic halothane exposures
of 50-150 ppm can cause liver and kidney damage (Ref.
5.4.). At sub-anesthetic levels, enflurane exposure produced
early mortality and detrimental weight effects in
mice although the same tests had no apparent effect on
rats or guinea pigs (Ref. 5.5.).
1.1.3. Workplace exposure
Enflurane and halothane are two of the three most commonly
used anesthetics. Occupational exposure may occur wherever
anesthetics are used such as operating rooms, veterinarian
hospitals, and dental offices. The number of people
potentially exposed is estimated to be 215,000 (Ref.
5.4.).
1.1.4. Physical properties (Ref. 5.4. unless otherwise stated)
| halothane |
|
| molecular weight: |
197.4 |
| boiling point: |
50°C |
| specific gravity: |
1.871 (Ref. 5.6.) |
| vapor pressure: |
243 mm Hg (20°C) |
| synonyms: |
2-bromo-2-chloro-1,1,1-trifluoroethane, fluothane |
| formula: |
CHClBrCF3 |
|
|
| enflurane |
|
| molecular weight: |
184.5 |
| boiling point: |
56°C |
| specific gravity: |
1.52 |
| vapor pressure: |
175 mm Hg (20°C) |
| synonyms: |
2-chloro-1,1,2-trifluoroethyldifluoro-methylether,
ethrane |
| formula: |
CHFClCF2OCHF2 |
1.2. Limit defining parameters
1.2.1. Detection limit of the analytical procedure
The detection limit of the analytical procedure is 6.08
ng/injection for enflurane and 3.74 ng/injection for halothane.
These are the amounts which give peaks about 5
times the baseline noise. (Section 4.1.1.)
1.2.2. Detection limit of the overall procedure
The detection limit for the overall procedure is 3.04
µg/sample (0.04 ppm or 0.3 mg/m3) for enflurane and 1.87
µg/sample (0.023 ppm or 0.19 mg/m3) for halothane. This
is the amount of analyte spiked on a charcoal tube which
allows recovery of an amount equal to the detection limit
of the analytical procedure. (Section 4.1.2.)
1.2.3. Reliable quantitation limit
The reliable quantitation limit is 3.04 µg/sample (0.04
ppm or 0.3 mg/m3) for enflurane and 1.87 µg/sample (0.023
ppm or 0.19 mg/m3) for halothane. This is the smallest
amount of analyte which can be quantitated within the
requirements of a recovery of at least 75% and a precision
of ±25% or better. (Section 4.2.)
The reliable quantitation limit and detection limits reported in
this method are based on optimization of the instrument for the
smallest possible amount of analyte. When the target concentration
is exceptionally higher than these limits, they may not be
attainable at the routine operating parameters.
1.2.4. Sensitivity
The sensitivity of the analytical procedure over the concentration
range representing 0.5 to 2 times the target
concentration is 496 area counts/(µg/mL) for enflurane and 534 area
counts/(µg/mL) for halothane. The sensitivity is determined by the
slope of the calibration curves. (Section 4.4.) Sensitivity will vary somewhat
with the particular instrument used for analysis and method of integration.
1.2.5. Recovery
The recovery of the analyte from the collection medium must be 75% or greater.
The average recovery from spiked charcoal tubes over the range of 0.5 to 2 times
the target concentration was essentially 100% for both enflurane and halothane.
(Section 4.6.)
1.2.6. Precision (analytical method)
The pooled coefficients of variation obtained from replicate
determinations of analytical standards at 0.5, 1, and
2 times the target concentration are 0.059 for enflurane
and 0.029 for halothane. (Section 4.3.1.)
1.2.7. Precision (overall procedure)
The precision at the 95% confidence level for the 15-day
storage test is ±15.6% and ±12.6% for enflurane and halothane,
respectively. (Section 4.3.2.) These values include an
additional ±5% for sampling error. The overall
procedure must provide results at the target concentration
that are ±25% or less at the 95% confidence level.
