|
<<< Back to Sampling and Analytical Methods |
Printing Instructions
|
For problems with accessibility in using figures, illustrations and PDF in
this method, please contact
the SLTC at (801) 233-4900. These procedures were designed and tested for
internal use by OSHA personnel.
Mention of any company name or commercial product does not constitute
endorsement by OSHA. |
Marine Project
[516
KB PDF]
|
| The Marines Project: A Laboratory Study of Diffusive Sampling/Thermal
Desorption/Mass Spectrometry Techniques for Monitoring Personal Exposure to Toxic Industrial
Chemicals |
|
| April 2002 |
Warren Hendricks |
Methods Developments Team
Industrial Hygiene Chemistry Division
OSHA Salt Lake Technical Center
Salt Lake City, UT 84115-1802
|
- Introduction
The Force Medical Project Advanced Concept Technology Demonstration program was
established in 1999 to
determine the utility of existing and emerging technologies to monitor U.S. military personnel for exposure to
toxic chemical substances. The technologies include real-time and non-real-time individual
chemical exposure samplers and alarms, and biological detection systems. The demand for such samplers, alarms,
and detection systems is evident in Presidential Review Directive 5, Department of
Defense Directive 6490.2, Department of
Defense Instruction 6940.3, and in a National Academy of Science study entitled "Strategies to Protect the
Health of Deployed U.S. Forces". The toxic chemical substances to be monitored include chemical warfare
agents (CWAs) and toxic industrial chemicals (TICs). Monitoring exposures to CWAs is
the primary objective
of the program. Monitoring TIC exposures is viewed as an important, but secondary, benefit of the program.
The Marine Corps was designated as technical manager for the program and has proposed use of diffusive
sampling with analysis by thermal desorption and gas chromatography/mass spectrometry (GC/MS) to develop
part of the monitoring program. Diffusive sampling has been shown to be a useful sampling technique for
many
volatile chemicals that does not require the use of cumbersome sampling pumps.
Thermal desorption is an excellent means to extract many chemicals from the sampling
medium with high efficiency. GC/MS is a
method of analysis with the capability to simultaneously identify and to quantitate the vast number of possible
toxic chemical exposures.
The U.S. Occupational Safety and Health Administration’s Salt Lake Technical Center
(OSHA SLTC) laboratory
was contacted by the Marine Corps and asked to participate in a project to test two prototype diffusive
samplers. SLTC was asked to perform laboratory research to test the sampling performance of the samplers
with selected TICs, and also to test the proposed thermal desorption and GC/MS analytical
technique. OSHA
has no legal authority to protect the health of military personnel, but a decision was made to participate in the
interests of interagency cooperation and in research that should result in a versatile sampling and analytical
technique that may have application in OSHA’s workplace monitoring program.
- Reagents
The following TICs were jointly selected by the Marines Corps and by OSHA.
TICs
Benzene, (Bz), [CAS 71-43-2], Aldrich Chemical Company (Milwaukee, WI), 99.0%,
A.C.S. Grade, lot no. BU 03051PS.
Ethylbenzene, (EtBz), [CAS 100-41-4], Aldrich Chemical Company (Milwaukee, WI), 99.8%, anhydrous, lot no. HI 03545DI.
1, 1, 2, 2-Tetrachloroethane, (TCA), [CAS 79-34-5], Aldrich Chemical Company
(Milwaukee,
WI), 98%, lot no. 08330EI.
Mesitylene, (1, 3, 5-Trimethylbenzene), (TMB), [CAS 108-67-8], Aldrich Chemical Company
(Milwaukee, WI), 98%, lot no. 00608TU.
(R)-(+)-Limonene, (1-methyl-4-isopropenyl-1-cylohexene), (LIM),
[CAS 5989-27-5], Aldrich Chemical Company (Milwaukee, WI), 97%, lot no. 1006CI.
Undecane, (UND), [CAS 1120-21-4], Aldrich Chemical Company (Milwaukee, WI), 99+%, lot
no. 1254AI.
2,2-Dichlorovinyldimethyl phosphate, (DDVP, also known as Dichlorvos), [CAS 62-73-7], Pfaltz
and Bauer, Inc. (Waterbury, CT), 99%, lot no. 111084-2.
A TIC mixture was prepared in the following proportions: Bz 36 mL; EtBz 36 mL; TCA 20
mL; TMB 36 mL; LIM 37 mL; UND 42 mL; and DDVP 22 mL. This mixture was used to generate
test atmospheres, and to prepare analytical standards and test samples.
The following analytical reagents were used in analysis of samples.
Analytical Reagents
Toluene-d8, (tol-d8), [CAS 2037-26-5], Scott Specialty Gases (Longmont, CO),
Certified 0.1%
in nitrogen gas, lot no. 11413CI. This gas was used as an internal standard for thermally-desorbed samplers.
Tol-d8 was selected for use as an internal standard because it does not
exist in nature.
Methyl alcohol, [CAS 67-56-1], Fisher Scientific (Fair Lawn, NJ), HPLC Grade, lot no. 011264.
This material was used to dilute the TIC mixture.
Carbon disulfide, [CAS 75-15-0], Aldrich Chemical Company (Milwaukee, WI), 99.9+%, low
benzene, lot no. TI 01762PI. This material was a component of the solution used to desorb
samples analyzed by GC/FID.
N,N-Dimethylformamide, (DMF), [CAS 68-12-2], Fisher Scientific (Fair Lawn, NJ), Certified
A.C.S. Grade, lot no. 902902. This material was a component of the solution used to desorb
samples analyzed by GC/FID.
Dodecane, [CAS 112-40-3], Aldrich Chemical Company (Milwaukee, WI), 99+%, lot no. EI
03040LU. This material was used as an internal standard for solvent-desorbed
samples.
A solution of 99% carbon disulfide and 1% DMF was used to desorb samples that were
analyzed by GC/FID. Dodecane (0.5 µL/mL) was added to this solution for use as an internal
standard.
- Air Samplers
Air samplers included project and non-project samplers. Project samplers were those that the Marine Corps
selected for testing. Non-project samplers were selected for testing by OSHA
SLTC.
Project Samplers
Ultra Passive Sampler, SKC, Inc. (Eighty Four, PA). A prototype diffusive sampler based on
the SKC 575 Series of samplers and containing 300 mg of Tenax TA, lot no. 1665.
GoreSorber, W.L. Gore and Associates (Elkton, MD). A proprietary diffusive
sampler
containing 30 mg of Tenax TA in each of two PTFE-like cartridges. Each GoreSorber sampler
had a barcode label with a unique number beginning with the letters "AA", or "AAA".
