| Method no.: |
PV2132 |
| |
|
| Control no.: |
T-PV2132-01-0403-M |
| |
|
| Target concentration: |
1 mg/m3 |
| |
|
| Procedure: |
Samples
are collected by drawing a known volume of air through glass sampling tubes
containing Chromosorb 106. Samples are
extracted with 99:1 carbon disulfide: N,N-dimethylformamide and analyzed
by GC using a flame ionization detector. |
| |
|
| Recommended sampling time and sampling rate: |
240 min at 0.2 L/min (48 L) |
| |
|
| Reliable quantitation limit: |
43 µg/m3 |
| |
|
| Status of method: |
Partially
evaluated method. This method has been subjected to established evaluation
procedures of the Methods Development Team and is presented for information and
trial use. |
| |
|
| March 2004 |
Mary
Eide |
| |
|
Methods Development Team
Industrial Hygiene Chemistry Division
OSHA Salt Lake Technical
Center
Sandy UT 84070-6406
|
1. General Discussion
1.1 Background
1.1.1 History
Air samples were received at OSHA Salt Lake Technical Center
collected on Chromosorb 106 tubes with an analysis request for di(ethylene
glycol) ethyl ether acrylate (DEGEEA). Chromosorb 106 is a nonpolar collection medium, therefore, carbon
disulfide was first selected for the extraction solvent. While the extraction efficiency of dry spiked
samples was 98.2%, spiked samples exposed to humid air had a loss of the DEGEEA
in solution as they sat after shaking, such that the recovery after 6 hours was
83.2%. This indicated that there was a
need for a polar modifier in the extraction solvent. A mixture of 99:1 carbon disulfide:N,N-dimethylformamide
(CS2:DMF) was tried next as the extraction solvent. The average recovery was 98.6%, and there was
no loss of DEGEEA, in humidified samples, as the samples sat on the autosampler. The analyte was found to be well retained on
the Chromosorb 106 tubes, with a retention efficiency recovery of 98.6% and the
storage stability recovery was 97.8% on day 14 of ambient storage.
1.1.2 Toxic
effects (This section is for
information only and should not be taken as the basis of OSHA policy.)1
DEGEEA
is a contact irritant affecting the skin, eyes, upper respiratory tract, and
mucous membrane. DEGEEA may cause
sensitization through skin contact. Prolonged skin exposure may cause defatting and dermatitis.
1.1.3 Workplace exposure2
DEGEEA is used as a reactive multifunctional monomer for UV curing of inks and coatings.
1.1.4 Physical properties and other descriptive information3,4
| synonyms: |
acrylic
acid, 2-(2-ethoxyethoxy)ethyl ester; carbitol acrylate; ethylcarbitol acrylate; ethyldiglycol
acrylate; 2-(2-ethoxyethoxy)ethyl acrylate; ethyl acrylate ester monomer;
2-propenoic acid, 2-(2-ethoxy) ethyl ester; Reactomer RC-20 |
| CAS number: |
7328-17-8 |
| IMIS5 |
D945 |
| molecular weight: |
188.22 |
| density (g/mL): |
1.016 |
| melting point: |
-66 °C |
| boiling point: |
95 °C |
| appearance: |
clear liquid |
| flash point: |
113 °C (235 °F) (closed cup) |
| odor: |
ester-like |
| molecular formula: |
C9H16O4 |
| solubility: |
soluble in most organic solvents |
| autoignition temperature: |
210 °C (410 °F) |
structural formula:

This
method was evaluated according to the OSHA SLTC "Evaluation Guidelines for Air
Sampling Methods Utilizing Chromatographic Analysis"6. The Guidelines define analytical parameters;
specify required laboratory tests, statistical calculations and acceptance
criteria. The analyte air concentrations
throughout this method are based on the recommended sampling and analytical
parameters.