1.3. Advantages
1.3.1. The sampling apparatus is portable and easy to use.
1.3.2. No loss of analytes is observed after 2 weeks of storage.
1.3.3. The samples can be reanalyzed several times.
1.3.4. The analytical method is easy and convenient.
1.4. Disadvantages
1.4.1. The method has not been field tested.
1.4.2. The solvent used for desorption is toxic and flammable so
it must be handled with care.
2. Sampling Procedure
2.1. Apparatus
2.1.1. A personal sampling pump which can be calibrated within
±5% of the 0.1 L/min flow rate.
2.1.2. Coconut shell charcoal tubes (glass tubes 7-cm
length, 6-mm o.d., 4-mm i.d.) are used. They contain
100-mg and 50-mg portions of charcoal and the tube
is heat sealed at both ends. For this evaluation,
tubes from SKC lot 120 charcoal were used.
2.2. Reagents
None required
2.3. Technique
2.3.1. Immediately before sampling, break open the ends of
the charcoal tubes and connect two charcoal tubes in
series per sample. The small, red, plastic SKC cap
with the closed end removed is convenient for this.
Position the 50-mg portion of the tubes nearest the
sampling pump. All tubes should be from the same lot
of charcoal.
2.3.2. Properly label each pair of tubes to distinguish the
front and backup tubes.
2.3.3. The tubes are connected to the sampling pump with a
piece of flexible tubing. The tube nearest the pump
should be labeled as the backup portion of the
sample.
2.3.4. The tubes should be placed vertically to avoid
channeling through the charcoal.
2.3.5. The exposed, broken end of the first charcoal tube
should be enclosed in a tube holder to avoid injury
to the worker.
2.3.6. Air being sampled should not pass through any tubing
before entering the first charcoal tube.
2.3.7. After sampling, separate the pairs of tubes and seal
each tube with plastic caps.
2.3.8. With each set of samples, submit at least one blank
tube. The blank should be treated in the same manner
as the samples (break ends, seal, transport) except
no air is drawn through it.
2.3.9. Transport the samples and paperwork to the lab for
analysis.
2.3.10. If bulk samples are to be submitted, they must be
sent in a mailing package separate from any air
samples being submitted. This is to avoid
contamination of the air samples if the bulk should
leak or get broken in transit.
2.4. Breakthrough
Studies to determine the 5% breakthrough values were done for
enflurane and halothane simultaneously at twice the target
concentration using an entire charcoal tube ("A" and "B" portion)
with a flow of 0.1 L/min. The average breakthrough air volumes
were 18.6 L for enflurane and 16.4 L for halothane. Average
capacities of the tubes were 0.255 mg of enflurane and 0.252 mg for
halothane. (Section 4.5.)
2.5. Desorption efficiency
The desorption efficiency was determined at several loadings of
enflurane and halothane by liquid injection of CS2 solutions of
these analytes. The amount of analytes added corresponded to the
detection limit of the analytical procedure, 0.5, 1, and 2 times
the target concentration. The "A" and "B" portions of each tube
were placed in one vial and desorbed with 1 mL of CS2/internal
standard solution. The desorption efficiencies for all loadings
and both analytes were essentially 100%. (Section 4.6.) Desorption
efficiencies may vary from one lot of charcoal to another.
2.6. Recommended air volume and sampling rate
The recommended air volume is 10 L at a flow rate of 0.1 L/min or
lower.
2.7. Interferences
2.7.1. Charcoal will adsorb many organic vapors so organics being
used in significant quantities near the sampling area
should be listed as possible interferences.
2.7.2. High relative humidity and temperature will decrease the
capacity of the charcoal for halothane and enflurane. If
these conditions are encountered, it may be advisable to
decrease the total volume of air sampled. The recommended
10-L air volume was based on sampling air at 22°C and 80%
relative humidity.
2.8. Safety precautions
2.8.1. Wear eye protection when breaking the ends of the charcoal
tubes.
2.8.2. Place the sampling apparatus on the employee so as not to
interfere with the work being done.
2.8.3. Adequate precautions should be taken at all times to prevent
injury from the broken ends of the charcoal tubes.
3. Analytical Method
3.1. Apparatus
3.1.1. Gas chromatograph (GC) equipped with a flame ionization
detector.