Non-Project Samplers
Ultra Passive Sampler-Reduced Sampling Rate, (SKC Ultra RSR), SKC, Inc. (Eighty Four,
PA). A prototype diffusive sampler based on the SKC 575 Series of samplers and containing
300 mg of Tenax TA, lot no. 1665. This sampler is similar to the SKC Ultra Passive Sampler
but has fewer holes in the inlet. These samplers were tested in a limited number of sampling
rate experiments.
Perkin Elmer (PE)-type sampler, Marks international, (Pontyclum, UK). A diffusive sampler
containing 200 mg of Tenax TA, or 400 mg of Carbopack B, or 300 mg of Chromosorb 106. Each sampler had a unique serial number that was
etched onto the tube beginning with the letters "Mi". PE-type diffusive sampling caps
containing membranes were used with these
samplers. These samplers were not obtained directly from PE, but are referred to in this work
as PE samplers. PE samplers have lower sampling rates than the project samplers and,
therefore, should be capable of sampling for longer times. PE samplers containing sorbents
other than Tenax TA were tested in limited work to provide supplementary information.
Carbon Molecular Sieve (CMS) sampling tubes, SKC, Inc. (Eighty Four, PA). An active
sampling tube containing two sections (75/150 mg) of Anasorb CMS, SKC Catalog no. 226-121, lot no. 1879. These samplers were used to help establish actual concentrations of test
atmospheres.
575-002 Passive Samplers, SKC, Inc. (Eighty Four, PA). A diffusive sampler
containing 500
mg of Anasorb 747. Various lot nos. including lot no. 1840. These samplers were used as
controls.
3520 Organic Vapor Monitor (OVM), 3M (St. Paul, MN). A diffusive sampler containing two
charcoal wafers. Various lot nos. including lot no. 10-02 112010. These samplers were also
used as controls.
The Marine Corps requested that OSHA partially test SKC Ultra RSR samplers. The
number of available SKC
Ultra RSR samplers was insufficient for other than preliminary testing. The project samplers and PE Tenax TA
samplers were those most tested in this work.
- Apparatus
 |
| Figure 1. Perkin Elmer TurboMatrix ATD. |
Thermally-desorbed samples were analyzed using a Perkin Elmer (Norwalk, CT)
TurboMatrix ATD (equipped
with internal standard addition option) connected to the electronic-pressure controlled volatiles interface inlet of
an Agilent Technologies (Wilmington, DE) 6890 Series GC system and an Agilent 5973 Network Mass Selective
Detector (MSD). The ATD (automatic thermal desorber) and GC carrier gas was helium. Thermal- desorption tubes were loaded on the ATD carrousel-tray so that the
end with the groove was upward. The groove identifies the front of the tube. ATD conditions:
thermal-desorption tubes containing Tenax TA were desorbed at 275°C for ten min
following a 1-min ambient purge. Tubes
containing Chromosorb 106 were desorbed at 225°C, and Carbopack B at 350°C for the same times. The
focusing trap was flash heated from -30°C to 300°C and maintained at the upper temperature for 2 min. The GC
transfer line temperature was 225°C and the valve temperature was 225°C. The
inlet split flow was 22 mL/min, the desorb flow 58 mL/min, the internal standard tube load flow 23
mL/min, and the internal standard loop flow
1.3 mL/min. GC conditions: a Restek (Bellefonte, PA) Rtx-5 capillary GC column (30-m × 0.25-mm
i.d. × 0.25-µm df) was used for this work. The GC column was temperature programmed from 40°C
(following a one-min hold) at 20°C/min to 230°C. The GC column was operated in the constant flow mode at 0.8
mL/min. The GC
inlet temperature was 230°C, and the inlet split ratio was usually 67 to 1. MSD conditions: thermal auxiliary
2 temperature was 280°C, MS source was 230°C, MS quad was 150°C. The MSD was
operated in the full scan mode from 24 to 350 AMU. Typically, a delay of 2 min was used to allow air and solvent used to prepare
analytical standards to clear the MSD before the filament was energized. Electron multiplier
(EM) voltage was
automatically set by the MSD software via an "autotune" performed each week of
operation. A photograph of the ATD is shown in Figure 1.
SKC CMS, SKC 575-002, and 3M 3520 samplers were analyzed (after desorption with carbon
disulfide/DMF solution for one hour) using a Hewlett Packard (Avondale, PA) 5890 GC equipped with a Restek Rtx-5 capillary
column (60-m × 0.32-mm i.d. × 1.5-µm df), an automatic liquid injector, and an FID. The GC column was
temperature programmed from 40°C (following a one-min hold) at 6°C/min to 190°C. The FID was maintained
at 250°C and the injector at 220°C. The GC column flow was 4.0 mL/min hydrogen. The inlet split ratio was
50 to 1. The FID gases were 35-mL/min hydrogen, 415-mL/min air, and 30-mL/min auxiliary nitrogen.
Test samples were collected from dynamically generated test atmospheres using an apparatus constructed
from stainless steel. The apparatus consisted of either a large (cross-section area: 161
cm2) or a small (cross-section area: 40 cm2) exposure
chamber designed to permit exposure of a large number of diffusive
samplers to test atmospheres. The chambers were of different size to permit sample exposure at
significantly different
face velocities. A stainless-steel manifold was connected in-line and following the exposure chambers to permit
collection of active samples.
Humid air for use with controlled-test atmospheres was generated using a Miller-Nelson Research (Monterey,
CA) Model HCS 501 Flow-Temperature-Humidity Control System. This system was equipped with a 500-L/min
mass flow controller.
Relative humidity and temperature of the test atmospheres were monitored with an Omega (Stamford, CT)
Digital Thermo-Hygrometer meter and probe. The probe was calibrated by the manufacturer. Pressure within
the exposure chambers was monitored with an Omega meter and pressure transducer. The transducer was
calibrated to read ambient barometric pressure before each run. Ambient barometric pressure was
measured with a Princo (Southampton, PA) NOVA mercury barometer.
Dilution air flow rates (50-360 L/min) were measured with an Equimeter (Rockwell International, Pittsburgh, PA)
No. 750 gas meter. Equimeter readings at several different flow rates were compared to those of a Singer
(Philadelphia, PA) DTM 115 gas meter (which had been tested by the local natural gas distributor and found
to be accurate) that was connected in series before the Equimeter. Both meters gave very similar readings.
The TIC mixture was metered into the system with an Isco (Lincoln, NE) 100 DM syringe pump equipped with
a cooling/heating jacket and an insulation cover package. The pump was operated in the constant flow mode. The temperature of water in the cooling/heating jacket was maintained at 23°C with a Forma Scientific
(Marietta, OH) Model 2006 CH/P Bath and Circulator.
