OSHA IMIS Code Number: T137

Chemical Abstracts Service (CAS) Registry Number: 79-94-7

Monitoring Methods used by OSHA

Primary Laboratory Sampling/Analytical Method (SLC1):
  • sampling media: Glass Fiber Filter (37 mm)
    analytical solvent: Isooctane/Isopropanol/Methanol (85:10:5)
    maximum volume:100 Liters
    maximum flow rate: 1.0 L/min
    current analytical method: High Performance Liquid chromatography HPLC/U.
    method reference: 2 (OSHA In-House File)
    method classification: Partially Validated
Conditions: Column Alltech CN; Solvent (85/10/5) Isooctane/isopropanol/methanol; Flow 1.1 mL/min; UV det 254 & 280 nm; Ret time 6 min; DL 0.5 ug/smpl