| N,N-DIMETHYLETHYLAMINE |
|
| Method no.: |
PV2096 |
|
| Control no.: |
T-PV2096-01-8711-CH |
|
| Matrix: |
Air |
|
| Target concentration: |
There is no OSHA PEL or TLV for N,N-dimethylethylamine. A target
concentration of 25 ppm was used for this study. |
|
| Procedure: |
Samples are collected by drawing a known volume of air through a jumbo
alumina tube, 400 mg reference section with a 200 mg backup section, lot 391. Samples are desorbed for 30 minutes
with 2 mL deionized water, which has been neutralized to pH of 7 and then analyzed by gas chromatography with a
nitrogen-phosphorous detector (GC-NPD). |
|
| Air volume and sampling rate studied: |
4 liters at 0.1 L/min |
|
| Special requirements: |
The capacity of the different lots of alumina tubes
varied greatly. If a different lot of alumina tube is used, the capacity should be checked. |
|
| Status of method: |
Stopgap method. This method has been only partially evaluated and is
presented for information and trial use. |
|
| Date: |
November, 1987 |
|
| Chemist: |
Mary E. Eide |
SOLVENTS BRANCH
OSHA ANALYTICAL LABORATORY
SANDY, UTAH |
1. General Discussion
1.1 Background
1.1.1 History of procedure
The OSHA lab has been recommending collection of N,N-dimethylethylamine with a 0.1 N H2SO4
bubbler following the NIOSH method (S 152) for collection of triethylamine (Ref 5.1). A paper by P.W. Langvardt and
R.G. Melcher recommends collection and analysis of ethanolamines and isopropanolamines using alumina tubes (Ref
5.2). Adsorbent tubes are more convenient, so alumina tubes were tried as a collection media for
N,N-dimethylethylamine. Different means of desorption and analysis were used than in the paper (Ref 5.2), since
the analyte of interest was N,N-dimethylethylamine. Alumina tubes in this study were desorbed with either 0.1 N H2SO4
or deionized water. Analysis was by GC-NPD. Two different lots and sizes of alumina tubes were tried, because the
capacity was different for the two lots. Lot 190 had a greater capacity, 6 liters with humid air (93% RH), but is
no longer available, so lot 391 jumbo tubes were evaluated. The desorption efficiency and the storage were good for
both lots studied.
1.1.2 Potential workplace exposure (Ref 5.3) N,N-dimethylethylamine is used in urea- and melamine-based enamels, as
an antilivering agent. N,N-dimethylethylamine is used in foundries.
1.1.3 Toxic Effects (This section is f or information purposes and should not be taken as the basis for OSHA
policy.) (Ref 5.3) N,N-dimethylethylamine has toxic effects similar to triethylamine, which causes corneal
damage, pulmonary irritation, cellular necrosis of the liver and kidneys, and skin irritation.
1.1.4 Physical properties (Ref 5.4):
Structure:

| Molecular weight: |
73.14 |
| Density: |
0.675 |
| Boiling point: |
37°C |
| Melting point: |
-140°C |
| Odor: |
strong ammoniacal odor |
| Color: |
clear liquid |
| Molecular formula: |
C4H11N |
| CAS: |
598-56-1 |
| IMIS: |
0915 |
| Flash point: |
-36°C |
1.2 Limit defining parameters
1.2.1 The detection limit of the analytical procedure is 5 ng, with a 1 µL injection volume. This is the smallest
amount which could be detected under normal operating conditions.
1.2.2 The overall detection limit is 0.3 ppm based on a 10 liter air volume.
1.3 Advantages
1.3.1 The sampling procedure is convenient.
1.3.2 The analytical method is reproducible and sensitive.
1.3.3 Reanalysis of samples is possible.
1.3.4 It may be possible to analyze other compounds at the same time.
1.3.5 Interferences may be avoided by proper selection of column and GC
parameters.
1.4 Disadvantages
None known.
2. Sampling procedure
2.1 Apparatus
2.1.1 A calibrated personal sampling pump, the flow of which can be determined within ±5% at the recommended
flow.
2.1.2 Alumina tubes, lot 391, containing 400-mg adsorbing section with a 200 mg backup section separated by a 2-mm
portion of urethane foam, with a silane treated glass wool plug before the adsorbing section and a 3 mm plug of
urethane foam at the back of the backup section. The ends are flame sealed and the glass tube containing the
adsorbent is 11-cm long, with a 8-mm O.D. and 6-mm I.D., SKC tubes or
equivalent.