1.2 Detection
limit of the overall procedure (DLOP) and reliable quantitation limit (RQL)
The
DLOP is measured as mass per sample and expressed as equivalent air
concentration, based on the recommended sampling parameters. Ten samplers were spiked with equal
descending increments of analyte, such that the highest sampler loading was 9.14
µg of DEGEEA. This is the amount spiked on
a sampler that would produce a peak at least 10 times the response for a sample
blank. These spiked samplers were
analyzed with the recommended analytical parameters, and the data obtained used
to calculate the required parameters (standard error of estimate and slope) for
the calculation of the DLOP. The slope
was 715 and the SEE was 148. The RQL is
considered the lower limit for precise quantitative measurements. It is determined from the regression line
parameters obtained for the calculation of the DLOP, providing 75% to 125% of
the analyte is recovered. The DLOP and
RQL were 0.62 µg and 2.07 µg, respectively. The recovery at the RQL was 98.5%.
Table 1.2
Detection Limit of the Overall Procedure for DEGEEA
|
mass per sample
(µg) |
area counts
(µV-s) |
|
| 0.00 |
0 |
| 0.91 |
550 |
| 1.83 |
1021 |
| 2.74 |
1635 |
| 3.66 |
2216 |
| 4.57 |
2872 |
| 5.49 |
3640 |
| 6.40 |
4339 |
| 7.32 |
4939 |
| 8.23 |
5720 |
| 9.14 |
6859 |
|

Figure 1.2.1 Plot of data to determine the DLOP/RQL for DEGEEA.
(y = 715x - 226) |
Below is a chromatogram of DEGEEA near the RQL. The recovery at the RQL was 98.5%.

Figure 1.2.2 Chromatogram of the DEGEEA near the RQL. (Key: (1) DEGEEA, (2) interferant) |
2. Sampling Procedure
All safety
practices that apply to the work area being sampled should be followed. The sampling equipment should be attached to
the worker in such a manner that it will not interfere with work performance or
safety.
2.1 Apparatus
2.1.1 Samples
are collected using a personal sampling pump calibrated, with the sampling
device attached, to within ±5% of the recommended flow rate.
2.1.2 Samples
are collected with 7-cm × 4-mm i.d. × 7-mm o.d. glass sampling tubes packed with two
sections (100/50 mg) of Chromosorb 106. The sections are held in place with foam plugs and with a glass wool
plug at the front. For this evaluation,
commercially prepared sampling tubes were purchased from SKC, Inc. (Catalog no.
226-110, lot 2573).
2.2 Reagents
None required.
2.3 Technique
2.3.1 Immediately
before sampling, break off the ends of the flame-sealed tube to provide an
opening approximately half the internal diameter of the tube. Wear eye protection when breaking the
tube. Use tube holders to minimize the
hazard of broken glass. All tubes should
be from the same lot.
2.3.2 The
smaller section of the adsorbent tube is used as a back-up and is positioned
nearest the sampling pump. Attach the
tube holder to the sampling pump so that the adsorbent tube is in an
approximately vertical position with the inlet facing down during
sampling. Position the sampling pump,
tube holder and tubing so they do not impede work performance or safety.
2.3.3 Draw the air to be sampled directly into the inlet of
the tube holder. The air being sampled
is not to be passed through any hose or tubing before entering the sampling
tube.
2.3.4 After sampling for the appropriate time, remove
the adsorbent tube and seal it with plastic end caps. Seal each sample end-to-end with an OSHA-21
form as soon as possible.
2.3.5 Submit
at least one blank sample with each set of samples. Handle the blank sample in the same manner as
the other samples except draw no air through it.
2.3.6 Record
sample air volumes (liters), sampling time (minutes), and sampling rate (L/min)
for each sample, along with any potential interferences on the OSHA-91A form.
2.3.7 Submit
the samples to the laboratory for analysis as soon as possible after
sampling. If delay is unavoidable, store
the samples at refrigerator temperature. Ship any bulk samples separate from the air samples.