3.1.2. A GC column is needed which is capable of separating carbon
disulfide, enflurane, halothane, and if used, an internal
standard compound. For this evaluation, a 20 ft
stainless steel column packed with 10% Tergitol NP-35 on
80/100 Chrom P-AW DMCS was used.
3.1.3. A suitable method for peak area integration.
3.1.4. Small vials with Teflon-lined caps are needed for desorption
of charcoal. Two-milliliter vials are preferable.
3.1.5. Microliter syringes; 2-µL and 10-µL
3.1.6. Pipettes of various sizes for desorption and dilution are
needed. A solvent dispenser is convenient for sample
desorption.
3.1.7. Volumetric flasks; 5-, 10-, 25-mL
3.2. Reagents
3.2.1. Carbon disulfide; reagent grade
3.2.2. Halothane; USP grade
3.2.3. Enflurane; USP grade
3.2.4. For this evaluation, benzene was used as an internal
standard (1 µL/mL in CS2). Since it may not be feasible to
use benzene in other laboratories, several other compounds
were tried but found to be unsatisfactory either because
of interferences or considerable lengthening of analysis
time. The compounds tried were hexane, cyclohexane, methyl
cyclohexane, heptane, isooctane, octane, nonane, bromomethane,
chloroform, methyl chloroform, Freon 113, trichloroethylene,
and tetrachloroethylene. An external
standard method may be used instead of an internal standard
method.
3.2.5. GC grade nitrogen or helium, hydrogen and air
3.3. Standard preparation
Standards should be prepared at concentrations corresponding to
80.7 µg/mL for halothane and 75.5 µg/mL for enflurane. This is
representative of samples taken at the target concentration assuming
a 10-L air volume.
3.4. Sample preparation
3.4.1. Both sections of charcoal in the first tube are placed in
one vial while both sections of the second tube (backup
tube) are placed in a separate vial. One milliliter of
desorbing solvent is added to each vial.
3.4.2. The vials are immediately capped and shaken intermittently
over a 30-min period before analysis.
3.5. Analysis
3.5.1. Conditions for this evaluation
| GC temperatures |
|
| oven: |
70°C |
| injector: |
150°C |
| detector: |
150°C |
|
|
|
|
| gas flow rates |
|
| carrier (He): |
25 mL/min. |
| hydrogen: |
23 mL/min. |
| air: |
250 mL/min. |
|
|
|
|
| retention times |
|
| enflurane: |
12.61 min |
| halothane: |
13.83 min. |
| benzene: |
19.84 min |
|
|
|
|
| injection size: |
2 µL |
| detector: |
flame ionization |
| chromatogram: |
Figure 4.8. |
3.5.2. Peak areas are measured by a suitable technique.
3.5.3. A calibration method for our data system was prepared by setting
the peak areas from the analytical standards equal to the corresponding air
concentration in ppm assuming a 10-L air volume, using 1 mL for desorption.
Example:
| Halothane standard: |
86 µg/mL |
| Assume 10-L air volume |
|
1-mL desorption volume: 99% desorption efficiency; 24.46 is molar
volume of an ideal gas at 25°C and 760 mm Hg; 197.39 MW of Halothane
ppm = (86 µg/mL)(1 mL)(24.46)/(10 L)(197.39)(0.99)
ppm (Halothane) = 1.08
By calibrating the area of the analyte to its corresponding
air concentration, the areas obtained from
actual samples can be expressed in ppm assuming
a 10-L air volume. The correction for these
results to the actual air volume for a sample
is discussed in Section 3.7.1.
3.5.4. Any samples above the target concentration
must be confirmed by GC/MS or another suitable
method. Retention time on one column is not
considered proof of identity.
3.6. Interferences
3.6.1. Any compound which has the same retention time
as halothane or enflurane will be an
interference.
3.6.2. The solvent, CS2, generally will contain
impurities which may be significant at the
analytical levels.