TIC vapors were generated by pumping the liquid mixture into a heated glass tube where it evaporated into the
dilution air stream. This vapor generator consisted of a 10-cm length of ¼-inch
diameter glass tubing with a small hole in the side. The hole was just large enough for 1/16-inch diameter tubing to be inserted. The glass
tubing was placed inside a ½-inch stainless steel Swagelok (Solon, OH) tee wrapped with heating tape that
was heated when electricity was applied with a variable-voltage transformer. The 1/16-inch tubing entered the
third port of the tee through an adaptor and was inserted about ⅛-inch (approximately in the center) into the
glass tubing through the small hole. Solvent was pumped through the 1/16-inch tubing and into the glass tubing.
The liquid flow rate was slow enough (0.3 to 20 µL/min) so that it did not accumulate in the heated evaporation
tube. The entire dilution air stream passed through the tee and swept generated vapors into the apparatus.
 |
| Figure 2. Diffusive sampler exposure apparatus with small
exposure chamber. |
The following is a description of the arrangement of the apparatus that was placed in a walk-in hood. Liquid
from which vapors were to be dynamically generated was pumped with a precision Isco syringe pump (an
identical pump and a small solvent mixing tee was available when its use was desired) into a heated glass tube
where it evaporated. The generated vapors were swept from the glass tube with dilution air. Stainless steel
tubing (½-inch o.d.) connected with stainless steel Swagelok fittings was used to
transfer the test atmosphere.
The dilution air was humidified (if desired) using a Miller-Nelson Flow-Temperature-Humidity
controller. The vapor/dilution air mixture then passed into a 3×24-inch stainless steel mixing chamber. The test atmosphere
next passed through ½-inch ball valves where it could be either diverted to waste, or
directed into the exposure
chamber. An additional ball valve allowed the chamber to be purged with room air. The transfer tubing diameter
was increased from ½ inch to 1 inch at this point using a Swagelok adaptor attached to the chamber inlet. Tube and fitting diameter was increased to 1 inch after
this fitting to help reduce any increased pressure to ambient. The 1-inch o.d. chamber inlets have small
stainless steel deflectors to help insure that the test atmosphere completely fills the sampling
chambers. Stainless steel screens were placed inside the chamber
for the same purpose. This design should cause air flow through the chamber to be somewhat turbulent. Face velocities of the test atmospheres were calculated by
dividing the volumetric flow of each atmosphere by the cross-sectional area available for air flow in each
chamber. The cross-sectional area available for air flow was the cross-sectional area of each chamber reduced
by the cross-sectional areas of the samplers. A photograph of the apparatus with the small sampling
chamber is shown in Figure 2. Exposure chambers are used with removable doors (not shown in the photograph) to completely seal them when used with test
atmospheres.
- Experimental
This laboratory research was performed partially based on techniques and tests described in OSHA methods
development protocols1,2. The TIC levels studied in this work were jointly
selected by the Marine Corps and by OSHA. They are not associated with any military or OSHA exposure standard.
Preparation of Samples
 |
| Figure 3. Ultra sampler. |
SKC Ultra samplers were prepared for analysis by prying open the back of the sampler with a screwdriver to reveal
the small end of the built-in aluminum funnel containing the sorbent. The small end of the funnel is sized to fit
inside the thermal-desorption tube. The Tenax TA was then carefully transferred into the back of a clean, empty
thermal-desorption tube using the funnel. The front of a thermal-desorption tube is the end with the groove and
the back is the opposite end. A gauze screen was then carefully placed on top of the Tenax TA with the aid of a
3/16-inch glass rod and a small screwdriver. The tube was not completely filled with the sorbent and an empty
space of approximately ½ inch remained in the end of the tube. A retaining spring was placed into this space,
and then seated on top of the gauze screen and slightly below the end of the tube using the small end of the
funnel to press the spring into position. A photograph of an SKC Ultra sampler is shown in Figure 3.
 |
| Figure 4. GoreSorber sampler. |
GoreSorber samplers were prepared for analysis by transferring one of the PTFE-like cartridges into the end
of a clean, empty thermal-desorption tube and then pushing it forward with a
3/16-inch glass rod. The other cartridge can then be placed in the same tube, or
analyzed separately if desired. Both GoreSorber cartridges were usually analyzed in the same thermal-desorption tube in this work. A retaining spring is placed
into the end of the tube, and then seated slightly below the end of the tube with the funnel from an SKC Ultra
sampler. A photograph of a GoreSorber sampler and a thermal-desorption tube is shown in Figure 4. The
tube endcaps shown are used to seal the samples for
analysis with the ATD.
 |
| Figure 5. PE Tenax TA sampler. |
PE samplers require no preparation for analysis. A photograph of one of these samplers is shown in Figure
5. A diffusive sampling cap with membrane is shown in the upper right section of the figure. A gauze screen and
retaining spring is shown in the right center. The vial shows the amount of sorbent contained within the tube. PE samplers are thermally preconditioned by the vendor
and can be reused numerous times after thermal reconditioning.
CMS active sampling tubes were prepared for analysis by placing each section of sorbent into separate 2-mL
glass vials and then adding 1 mL of 99/1 carbon disulfide/DMF solution. These vials were allowed to
stand one hour before analysis, and were shaken by hand several times during this time.
SKC 575-002 samplers were prepared for analysis by adding 2 mL of 99/1 carbon disulfide/DMF solution
through one of the two sampler ports designed for that purpose. The sampler was then
shaken on an SKC Sorbent Extractor (also designed for that purpose) for one hour before analysis.
3M 3520 OVMs were prepared for analysis by removing the two charcoal wafers, placing each wafer into a
separate 4-mL glass vial, adding 2 mL of 99/1 carbon disulfide/DMF solution to each
vial, and then desorbing the sample on a tube rotator for one hour before analysis. The wafers were removed from the sampler to
eliminate any possibility of desorption solvent loss from in situ desorption.
Analytical Standards
Media standards for thermally-desorbed samples were prepared by spiking 5-µL aliquots of a series of diluted
TIC mixtures onto thermal-desorption tubes containing the same medium used for
sampling. Standards used for SKC Ultra and GoreSorber samplers were spiked on the back of the thermal-desorption tube. Clean,
room air was drawn through these tubes at 50 mL/min for about 10 seconds immediately after they were spiked to
help ensure that the spiked TICs resided on the sampling medium. This air entered the thermal-desorption tube
at the end spiked. The source of the media desorption tubes used to prepare standards for SKC Ultra and
GoreSorber samplers was SKC Ultra samplers that had been previously analyzed, and then reconditioned for
10 min at 275°C using the ATD tube-conditioning feature. These thermal-desorption tubes were reused
approximately three or four times before recycling. Recycling thermal-desorption tubes was accomplished by
removing the retaining spring, the gauze screen, and the sorbent from previously analyzed samples. The empty
tubes were washed twice with methyl alcohol and then air-dried overnight before
reuse.