2.2 Sampling technique
2.2.1 The ends of the alumina tubes are opened immediately before sampling.
2.2.2 Connect the alumina tube to the sampling pump with flexible tubing.
2.2.3 Tubes should be placed in a vertical position to minimize channeling, with the smaller
section towards the pump.
2.2 4 Air being sampled should not pass through any hose or tubing before entering the alumina tube.
2.2.5 Seal the alumina tube with plastic caps immediately after sampling. Seal each sample lengthwise with OSHA
Form-21.
2.2 6 With each batch of samples, submit at least one blank tube from the same lot used for air samples. This tube
should be subjected to exactly the same handling as the samples (break ends, seal, and transport) except that no
air is drawn through it.
2.2.7 Transport the samples (and corresponding paperwork) to the lab for analysis.
2.2.8 Bulks submitted for analysis must be shipped in a separate container
from the samples.
2.3 Desorption efficiency
A desorption efficiency of the 400/200 mg size alumina tube, lot 391, was performed with spikes of 0.3375, 0.675,
and 1.35 mg of N,N-dimethylethylamine. The tubes were stored at room temperature overnight. They were opened, each
section placed into a separate 4 mL vial, and desorbed with 2 mL deionized water. They were allowed to desorb for
30 minutes with occasional shaking, and then analyzed by GC-NPD (Table 2.3). The overall average desorption was
97.9%.
Table 2.3
Desorption Study Lot 391 |
|
| Tube# |
0.3375mg |
% Recovered 0.675mg |
1.35mg |
|
| 1 |
97.3 |
97.2 |
93.1 |
| 2 |
100 |
103 |
93.7 |
| 3 |
99.8 |
98.4 |
96.6 |
| 4 |
102 |
101 |
95.8 |
| 5 |
97.4 |
97.5 |
98.6 |
| 6 |
99.5 |
97.9 |
93.3 |
| average |
99.3 |
99.2 |
95.2 |
| |
|
|
|
| overall average |
97.9 |
|
| standard deviation |
±2.80 |
|
|
2.4 Retention efficiency
Retention efficiency of lot 391 jumbo alumina tubes was performed by spiking 0.675 mg N,N-dimethylethylamine onto
alumina tubes and drawing 4 liters of humid air, 94 % RH, through them. They were opened, each section placed into a
separate 4 mL vial, and desorbed with 2 mL deionized water . They were allowed to desorb for 30 minutes with
occasional shaking, and then analyzed by GC-NPD. While there was N,N-dimethylethylamine found on the backup portion
of the alumina tubes, the amount recovered averaged 96.3% (Table 2.4). The recoveries are corrected for desorption
efficiency.
Table 2.4
Retention Efficiency Lot 391 |
|
| Tube # |
% Recovered 'A' |
% Recovered 'B' |
% Total |
|
| 1 |
89.5 |
9.3 |
98.8 |
| 2 |
79.9 |
18.5 |
98.2 |
| 3 |
78.1 |
16.6 |
94.7 |
| 4 |
83.7 |
12.3 |
96.0 |
| 5 |
88.9 |
4.8 |
93.7 |
| 6 |
lost in analysis |
|
|
| |
|
|
|
|
|
average |
97.7 |
|
2.5 Storage
Storage stability for lot 391 jumbo alumina tubes was performed by spiking 0.675 mg N,N-dimethylethylamine on the
tubes, and storing them at room temperature. The tubes were opened, each section placed into separate 4 mL vials,
and desorbed with 2 mL deionized water. They were allowed to desorb for 30 minutes with occasional shaking, and then
analyzed by GC-NPD. The storage stability remained above 95% for the 14 days studied (Table 2.5). The values in the
table are corrected for desorption efficiency.
Table 2.5
Storage with lot 391 Alumina Tubes |
|
| Day |
% Recovered |
|
| 5 |
99.6 |
| 5 |
97.1 |
| 5 |
96.2 |
| 14 |
96.4 |
| 14 |
95.9 |
| 14 |
101 |
| |
|
| average |
97.7 |
|
2.6 Air volume and sampling rate studied
2.6.1 The air volume studied was 4 liters.
2.6.2 The sampling rate studied was 0.1 liter per minute.
2.7 Interferences
Suspected interferences should be listed on sample data sheets.
2.8 Safety precautions
2.8.1 Sampling equipment should be placed on an employee in a manner that does not interfere with work performance
or safety.