3.7. Calculations
3.7.1. The values obtained from the calibration
method described in Section 3.5.3. will be in
ppm based on a 10-L air volume. To convert this
ppm value to a ppm value based on the actual
air volume for each sample, the actual air
volume is divided by 10 L and that quotient is
divided into the ppm value from the calibration
method.
| Example: |
1 ppm from calibration method |
|
12-L air volume |
12 L 10 L |
= 1.2 |
|
1 ppm 1.2 |
= 0.8 ppm |
Reported result = 0.8 ppm
3.7.2. If there is analyte present in the backup tube
then the same air volume correction must be
made. This corrected ppm value is added to that
value from the front tube and the total is
reported. If the corrected ppm value from the
backup tube is 25% or more of the corrected
value from the front tube, then a note should
be placed on the analysis report sheets
indicating that the particular sample should be
considered saturated.
3.8. Safety precautions
3.8.1. Work in a hood when using solvents.
3.8.2. Keep volumetrics and vials containing solvents
away from sources of high temperature such as
GC injectors and detectors.
3.8.3. Avoid skin contact with all solvents.
3.8.4. Wear safety glasses at all times.
4. Backup Data
4.1. Detection limits
4.1.1. Analytical procedure
The detection limits were determined by injecting 2 µL of
an analytical standard containing 3.04 µg/mL enflurane and
1.87 µg/mL halothane. The chromatographic peaks obtained
for the analytes were approximately 5 times the baseline
noise. Therefore, the detection limit of the analytical
procedure is 6.08 ng/injection for enflurane and 3.74
ng/injection for halothane (Figure 4.1.1.).
4.1.2. Overall procedure
The detection limit of the overall procedure is 3.04
µg/sample (0.040 ppm or 0.30 mg/m3) for enflurane and 1.87
µg/sample (0.023 ppm or 0.10 mg/m3) for halothane. These
are the amounts of respective analyte spiked on a charcoal
tube which allow recovery of amounts equal to the respective
detection limits of the analytical procedure (Table
4.6.1. and 4.6.2., Figure 4.6.1. and 4.6.2.). These limits
assume the injection size (2 µL) of the analytical
method.
4.2. Reliable quantitation limit
The reliable quantitation limit is 3.04 µg/sample (0.040 ppm or
0.3 mg/m3) for enflurane and 1.87 µg/sample
(0.023 ppm or 0.19 mg/m3) for halothane. This is the smallest amount which can be
spiked on a charcoal tube, which will give at least 75% recovery
and a precision within ±25% at the 95% confidence level. The
precision (1.96 SD) of the data listed in Tables 4.6.1 and 4.6.2. for
the above concentrations is ±3% for halothane and ±2% for enflurane.
This limit was determined assuming the injection size (2 µL)
of the analytical method.
4.3. Precision data
4.3.1. Analytical method
Precision was determined by replicate injections of analytical
standards prepared at concentrations equivalent to
0.5, 1, and 2 times the target concentration. The pooled
coefficient of variation is 0.059 for enflurane and 0.029
for halothane.
Table 4.3.1.1. Analytical Precision for Halothane
|
| × target conc. |
0.5× |
1× |
2× |
|
| area counts |
12002 |
24316 |
45953 |
|
11851 |
24070 |
47323 |
|
11921 |
24047 |
45316 |
|
11322 |
25971 |
46503 |
|
12423 |
24196 |
45369 |
|
12073 |
23356 |
45685 |
|
|
 |
11932 |
24326 |
46025 |
| SD |
359 |
873 |
770 |
| CV |
0.030 |
0.036 |
0.017 |
|
|
= 0.029 |
|
Table 4.3.1.2. Analytical Precision of Enflurane
|
| × target conc. |
0.5× |
1× |
2× |
|
| area counts |
7318 |
16791 |
33271 |
|
7377 |
19266 |
33832 |
|
7188 |
15538 |
32004 |
|
7024 |
15989 |
33310 |
|
7732 |
15776 |
33256 |
|
7464 |
15096 |
31904 |
|
|
 |
7351 |
16409 |
32929 |
| SD |
242 |
1508 |
786 |
| CV |
0.033 |
0.092 |
0.030 |
|
|
= 0.059 |
|
4.3.2. Overall procedure
The precision for the overall procedure includes ±5% for
sampling and any storage effects. This precision was
found by plotting the storage data from Table 4.7.1.,
Figures 4.7.1. - 4.7.4. The precision for enflurane is
±16.9% (Figure 4.7.2.) and ±13.5% for halothane (Figure
4.7.4.).