Analytical standards for PE samplers containing Chromosorb 106 or Carbopack B were prepared using
reconditioned PE samplers containing the same medium used for sampling. PE samplers were thermally
reconditioned by heating them at the same temperature used for desorption for 10 min. Standards for PE Tenax
TA samplers were prepared using reconditioned SKC Ultra sampling medium. Standards for all PE samplers
were prepared by liquid spiking the TIC mixture on the front of the tube.
Only limited work was performed to prepare standards using GoreSorber cartridges. Liquid spiking the
cartridges would require a needle hole to be punctured in the PTFE-like cartridge wall and this would cause
standards to be significantly different than samples. Some success was obtained by placing the cartridge in
a glass vial, spiking the liquid standard into the vial, and allowing the vial to stand overnight at 40°C before
transfer to the thermal-desorption tube and subsequent analysis. This technique was not fully
developed.
Analytical standards for solvent-desorbed samples were prepared by injecting microliter volumes of diluted TIC
mixtures into the same volume of 99/1 carbon disulfide/DMF desorption solution used to desorb the samples.
MSD Calibration
GC/MSD calibration curves were prepared for each TIC (for each sample set) by first calibrating the MSD
software in terms of mass per sample with a single standard. MSD response was plotted
against actual TIC mass per standard using an electronic spreadsheet. Sample results were calculated using the
calibration-curve equation obtained for each TIC. The calibration range, in most cases, was one-half to two times the expected
sample concentration. The MSD is saturated when approximately 50 ng of a TIC reaches the detector and
accurate quantitation becomes impossible after this point. A combination of GC injector split and/or ATD split
should be employed to reduce mass reaching the MSD when samples contain sufficiently high TIC levels.
Alternatively, MSD EM voltage could manually be reduced to prevent premature saturation. All MSD results
were calculated without internal standard correction.
GC/FID calibration curves were also prepared using an electronic spreadsheet for samples analyzed by solvent
desorption. A Waters Associates Millennium Chromatography Manager data system was used to measure
FID response. Dodecane was employed as an internal standard in the analysis of these samples.
MSD Detection Limits
Detection limits for the studied TICs were determined using the procedure described in OSHA methods
development guidelines3. A series of standards was prepared with the highest giving a MSD response
approximately ten times the MSD baseline noise. These standards were analyzed and MSD response plotted
against mass injected on-column. The detection limit was defined as the mass equivalent to three times the
SEE (standard error of estimate) of the regression line. SEE is the dispersion of data about a
regression line, and it is mathematically similar to standard deviation for a data set.
Desorption Efficiency
Desorption efficiency studies can reveal adverse interactions between TICs and the sampling media. Desorption
efficiencies for components of the TIC mixture were determined by liquid spiking the samplers with aliquots of
the TIC mixture. No desorption efficiency work was performed for PE samplers because these spiked tubes
would be identical to analytical standards. The spiking technique was routine except for GoreSorber samplers. The GoreSorber cartridges were removed from the outer pouch, one of
the two cartridges was liquid spiked by penetrating the PTFE-like shell with a syringe needle, and then both were placed in a sealed glass vial. All
spiked samples were allowed to equilibrate overnight before analysis. SLTC experience has shown that
desorption efficiency results from wet sampling media can occasionally be different than from dry media. This
is unlikely for Tenax TA because that sorbent is known to have low affinity for water. Desorption from wet SKC
Ultra samplers was, however, tested to a limited degree. Wet media were prepared by sampling a clean
atmosphere containing 80% relative humidity for four hours with dry samplers. Wet media were liquid spiked
with aliquots of the TIC mixture after exposure to humid air. Dry samplers were as received from the vendors.
Desorption efficiency studies were performed using both wet and dry samplers for CMS sampling tubes, SKC
575-002 samplers, and 3M 3520 OVMs.
Sampling Rate
Sampling rates for the components of the TIC mixture were determined for project, non-project, and control
samplers. Sampling rates were determined at ambient temperature and pressure, but all sampling rates
presented in this work are expressed at 760 mmHg and 25°C. Three of each
type of sampler was exposed to controlled test atmospheres for increasing time periods in the small exposure chamber for these experiments. The exposure times ranged from two to as long as 16 hours. The
relative humidity of the test atmospheres was approximately 80% at ambient temperature. The face velocity of test atmospheres
through the small exposure chamber was approximately 0.4 m/s. The concentration of each TIC in the test atmospheres was
approximately 4 mg/m3. Sampler orientation was parallel to the flow direction of the test atmosphere, except
for PE samplers which was perpendicular. This orientation of PE samplers was
necessitated because of space limitation within the small chamber.
Long-Term Sampling Capacity
A long-term sampling capacity experiment was performed with SKC Ultra,
GoreSorber, and PE Tenax TA
samplers. Twenty-one of each sampler were placed in the large exposure chamber, and three of each sampler
were removed at approximately eight-hour intervals over a total exposure time of 54 hours. The relative humidity
of the test atmospheres was approximately 80% at ambient temperature. The face velocity of the test
atmosphere through the large exposure chamber was approximately 0.1 m/s. The concentration of each TIC
in the test atmosphere was approximately 0.4 mg/m3. Sampler orientation was parallel to the flow direction of
the test atmosphere, except for PE Tenax TA samplers which was perpendicular.
Reverse Diffusion
A reverse diffusion experiment was performed by exposing SKC Ultra,
GoreSorber, and PE Tenax TA samplers
(six of each) to a test atmosphere containing approximately 4 mg/m3 of each TIC for four hours. The relative
humidity of the test atmosphere was approximately 80% at ambient temperature and the face velocity was
about 0.4 m/s. Following the four-hour exposure time, the samplers were removed and the chamber was
flushed with clean, humid air. Then, three of each sampler was replaced in the
chamber and exposure continued for an additional four hours with clean, humid air. The analytical results for samples exposed to the
test atmosphere and then additionally to clean air were compared to results for samples exposed to only the
test atmosphere.
Storage Stability
SKC Ultra and GoreSorber storage stability samples were simultaneously prepared by sampling a test
atmosphere with 33 of each sampler. Thirty-three PE Tenax TA samplers were used to sample another, but
similar, test atmosphere. The concentration of the test atmospheres was approximately 4
mg/m3 for each TIC, and the exposure times were four hours. The relative humidity of the test atmospheres was approximately 80%
at ambient temperature and the face velocity was about 0.4 m/s. Three of each type of sampler was analyzed
on the day they were generated, and the remaining 30 of each type were split into two sets of 15 each. One
set of each type of sampler was stored in a laboratory oven maintained at 40 °C to simulate hot ambient
environments, and the other set in a refrigerator at -4°C to simulate cold environments, and to test if refrigeration
could stabilize any TIC that might degrade at the higher temperature. The samplers were all sealed using
original packaging materials as received from vendors. Three of each set of each type of sampler were removed
from storage and analyzed at approximately three-day intervals.