2.4 Extraction efficiency
The
extraction efficiency was determined by spiking front sections of Chromosorb
106 tubes with DEGEEA at 0.1 to 2 times the target concentration, based on a
48-L air volume, for a loading of 4.57 to 91.4 µg/sample. These samples
were stored overnight at ambient temperature and then extracted with 1 mL of
extracting solvent for 30 minutes with shaking, and analyzed by GC-FID. The mean extraction efficiency over the
studied range was 98.6%. The wet extraction efficiency was determined at 1
times the target concentration by liquid spiking the analyte onto Chromsorb 106
tubes which had 48-L humid air (absolute humidity of 15.9 mg/L of water, about
80% relative humidity at 22.2 °C) drawn through them immediately before
spiking. The mean recovery for the wet
samples was 98.9%.
Table 2.4
Extraction Efficiency (%) of DEGEEA
|
level
|
sample number
|
mean
|
x target
concn |
µg per
Sample |
1 |
2 |
3 |
4 |
5 |
6 |
|
|
| 0.1 |
4.57 |
97.4 |
97.2 |
98.2 |
98.3 |
97.0 |
97.6 |
97.6 |
| 0.5 |
22.9 |
99.1 |
99.3 |
98.5 |
98.4 |
98.8 |
99.0 |
98.9 |
| 1.0 |
45.7 |
99.1 |
98.5 |
98.9 |
98.6 |
99.2 |
98.8 |
98.9 |
| 1.5 |
68.6 |
99.0 |
98.3 |
99.3 |
98.7 |
99.4 |
97.9 |
98.8 |
| 2.0 |
91.4 |
98.9 |
99.0 |
99.1 |
99.0 |
99.4 |
98.8 |
99.0 |
| |
|
|
|
|
|
|
|
|
| 1.0 (wet) |
45.7 |
98.9 |
99.0 |
98.6 |
98.5 |
99.1 |
99.3 |
98.9 |
|
2.5 Retention efficiency
Six
Chromosorb 106 tubes were spiked with 91.4 µg (1.9 mg/m3) of DEGEEA in the front sections, and
allowed to equilibrate for 6 h. The
tubes had 48-L humid air (absolute humidity of 15.9 mg/L of water, about 80%
relative humidity at 22.2 °C) pulled through them at 0.2 L/min. The samples were extracted and analyzed.
The mean recovery was 98.6%. There was no analyte found on the back-up
section of any of the tubes.
Table 2.5
Retention Efficiency (%) of DEGEEA |
|
section
|
sample number
|
mean
|
| |
1 |
2 |
3 |
4 |
5 |
6 |
|
|
| front of spiked tube |
99.1 |
98.3 |
99.0 |
98.8 |
98.5 |
97.9 |
98.6 |
| rear of spiked tube |
0.0 |
0.0 |
0.0 |
0.0 |
0.0 |
0.0 |
0.0 |
| total |
99.1 |
98.3 |
99.0 |
98.8 |
98.5 |
97.9 |
98.6 |
|
2.6 Sample Storage
Fifteen
Chromosorb 106 tubes were each spiked with 45.7 µg (0.95 mg/m3) of DEGEEA. They were allowed to equilibrate for 6h, then
they had 48 L of air, with an absolute humidity of 15.7 milligrams of water per
liter of air (about 80% relative humidity at 23 °C), drawn through them.
Three samples were analyzed immediately, and the rest were sealed. Six were stored at room temperature (23 °C),
while the other six were stored at refrigerated temperature (4 °C). Three samples stored at room temperature and
three samples stored at refrigerated temperature were analyzed after 7 days and
the remaining three of each set after 14 days. The amounts recovered indicate good storage stability for the time
period studied.
