| Standard Number: | 1910.1000 |
|
February 21, 1989
MEMORANDUM FOR: REGIONAL ADMINISTRATORS
ATTN: ARA's for Technical Support
THRU: LEO CAREY Director
Office of Field Programs
FROM: EDWARD J. BAIER Director
Directorate of Technical Support
SUBJECT: Cyclones Supplied by SKC Inc.
The cyclones shipped with the SKC medium flow air sampling pumps for
respirable dust sampling are made of metal and have been calibrated at a flow
rate of 1.9 liters per minute (LPM) which is different from the 1.7 LPM
required for the nylon cyclone.
SKC indicated that these metal cyclones were purchased from Sweden and it was designed from the collection criteria established by the British Medical Research Council (BMRC). The nylon cyclones supplied to our CSHOs are based on the collection criteria recommended by the American Conference of Governmental Industrial Hygienists (ACGIH). Their major difference is the cut-off size. The nylon cyclone has a 50% collection efficiency at 3.5 micrometers while the SKC cyclone has a 50% collection efficiency at 5.0 micrometers (see attached table for details). Since the two collection efficiency curves are quite different and are not compatible, the SKC cyclones are not to be used for collection of respirable dust. All SKC cyclones should be returned to the OSHA Cincinnati Laboratory for replacement by the nylon cyclones. If your staff is interested in respirable dust sampling, the ACGIH publication "Air Sampling Instruments" has an excellent chapter on this subject (Size-selective Health Hazard Sampling by Morton Lippmann).