Factor Test
A 16-run factor test was performed using a modified version of the test discussed in the NIOSH protocol for
evaluation of diffusive samplers4. This test is performed primarily to observe the degree with which sampling
rates change under the environmental conditions of the test. These conditions are significantly altered following
a prescribed regimen for each run of the test. The pooled RSD of the sampling rates is called sampling rate
variation5 (SRV) by OSHA SLTC, and it has been established as a measure of
sampling error for diffusive samplers. SRV is analogous to the often-cited ±5% sampling pump error used to estimate sampling error for
active samplers.
Six factors were identified in the NIOSH protocol as having the potential to affect sampler performance. These
factors are analyte concentration, exposure time, face velocity, relative humidity,
interferant, and sampler orientation. Sixty-four experimental runs (26) would be required to fully evaluate combinations of each factor
at two levels. NIOSH recognized that this would be an excessive number of tests, and has devised a 16-run
fraction of the full factorial that is capable of revealing any of these factors
having a significant effect, free of two-factor interactions, on sampler performance. Some two and three-factor interactions can also be screened by
this design. The test is based on the comparison of each factor effect to experimental error so that the
significance of that effect can be determined.
Experimental design and conditions are shown in Table 1. Interferant was provided by the components of the
TIC mixture, for example, if EtBz was being examined then Bz, TCA, TMB, LIM,
UND, and DDVP were the interferants, and the levels were declared either high or low. Each of the samplers has a slightly different
cross-sectional area and this difference will slightly affect the face velocity to which that sampler is exposed. The
following calculation is applicable only to high face velocity (1.5 to 2.1 m/s)
experiments performed in the small exposure chamber. Experiments at low face velocity were performed in the large
chamber, and the differences in sampler cross-sectional area are small compared to chamber cross-sectional area. Face velocity shown
in Table 1 is for SKC Ultra samplers. Multiply the value by 0.90 to calculate parallel orientation face velocity
for GoreSorber samplers, and by 1.05 for perpendicular orientation. Exposure chamber size limitation
necessitated that only perpendicular orientation be used for PE Tenax TA samplers. PE sampler
orientation was either upward so that the test atmosphere first encountered the diffusion cap, or the sampler was inverted
so that the test atmosphere first encountered the back endcap. Upward orientation was designated
perpendicular orientation, and inverted designated parallel. Multiply by 0.91 to calculate parallel
orientation face velocity for PE Tenax TA samplers, and by 0.86 for perpendicular
orientation.
Table 1
Experimental Design of the Factor Test |
run
no. |
concn
(mg/m3) |
RH
(%,ºC) |
inter
level |
time
min |
face vel
(m/s)* |
sampler
orien |
| 1 |
2 |
24,25 |
low |
240 |
1.8 |
perp |
| 2 |
7 |
10,22 |
low |
60 |
0.1 |
perp |
| 3 |
2 |
81,21 |
low |
60 |
1.5 |
paral |
| 4 |
7 |
75,23 |
low |
240 |
0.1 |
paral |
| 5 |
2 |
9,22 |
high |
240 |
0.1 |
paral |
| 6 |
7 |
26,24 |
high |
62 |
1.5 |
paral |
| 7 |
2 |
75,21 |
high |
60 |
0.1 |
perp |
| 8 |
7 |
66,27 |
high |
240 |
1.8 |
perp |
| 9 |
7 |
78,22 |
high |
60 |
0.1 |
paral |
| 10 |
2 |
81,20 |
high |
240 |
1.5 |
paral |
| 11 |
7 |
8,23 |
high |
240 |
0.1 |
perp |
| 12 |
2 |
7,22 |
high |
60 |
2.1 |
perp |
| 13 |
7 |
66,25 |
low |
60 |
1.8 |
perp |
| 14 |
2 |
72,22 |
low |
240 |
0.1 |
perp |
| 15 |
7 |
22,26 |
low |
254 |
1.5 |
paral |
| 16 |
2 |
11,19 |
low |
60 |
0.1 |
paral |
| inter = interferant, orien = orientation, perp = perpendicular, paral =
parallel, vel = velocity, *face vel is for Ultra samplers |
|
Precision and Accuracy
The NIOSH acceptability criterion (published in their diffusive sampler evaluation
protocol6) for accuracy is that the method provide results within ±25% of the reference value at the 95% confidence level over the range 0.5
to 2 times the target level of the method. NIOSH specified which data to test against the criterion and those
specifications were followed as closely as possible with data obtained in this work. Relative standard
deviations (RSD) obtained from the analysis of three replicate samples of each type exposed at 2, 4, and 7
mg/m3 for each TIC were pooled after first examining them for homogeneity at the 95% confidence level with
the Cochran Test. RSDs from samples exposed at 0.4 mg/m3 were pooled separately to evaluate precision and
accuracy at lower levels.
OSHA methods acceptance criteria require that candidate sampling and analytical methods provide sample
results at least 75% (±25%) of the method target level at the 95% confidence level, and that uncorrectable bias
be less than 10%. OSHA evaluates precision and bias using data from the storage stability test.
Precision and accuracy data were evaluated for SKC Ultra, GoreSorber, and PE Tenax TA samplers.
Packaging Integrity
An experiment was performed to determine if manufacturer’s packaging is sufficient to prevent contamination
of unused samplers placed in contaminated environments. Two each, SKC Ultra,
GoreSorber, and PE Tenax
TA samplers, unopened and as received from the vendors, were placed in the exposure chamber and allowed
to remain undisturbed for 130 days. The samplers were exposed to test atmospheres for approximately 120
hours during the 130 days. The TIC concentrations of test atmospheres varied, but the average was about 4
mg/m3 for each TIC.
Concentrations of Test Atmospheres
Previous OSHA SLTC work7 has shown that use of active samplers as an independent means to determine
concentrations of test atmospheres gave more consistent experimental results than did calculation of
concentrations from generation parameters. This is because of normal day-to-day variations in the operation
of test equipment.
Theoretical concentrations of test atmospheres were calculated from the flow rate at which the liquid TIC
mixture was pumped into the vapor generator, the TIC concentrations of the liquid mixture, and the dilution air
flow rate. Actual concentrations of test atmospheres were determined from the analytical results of CMS
sampling tubes. The CMS tube sampling rate was 0.01 L/min, and four samples were taken simultaneously
with diffusive samples for every run. Average CMS results for the 16-run factor test, for example, were 100%
of theoretical with a pooled RSD of 15%. Comparison of sampling rates for SKC 575-002 and 3M 3520 control
samplers determined in this work with sampling rates published in the literature allows confirmation of the
validity of this practice. A comparison of sampling rates is presented in Table
21.