Table 2.6 Storage Test for
DEGEEA |
|
| time (days) |
ambient storage recovery (%) |
refrigerated storage recovery (%) |
|
| 0 |
99.5 |
98.3 |
98.9 |
|
|
|
| 7 |
98.8 |
98.5 |
97.7 |
98.9 |
99.3 |
98.0 |
| 14 |
97.5 |
98.7 |
97.2 |
98.5 |
99.2 |
97.9 |
|
2.7 Recommended air volume and sampling rate
Based
on the data collected in this evaluation, 48-L air samples should be collected
at a sampling rate of 0.2 L/min for 240 minutes.
2.8 Interferences (sampling)
2.8.1 There
are no known compounds which will severely interfere with the collection of DEGEEA.
2.8.2 Suspected
interferences should be reported to the laboratory with submitted samples
3. Analytical Procedure
Adhere
to the rules set down in your Chemical Hygiene Plan. Avoid skin contact and inhalation of all
chemicals and review all appropriate MSDSs.
3.1 Apparatus
3.1.1 A gas
chromatograph equipped with an FID detector. For this evaluation, an Agilent 6890 GC was used.
3.1.2 A GC
column capable of separating DEGEEA from the extraction solvent, internal
standard, and any potential interferences. A 60-m × 0.32-mm i.d. DB-1 (1-µm df) capillary column (J&W Scientific,
Folsom, CA) was used in this evaluation.
3.1.3 An
electronic integrator or some other suitable means of measuring peak
areas. A Waters Millennium32
Data System and an Agilent integrator 3396 were used in this evaluation.
3.1.4 Glass
vials with poly(tetrafluoroethylene)-lined caps. For this evaluation 2-mL vials were used.
3.1.5 A dispenser capable of delivering 1.0 mL of extraction
solvent to prepare standards and samples. If a dispenser is not available, a 1.0-mL volumetric pipet may be used.
3.1.6 Volumetric flasks – 10-mL and other convenient
sizes for preparing standards.
3.1.7 Calibrated 10-µL or 20-µL syringe for preparing standards.
3.1.8 A mechanical shaker. An Eberbach mechanical shaker was used in
this evaluation.
3.2 Reagents
3.2.1 Di(ethylene
glycol) ethyl ether acrylate, technical grade. Aldrich 90% (lot 07619DF) was used in this evaluation.
3.2.2 Carbon
disulfide, Reagent grade. EM Science Omni-Solv® 99.99% (lot 43279343), was used in this
evaluation.
3.2.3 N,N-Dimethylformamide, anhydrous. Aldrich 99.8% (lot 04643LA) was used in this
evaluation.
3.2.4 p-Cymene, reagent grade. Aldrich 99% (lot 11703TR) was used in this
evaluation
3.2.5 The extraction solvent solution was 99:1 carbon disulfide:DMF with 0.25 µL/mL of p-cymene
as internal standard.
3.3 Standard preparation
3.3.1 Prepare
at least two stock standards by spiking microliter quantities of DEGEEA from a
microliter syringe into volumetric flasks containing the extraction solution. Working analytical standards are prepared by
serial dilutions of the stock standard with the extraction solvent. A stock standard of 5
µL/10 mL (0.5 µL/mL) is equivalent to 457 µg/mL,
based on the density and the purity of
90%. A 1:10 dilution of this stock standard is 45.7 µg/mL, which is equivalent to 0.95 mg/m3
based on a 48 L air volume.
3.3.2 Bracket sample concentrations with standard
concentrations. If, upon analysis,
sample concentrations fall outside the range of prepared standards, prepare and
analyze additional standards to confirm instrument response, or dilute high
samples with extraction solvent and reanalyze the diluted samples. For this evaluation, standards in the range
of 0.91 to 183 µg/mL were
used. A check standard from a second
source should be prepared to check the calibration.
3.4 Sample preparation
3.4.1 Remove
the plastic end caps from the sample tubes and carefully transfer each
adsorbent section to separate 2-mL vials. Discard the glass tube, urethane foam plug and glass wool plug.
3.4.2 Add
1.0 mL of extraction solvent to each vial using the same dispenser as used for preparation
of standards.