4.4. Sensitivity
The sensitivity is taken as the slope of the calibration curves
obtained by plotting data Tables 4.3.1.1. and 4.3.1.2. The
sensitivity for enflurane is 496 area counts/(µg/mL) (Figure 4.4.1.)
and 534 area counts/(µg/mL) for halothane (Figure 4.4.2.).
4.5. Sampler Capacity
Breakthrough is defined as the point during sampling when the
analyte concentration in the air downstream from the charcoal tube
is 5% of the concentration ahead of the tube. To determine the
breakthrough point, a vapor generation system was used.
The atmosphere in the system was 2 ppm of halothane and enflurane,
each, with approximately 80% relative humidity at 21°C. Sampling
this atmosphere was performed using a charcoal tube containing
both sections of charcoal.
Monitoring the downstream concentration of analytes was done by
using a gas sampling valve (GSV) connected to a GC. The sample in
the GSV was injected, at specified time intervals, into the GC
column in order to separate the two analytes. An FID was used for
detection.
Before a charcoal tube was put in line, a by-pass was used so the
analytes' concentrations downstream would be the same as ahead of
the tube. This allowed the monitoring of the analytes'
concentrations in the system, so the 100% value could be determined. When
the analytes' concentrations had stabilized, the air was diverted
from the by-pass to the charcoal tube at the start of a gas valve
sampling cycle; this is time zero. Since the air sampled at time
zero had not passed through the charcoal tube, it's concentration
of analytes was essentially 100%. Depending on the flow rate
through the GSV, and the GC run time, the concentration of the
analytes in the air sampled will drop to zero within the next few
GSV cycles.
Four separate breakthrough determinations were performed by
sampling the above mentioned atmosphere (2 ppm enflurane and
halothane, 80% R.H.) at a flow rate of 0.1 L/min through the charcoal
tubes. The average breakthrough air volumes are 18.5 L for
enflurane and 15.9 L for halothane. The average capacities of the
tubes were 0.255 mg of enflurane and 0.252 mg of halothane. The
data collected for one test is plotted in Figure 4.5.1. with the
(%) of halothane and enflurane in the downstream air vs. the total
time of the test.
Table 4.5. Sampler Capacity and Air Volume Sampled at 5% Breakthrough
|
| test |
enflurane |
halothane |
enflurane |
halothane |
|
mg |
mg |
L |
L |
|
| 1 |
0.270 |
0.286 |
19.8 |
17.6 |
| 2 |
0.241 |
0.248 |
17.5 |
15.8 |
| 3 |
0.226 |
0.231 |
15.4 |
15.9 |
| 4 |
0.282 |
0.241 |
21.1 |
14.4 |
|
4.6. Desorption efficiency
Desorption efficiencies were determined by injecting amounts of
enflurane and halothane in solution onto charcoal tubes. The
amounts were equivalent to 0.5, 1, and 2 times the target concentration.
Also, one set of tubes was prepared at the detection
limit of the analytical procedure. The desorption efficiencies
for both analytes over the concentration range evaluated was virtually
100% (Tables 4.6.1. and 4.6.2., Figures 4.6.1. and 4.6.2.)
Table 4.6.1.
Desorption Efficiencies for Halothane
|
| x target conc. |
0.02× |
0.5× |
1× |
2× |
| µg/sample |
1.87 |
42.1 |
84.2 |
168 |
|
| desorption |
101 |
100 |
102 |
102 |
| efficiency, % |
101 |
90 |
99 |
101 |
|
100 |
94 |
99 |
104 |
|
104 |
97 |
102 |
102 |
|
101 |
95 |
104 |
102 |
|
104 |
96 |
104 |
102 |
|
|
 |
102 |
95 |
102 |
102 |
|
|
= 100 |
|
|
Table 4.6.2.