- Results and Discussion
MSD Calibration
MSD calibration data provides a means to estimate the contribution of the analytical procedure to overall
method imprecision. Calibration was performed with MSD response at the following selected ions: Bz 78.1,
EtBz 91.1, TCA 83.0, TMB 105.1, LIM 68.1, UND 57.1, DDVP 109.0, tol-d8 98.1
m/z.
Table 2
Standard Concentrations (µg per standard) |
| std |
Bz |
EtBz |
TCA |
TMB |
LIM |
UND |
DDVP |
| 1 |
4.10 |
4.10 |
3.95 |
4.00 |
3.95 |
4.05 |
3.90 |
| 2 |
3.25 |
3.25 |
3.15 |
3.20 |
3.15 |
3.25 |
3.15 |
| 3 |
2.45 |
2.45 |
2.35 |
2.40 |
2.35 |
2.40 |
2.35 |
| 4 |
1.65 |
1.65 |
1.60 |
1.60 |
1.60 |
1.60 |
1.55 |
| 5 |
0.80 |
0.80 |
0.80 |
0.80 |
0.80 |
0.80 |
0.80 |
|
The concentration of standards used for MSD
calibration is shown in Table 2. MSD response data
obtained over this mass range is shown in Table 3. The
combined GC and ATD split was 135 to 1.
The ATD automatic internal standard addition feature was activated for this calibration work, but MSD
response data were not corrected for the internal standard. Internal standard data are presented for
precision information only. No results obtained by thermal desorption presented in this report were corrected for internal standard. One reason for not using the
internal standard option was that the uptake of tol-d8 was about 115% for SKC Ultra samplers, about 60% for
GoreSorber samplers, and about 105% for PE Tenax TA samplers, when compared to the uptake of
standards. Use of an internal standard would obviously bias analytical results. Analytical standards would have to be
prepared using blank samplers (including the PTFE-like shell for GoreSorber cartridges) in order to successfully
employ the internal standard addition feature. Another reason was that the internal standard addition
mechanism did not function properly until the work was fairly well advanced.
Tol-d8 was used in gas form after dilution by the vendor to 0.1% by volume in nitrogen gas. The ATD employs a 0.5-mL loop for internal standard
addition, therefore, theoretically 2 µg of tol-d8 was injected with each
standard.
Table 3
MSD Calibration Data (MSD response) |
| std no. |
Bz |
EtBz |
TCA |
TMB |
LIM |
UND |
DDVP |
tol-d8 |
| 1 |
28641636 |
43454042 |
14312933 |
38846452 |
13247783 |
17202956 |
23440848 |
569535 |
| 1 |
30643376 |
45257645 |
14858385 |
41196355 |
13683133 |
11815700 |
26580610 |
560406 |
| 1 |
30445154 |
44581695 |
14725691 |
40398411 |
13658798 |
17827207 |
27048673 |
547299 |
| ave |
29910055 |
44431127 |
14632336 |
40147073 |
13529905 |
17715288 |
25690044 |
559080 |
| RSD |
3.69 |
2.05 |
1.94 |
2.98 |
1.81 |
2.63 |
7.64 |
2.00 |
| |
|
|
|
|
|
|
|
|
| 2 |
26286654 |
36684939 |
12022473 |
33621744 |
11224307 |
14781093 |
20340921 |
546155 |
| 2 |
26096894 |
36726922 |
11965531 |
33337284 |
11180821 |
14781093 |
20191208 |
548792 |
| 2 |
26964735 |
37238709 |
12012397 |
33915663 |
11539267 |
14889760 |
20662368 |
562188 |
| ave |
26449428 |
36883523 |
12000134 |
33624897 |
11314798 |
14821084 |
20398166 |
552378 |
| RSD |
1.73 |
0.84 |
0.25 |
0.86 |
1.73 |
0.40 |
1.18 |
1.56 |
| |
|
|
|
|
|
|
|
|
| 3 |
19767705 |
27856077 |
9007271 |
25398185 |
8621183 |
11296994 |
14195664 |
562952 |
| 3 |
20450297 |
28375303 |
9257640 |
26174815 |
8986079 |
11737467 |
15031324 |
560481 |
| 3 |
20932034 |
29042523 |
9526675 |
26680928 |
9043854 |
11778283 |
15988129 |
575954 |
| ave |
20383345 |
28424634 |
9263862 |
26084643 |
8883705 |
11604248 |
15071706 |
566462 |
| RSD |
2.87 |
2.09 |
2.80 |
2.48 |
2.58 |
2.30 |
5.95 |
1.47 |
| |
|
|
|
|
|
|
|
|
| 4 |
13046706 |
19682445 |
6383122 |
18079294 |
5839289 |
8156706 |
10839043 |
559044 |
| 4 |
13481463 |
20146261 |
6516168 |
18157601 |
6210200 |
8293435 |
10137451 |
577749 |
| 4 |
13021364 |
19539209 |
6308035 |
17353238 |
5966648 |
7870290 |
10786881 |
520656 |
| ave |
13183178 |
19789305 |
6402442 |
17863378 |
6005379 |
8106810 |
10587792 |
552483 |
| RSD |
1.96 |
1.60 |
1.65 |
2.48 |
3.14 |
2.66 |
3.69 |
5.27 |
| |
|
|
|
|
|
|
|
|
| 5 |
7332528 |
10204191 |
3273858 |
9255749 |
3223915 |
4351088 |
4884348 |
610041 |
| 5 |
7410959 |
10084991 |
3208730 |
9378546 |
3195213 |
4288999 |
4553462 |
630170 |
| 5 |
7309025 |
10020740 |
3227673 |
9203708 |
3136032 |
4281767 |
5133310 |
630563 |
| ave |
7350837 |
10103307 |
3236754 |
9279334 |
3185053 |
4307285 |
4857040 |
623591 |
| RSD |
0.73 |
0.92 |
1.04 |
0.97 |
1.41 |
0.88 |
5.99 |
1.88 |
| |
|
|
|
|
|
|
|
|
pooled
RSD |
2.42 |
1.59 |
1.76 |
2.14 |
2.23 |
2.01 |
5.38 |
1.74 |
|
The individual RSDs in Table 3 were tested for homogeneity with the Cochran Test and, with the exception of
tol-d8 data for std no. 4, found to be homogeneous. The homogeneous RSDs were pooled and the result is
shown in Table 3. These pooled RSDs are similar to those obtained for FID calibration data which are typically
1 to 2%. Instrument calibration was not a significant source of analytical error.
An example calibration curve for TMB is shown in Figure 6. SEE (standard error of estimate) shown in Figure
6 is equivalent to 87 ng over the calibrated range. Calibration is not performed using total-ion MSD response;
it is accomplished using MSD software for optimum response and selectivity. A total-ion chromatogram for std
no. 4 is shown in Figure 7.