3.4.3 Immediately seal the vials with poly(tetrafluoroethylene)-lined caps.
3.4.4 Agitate the vials on a shaker or rotator for
30 minutes.
3.5 Analysis
3.5.1 Gas chromatographic conditions
GC conditions
|
| zone temperature: |
initial 50 °C,
hold 1 min, ramp at 10 °C/min to 170°C, hold 7 min |
| |
250 °C (injector) |
| |
250 °C (detector) |
| run time: |
20 min |
| column gas flow: |
3.2 mL/min (hydrogen) |
| injection size: |
1.0 µL (10:1 split) |
| column: |
60-m
× 0.32-mm i.d. capillary DB-1 (df = 1 µm) |
| retention times: |
4.0
min (carbon disulfide); 7.0 min (DMF); 11.8 min (p-cymene); 15.2
min (DEGEEA) |
FID conditions
|
| hydrogen flow: |
30 mL/min |
| air flow: |
400 mL/min |
| makeup flow: |
25 mL/min (nitrogen) |

Figure 3.5.1 A chromatogram of 45.7 µg/mL DEGEEA in 99:1 CS2:DMF with 0.25
µ/mL p-cymene internal
standard.
(Key: (1) CS2; (2)
benzene contaminant in the CS2; (3) DMF; (4) p-cymene; (5) DEGEEA) |
3.5.2 Peak areas are measured by an integrator or other suitable means.
3.5.3 An internal standard (ISTD) calibration method
is used. A calibration curve can be
constructed by plotting response of standard injections versus micrograms of
analyte per sample. Bracket the samples
with freshly prepared analytical standards over the range of concentrations.
Figure 3.5.2 Calibration
curve of DEGEEA. (y = 476x + 388) |
3.6 Interferences (analytical)
3.6.1 Any
compound that produces a GC response and has a similar retention time as the analyte
is a potential interference. If any
potential interferences were reported, they should be considered before samples
are extracted. Generally, chromatographic
conditions can be altered to separate an interference from the analyte.
3.6.2 When
necessary, the identity or purity of an analyte peak may be confirmed by GC-mass
spectrometry or by another analytical procedure. The mass spectrum in Figure 3.6.2 was from an
analytical standard run on an Agilent 6890 GC with a 5973 Mass Selective
Detector.

Figure 3.6.2 Mass spectrum of DEGEEA |
3.7 The
amount of analyte per sampler is obtained from the appropriate calibration
curve in terms of micrograms per sample, uncorrected for extraction
efficiency. This total amount is then
corrected by subtracting the total amount (if any) found on the blank. The air concentration is calculated using the
following formulas.
| where: |
CM is concentration by weight (mg/m3) |
| M is micrograms per sample |
| V is liters of air sampled |
| EE
is extraction efficiency, in decimal form |
4. Recommendations for Further Study
Collection,
reproducibility, and other detection limit studies need to be performed to make
this a fully validated method.
References
1. Material
Safety Data Sheet: Di(ethylene glycol) ethyl ether acrylate, Aldrich Chemical Co., Milwaukee, WI, (accessed 10/3/03).
2. Material
Safety Data Sheet: Diethylene glycol ethyl ether acrylate, Chemwatch, Victoria, Australia, (accessed 12/17/03).
3. Material
Safety Data Sheet: Diethylene glycol ethyl ether acrylate, Aldrich Chemical Co.,
Milwaukee, WI. (accessed 10/03/03).
4. Material
Safety Data Sheet: Ethyldiglycol acrylate, BASF Corporation (accessed 12/17/03).
5.
OSHA Chemical Sampling Information (accessed 12/17/03).
6. Burright,
D.; Chan, Y.; Eide, M.; Elskamp, C.; Hendricks, W.; Rose, M. C. Evaluation
Guidelines for Air Sampling Methods Utilizing Chromatographic Analysis; OSHA Salt Lake Technical Center,
U.S. Department of Labor: Salt Lake City, UT, 1999.
|