2.8.2 Safety glasses should be worn at all times.
2.8.3 Follow all safety practices that apply to the workplace being sampled.
3. Analytical method
3.1 Apparatus
3.1.1 Gas chromatograph equipped with a nitrogen-phosphorous detector.
3.1.2 GC column capable of separating the analyte from any interferences. A 6-ft glass column packed with 4%
Carbowax 20M, 0.8% KOH on Carbopack B was used for this study.
3.1.3 An electronic integrator or some other suitable method of measuring peak areas.
3.1.4 Four milliliter vials with Teflon-lined caps for sample desorption.
3.1.5 A 10 µL syringe or other convenient size for sample injection.
3.1.6 Pipets for dispensing the desorbing solution.
3.1.7 Volumetric flasks - 5 mL and other convenient sizes for preparing
standards.
3.2 Reagents
3.2.1 Purified GC grade nitrogen, hydrogen, and air.
3.2.2 Deionized water.
3.2.3 Reagent grade N,N-dimethylethylamine.
3.3 Sample preparation
3.3.1 Jumbo alumina tubes are opened and the front and back section of each tube are placed in separate 4 mL vials.
3.3.2 Each section of the tube is desorbed with 2 mL deionized water.
3.3.3 The vials are sealed immediately and allowed to desorb for 30 minutes
with occasional shaking.
3.4 Standard preparation
3.4.1 Standards are prepared by diluting a known quantity of N,N-dimethylethylamine with deionized water.
3.4.2 At least four separate standards should be made from at least two separate stock standards. A curve is
plotted, since the response can be non-linear on a NPD.
3.4.3 A standard of 337.5 µg/mL (0.5 µL/mL) N,N-dimethylethylamine in water corresponds to 57.62 ppm based on a 4
liter air volume, 2 mL desorption volume, and a 97.9% desorption efficiency
for lot 391 jumbo tubes.
3.5 Analysis
3.5.1 Gas chromatograph conditions.
| Flow rates (mL/min) |
Temperature (°C) |
| |
|
| Nitrogen: |
24 |
Injector: |
200 |
| Hydrogen: |
3 |
Detector: |
220 |
| Air: |
50 |
Column: |
120 |
| |
|
| Injection size: |
1 µL |
|
| Elution time: |
3.12 min |
|
| Chromatogram: |
 |
Figure 1. Standard of 675 µg/mL N,N-dimethylethylamine in water.
3.5.2 Peak areas are measured by an integrator or other suitable means.
3.6 Interferences (analytical)
3.6.1 Any compound having the general retention time of the analyte used is an interference. Possible interferences
should be listed on the sample data sheet. GC parameters should be adjusted if necessary so these interferences will
pose no problems.
3.6.2 Retention time data on a single column is not considered proof of chemical identity. Samples over the target
concentration should be confirmed by GC/Mass Spec or other suitable means.
3.7 Calculations
3.7.1 Standards are plotted on a curve and the micrograms/milliliter of the samples are taken from the curve.
3.7.2 To calculate the concentration of analyte in the air sample the following formulas are used:

3.7.3 The above formulas can be consolidated to form the following formula. To calculate the ppm of analyte in the
sample based on a 10 liter air sample:
| µg/mL |
= |
concentration of analyte in sample or standard |
| 24.46 |
= |
Molar volume (liters/mole) at 25°C and 760 mmHg. |
| MW |
= |
Molecular weight (g/mole) |
| DV |
= |
Desorption volume |
| 4 L |
= |
10 liter air sample |
| DE |
= |
Desorption efficiency |
3.7.4 This calculation is done for each section of the sampling tube and the results added
together.
3.8 Safety precautions
3.8.1 All handling of solvents should be done in a hood.
3.8.2 Avoid skin contact with all solvents.
3.8.3 Wear safety glasses at all times.
4. Recommendations for further study
Collection efficiencies should be performed. Any lot of alumina tubes other than the ones in this study should be
checked for retention efficiency before used in the field.
5. References
5.1 "NIOSH Manual of Analytical Methods", U.S. Department of Health, Education, and Welfare, Public Health
Service, Center for Disease Control, National Institute for Occupational Safety and Health, Second Edition, Vol. 3.,
Method S152.
5.2 Langvardt, P.W., Melcher, R.G., Anal. Chem., 1980, vol.
52, p. 669.
5.3 Proctor, N.H., Hughes, J.P., "Chemical Hazards of the Workplace", J.B. Lippincott Co., Philadelphia
PA, 1978, p. 497.
5.4 Weast, R.C., "Handbook of Chemistry and Physics", 67th Edition, CRC Press Inc., Boca Raton
FL, 1986, p. C247.
| |
|