Table 4
MSD Detection Limits (ng) |
| Bz |
EtBz |
TCA |
TMB |
LIM |
UND |
DDVP |
| 0.16 |
0.10 |
0.06 |
0.10 |
0.07 |
0.06 |
0.24 |
|
MSD Detection Limits
MSD detection limits are shown in Table 4. These amounts are mass injected on-column and
not mass per sample. The detection limits are based on precision of the analysis, and on
spectral quality of the resultant mass spectra.
An example of plotted data used to determine MSD detection limit for TMB is shown in Figure 8. A total-ion
chromatogram from the analysis of a standard spiked at approximately the MSD detection limits is shown in
Figure 9.
 |
|
Media standards for project samplers were prepared using the sampling medium from previously analyzed Ultra
samplers that have been thermally reconditioned. Analysis of reconditioned Ultra samplers gives clean
chromatograms. SKC conditions the Tenax TA to be packaged within Ultra samplers by solvent extraction and
subsequent heating to remove the solvent. The sorbent is not thermally conditioned at a sufficiently high
temperature, and the analysis of a sampler blank reveals many artifact peaks. An SKC representative has
stated that they intend to begin thermally conditioning the sorbent to reduce artifacts. Analysis of blank
GoreSorber samplers also shows a significant number of artifacts8. Total-ion chromatograms from the analysis
of SKC Ultra and GoreSorber blank samplers are shown in Figures 10 and 11. Chemicals identified on the SKC
Ultra blank include trioxane,1-methyl-2-propyl acetate, and 2-ethyl-1-hexanol. Present on the GoreSorber
blank are 2-ethyl-1-hexanol, 1-decene, and butylated hydroxytoluene. Attaining the detection limits in Table
4 for blank and field samples could be challenging because of co-eluting artifacts and other species. For
example, satisfactory MSD spectra for EtBz and for TCA were not obtained for blank SKC Ultra samplers
spiked at approximately the detection limits. Almost 2 ng of each TIC was required to obtain good spectra, and
this required spectral subtraction of the interfering species. Good spectra were obtained for the other TIC
components spiked on a blank SKC Ultra sampler at the stated detection limits. This data emphasizes the
need for samplers with low artifact levels.
Table 5
Desorption Efficiency From SKC Ultra Samplers (%) |
| |
Bz |
EtBz |
TCA |
TMB |
LIM |
UND |
DDVP |
| dry |
102.5 |
100.4 |
97.0 |
100.6 |
99.7 |
98.4 |
99.2 |
| wet |
92.3 |
93.1 |
94.1 |
94.3 |
95.6 |
95.7 |
99.7 |
|
Desorption Efficiency
Average desorption efficiencies for SKC Ultra samplers
are shown in Table 5. The mass range studied was 13
to 32 µg for each TIC per sample. Wet samplers were
prepared by exposing them to clean, humid air for four
hours prior to spiking. Some results from wet samplers
were lower than from dry samplers for unknown
reasons. These results were not used to correct SKC
Ultra sampler results presented in this report because it was assumed that desorption was 100%. This
assumption was supported by the fact that a second desorption of several SKC Ultra samples showed no TICs
present in significant amounts.
Table 6
Desorption Efficiency From GoreSorber Samplers (%) |
| |
Bz |
EtBz |
TCA |
TMB |
LIM |
UND |
DDVP |
| dry |
74.7 |
99.2 |
98.3 |
100.1 |
99.7 |
99.8 |
100.8 |
|
Average desorption efficiencies for GoreSorber samplers are shown in Table 6. The mass range
studied was 13 to 30 µg for each TIC per sample. A study using wet samplers was not performed. These
results were not used to correct GoreSorber results
presented in this report because it was assumed that desorption was 100%. This assumption was again
supported by the fact that a second desorption of several GoreSorber samples showed no TICs to be present
in significant amounts. The low desorption efficiency obtained for Bz was probably caused by the relatively low
affinity of Tenax TA for Bz, and its subsequent loss through the hole punctured in the PTFE-like shell of the
cartridge by liquid spiking the sample.
Table 7
Desorption Efficiency From Anasorb CMS (%) |
| |
Bz |
EtBz |
TCA |
TMB |
LIM |
UND |
DDVP |
| dry |
96.0 |
96.7 |
93.6 |
96.4 |
97.1 |
99.3 |
91.3 |
| wet |
94.2 |
95.8 |
80.4 |
95.2 |
92.7 |
98.3 |
68.6 |
|
Average desorption efficiencies for the front section (150 mg) of SKC Anasorb CMS sampling tubes are shown
in Table 7. The mass range studied was 10 to 165 µg of each TIC per sample. Wet samplers were prepared
by drawing clean, humid air through the samplers at 0.1 L/min for four hours prior to spiking. All CMS results
presented in this report were corrected using the appropriate desorption efficiency. Results for DDVP desorbed from wet samplers were low and would
necessitate use of another solvent or desorption technique in routine analytical work because a minimum
recovery of 75% is usually required for NIOSH and OSHA methods. The reason for low DDVP wet results is
unknown but could be due to hydrolysis. The low desorption was deemed adequate for this work because
samples were analyzed immediately after generation. Results for TCA from wet CMS were lower than from dry
CMS, but they were greater than 75% and, therefore, adequate for conditional use.
Table 8
Desorption Efficiency From SKC 575-002 Samplers (%) |
| |
Bz |
EtBz |
TCA |
TMB |
LIM |
UND |
DDVP |
| dry |
96.6 |
101.5 |
87.7 |
100.0 |
103.9 |
109.4 |
inter |
| wet |
94.0 |
95.9 |
47.3 |
94.0 |
95.2 |
100.3 |
inter |
|
inter=interference |
|
Average desorption efficiencies for SKC 575-002 samplers are shown in Table 8. The mass
studied was 13 µg of each TIC per sample. Wet samplers were prepared by exposing them
to clean, humid air for four hours prior to spiking. All SKC 575-002 samplers in this
study were exposed to humid air and, therefore, analytical results were corrected
using only the wet desorption efficiencies. A chromatographic interference prevented the analysis of DDVP. This co-eluting interference was identified as cyclohexyl isothiocyanate by GC/MS. The low wet desorption
efficiency obtained for TCA would necessitate use of another solvent or desorption technique in routine
analytical work, however, it was also deemed adequate for this work because these samples were analyzed
immediately after generation.
Table 9
Desorption Efficiency From 3M 3520 OVM Charcoal Wafers (%) |
| |
Bz |
EtBz |
TCA |
TMB |
LIM |
UND |
DDVP |
| dry |
96.7 |
101.5 |
91.2 |
94.8 |
94.4 |
95.0 |
86.2 |
| wet |
98.1 |
98.2 |
62.1 |
98.0 |
97.4 |
99.9 |
57.4 |
|
Average desorption efficiencies for 3M 3520 OVM charcoal wafers are shown in Table 9. The mass studied was 26 µg of each TIC per
sample. Wet samplers were prepared by exposing them to clean, humid air for four
hours prior to spiking. All 3M 3520 OVM samples in this study were exposed to humid
air and, therefore, analytical results were corrected using only the wet desorption efficiencies. The low wet
desorption efficiencies for TCA and DDVP would necessitate use of another solvent or desorption technique
in routine analytical work, however, they were again deemed adequate for this work because these samples
were analyzed immediately after generation.