Desorption Efficiencies for Enflurane
|
| x target conc. |
0.025× |
0.5× |
1× |
2× |
| µg/sample |
3.04 |
37.9 |
75 |
152 |
|
| desorption |
104 |
108 |
110 |
103 |
| efficiency, % |
104 |
100 |
108 |
112 |
|
102 |
104 |
108 |
106 |
|
103 |
104 |
102 |
108 |
|
101 |
110 |
110 |
103 |
|
103 |
110 |
106 |
106 |
|
|
 |
103 |
107 |
107 |
106 |
|
|
= 103 |
|
|
4.7. Storage data
Thirty-six samples were collected from a dynamically generated
atmosphere of approximately 1 ppm of both halothane and enflurane.
Relative humidity in the system was maintained at approximately
80%. Ten liters of this atmosphere were collected at 0.1 L/min.
Six of the samples were desorbed and analyzed the
same day as collected. Of the 30 remaining tubes 15
were stored at ambient temperatures and 15 were
refrigerated at approximately (5°C). Every third day,
3 samples from the ambient set and 3 from the
refrigerated set were desorbed and analyzed. The
results as % recovered are presented below and
plotted in Figures 4.7.1. - 4.7.4.
Table 4.7.1
Storage Tests for Halothane
|
| storage time |
% recovery |
| (days) |
(refrigerated) |
(ambient) |
|
| 0 |
82 |
88 |
84 |
90 |
84 |
90 |
| 3 |
95 |
93 |
101 |
93 |
94 |
97 |
| 6 |
96 |
91 |
99 |
89 |
94 |
100 |
| 10 |
98 |
99 |
102 |
94 |
99 |
100 |
| 12 |
98 |
98 |
99 |
96 |
98 |
95 |
| 15 |
100 |
99 |
98 |
94 |
97 |
99 |
|
Table 4.7.2. Storage Tests for Enflurane
|
| storage time |
% recovery |
| (days) |
(refrigerated) |
(ambient) |
|
| 0 |
105 |
107 |
103 |
105 |
101 |
103 |
| 3 |
91 |
92 |
107 |
96 |
93 |
97 |
| 6 |
99 |
94 |
110 |
94 |
108 |
104 |
| 10 |
106 |
105 |
107 |
100 |
106 |
112 |
| 12 |
98 |
99 |
101 |
97 |
101 |
96 |
| 15 |
110 |
113 |
109 |
lost |
96 |
107 |
|
Figure 4.1.1. Detection limits of the analytical procedure.
Figure 4.4.1. Calibration curve for enflurane.
Figure 4.4.2. Calibration curve for halothane.
Figure 4.5.1. Breakthrough curves.
Figure 4.6.1. Desorption efficiency for enflurane.
Figure 4.6.2. Desorption efficiency for halothane.
Figure 4.7.1. Ambient storage test for enflurane.
Figure 4.7.2. Refrigerated storage test for enflurane.
Figure 4.7.3. Ambient storage test for halothane.
Figure 4.7.4. Refrigerated storage test for halothane.
Figure 4.8. Chromatogram of halothane and enflurane at their respective target
concentrations.
5. References
5.1. Nicholson, John A.; Sada, Tsuneo; Aldrete, J. Antonio,
Anesthesia and Analgesia. 1975, 54(4), 449 - 454.
5.2. Mazur, J.F.; Podolak, G.E.; Esposito, G.G.; Rinehart, D.S.;
Glenn, R.E. Am. Ind. Hyg. Assoc. J. 1980, 41(5), 317 - 321.
5.3. Knill-Jones, R.P.; Moir, D.D.; Rodrigues, L.V.; Spence, A.A.
The Lancet. 1972, I, 1326 - 1328.
5.4. "Criteria for a Recommended Standard...Occupational Exposure
to Waste Anesthetic Gases and Vapors". NIOSH. (U.S.) 1977,
Publ. No. 77 - 140.
5.5. Stevens, Wendell C.; Eger, Edmond I.; White, Anne; Biava,
Claude G.; Gibbons, Robert D.; Shargel, Richard. Canadian
Anesthesiology Society Journal. 1977, 24(4), 479 - 490.
5.6. Stecher, Paul G. (Editor) "The Merck Index". 8th edition.
Rahway: Merck and Company, Inc.
|