Sampling Rate
Sampling rates were determined by exposing samplers to test atmospheres containing approximately 4
mg/m3 of each TIC (about 1 ppm) for increasing time intervals.
Sampling rates were calculated with the following equation and then converted to
mL/min:
ambient sampling rate = average mass collected/(concn of test atm × sampling time)
Mass was corrected for desorption efficiency only for SKC 575-002 and 3M 3520 OVM samplers.
Experimental sampling rates were determined at ambient temperature (Tamb) and barometric pressure
(Pamb), and were converted to their equivalent at 760 mmHg and 298K with the following equation:
sampling rate760 mmHg, 298K = sampling rateamb(298/Tamb)3/2(Pamb/760)
Sampling rates for the components of the TIC mixture were determined for SKC Ultra and GoreSorber project
samplers. Sampling rates were also determined for SKC Ultra RSR, PE Tenax TA, PE Chromosorb 106, and
PE Carbopack B non-project samplers. They were also determined for the 3M 3520 and SKC 575-002
samplers that were used as controls. DDVP was not thermally desorbed from PE Carbopack B samplers at
its recommended maximum desorption temperature. A chromatographic interference that eluted at the same
time as DDVP prevented its determination in SKC 575-002 samplers. Relative standard deviations
(RSD) were calculated for the three samplers of each type that were exposed for each time interval. RSD was excessive
in some cases, but all data were retained because of the small sample size and also to preserve the integrity
of precision results. Part of this work was to test and evaluate the analytical method.
Table 10
Sampling Rate Data for SKC Ultra Samplers |
| time |
Bz |
EtBz |
TCA |
TMB |
LIM |
UND |
DDVP |
| hours |
mL/
min |
RSD |
mL/
min |
RSD |
mL/
min |
RSD |
mL/
min |
RSD |
mL/
min |
RSD |
mL/
min |
RSD |
mL/
min |
RSD |
| 2 |
13.92 |
6.13 |
12.19 |
4.88 |
10.71 |
6.01 |
11.14 |
5.68 |
10.38 |
5.54 |
9.79 |
4.40 |
8.59 |
5.75 |
| 4 |
11.07 |
13.39 |
12.30 |
19.34 |
11.38 |
16.14 |
11.51 |
18.48 |
10.52 |
16.55 |
9.98 |
13.61 |
10.25 |
14.56 |
| 6 |
12.51 |
0.32 |
14.80 |
0.63 |
13.63 |
1.03 |
14.03 |
1.29 |
12.77 |
1.59 |
11.76 |
1.31 |
11.17 |
2.36 |
| 8 |
9.91 |
2.51 |
13.01 |
3.00 |
12.49 |
2.45 |
12.14 |
3.14 |
11.16 |
3.12 |
10.79 |
2.04 |
9.89 |
2.85 |
| 10 |
8.29 |
2.12 |
12.46 |
1.03 |
10.69 |
1.44 |
13.41 |
5.79 |
10.66 |
1.00 |
10.28 |
1.26 |
10.58 |
2.61 |
| 14 |
5.48 |
3.19 |
10.90 |
3.62 |
11.04 |
2.50 |
11.82 |
2.02 |
10.69 |
1.71 |
10.53 |
0.19 |
10.04 |
1.41 |
Table 11
Sampling Rate Data for GoreSorber Ultra Samplers |
| time |
Bz |
EtBz |
TCA |
TMB |
LIM |
UND |
DDVP |
| hours |
mL/
min |
RSD |
mL/
min |
RSD |
mL/
min |
RSD |
mL/
min |
RSD |
mL/
min |
RSD |
mL/
min |
RSD |
mL/
mi
n |
RSD |
| 2 |
9.29 |
7.64 |
23.78 |
4.93 |
21.58 |
5.50 |
21.02 |
5.06 |
19.11 |
4.53 |
21.35 |
4.66 |
22.12 |
6.68 |
| 4 |
4.16 |
6.93 |
19.32 |
11.07 |
19.74 |
14.45 |
18.46 |
13.75 |
16.25 |
12.50 |
18.91 |
13.23 |
21.27 |
15.37 |
| 6 |
1.30 |
13.17 |
20.12 |
12.84 |
20.38 |
14.21 |
19.34 |
15.23 |
16.20 |
15.01 |
19.96 |
12.84 |
21.31 |
17.78 |
| 8 |
0.83 |
3.19 |
11.53 |
8.24 |
15.97 |
4.52 |
15.55 |
4.63 |
13.07 |
3.66 |
18.07 |
2.75 |
21.58 |
1.80 |
| 10 |
0.77 |
14.99 |
7.11 |
26.00 |
13.74 |
6.81 |
12.23 |
9.42 |
9.97 |
12.46 |
15.54 |
8.24 |
21.19 |
6.12 |
| 14 |
0.45 |
11.58 |
6.04 |
13.77 |
10.64 |
8.57 |
10.72 |
8.31 |
9.32 |
9.26 |
14.70 |
8.09 |
18.20 |
6.48 |
Table 12
Sampling Rate Data for SKC Ultra RSR Samplers |
| time |
Bz |
EtBz |
TCA |
TMB |
LIM |
UND |
DDVP |
| hours |
mL/
mi
n |
RSD |
mL/
mi
n |
RSD |
mL/
mi
n |
RSD |
mL/
mi
n |
RSD |
mL/
mi
n |
RSD |
mL/
mi
n |
RSD |
mL/
mi
n |
RSD |
| 2 |
6.28 |
5.74 |
5.07 |
6.39 |
4.15 |
6.35 |
4.40 |
6.75 |
4.06 |
7.12 |
3.59 |
5.25 |
2.41 |
6.15 |
| 8 |
5.49 |
17.53 |
4.80 |
21.88 |
4.26 |
19.36 |
4.35 |
25.41 |
3.85 |
21.64 |
3.45 |
2.56 |
2.56 |
25.58 |
| 10 |
4.98 |
4.19 |
4.74 |
2.69 |
4.43 |
2.56 |
4.34 |
2.51 |
3.98 |
2.60 |
3.74 |
3.54 |
3.54 |
6.84 |
| 14 |
4.17 |
4.41 |
2.72 |
20.55 |
4.68 |
7.01 |
4.71 |
8.86 |
4.22 |
5.90 |
3.65 |
3.17 |
3.17 |
6.82 |
| | |