1. General Discussion
1.1 Background
1.1.1 History of procedure
This evaluation was undertaken to determine the effectiveness of the OVS-2 tube as a sampling device for carbofuran. It follows the procedure developed for carbaryl. (Ref. 5.1)
1.1.2 Toxic effects (This section is for information only and should not be taken as the basis of OSHA policy).
The toxic effects of carbamate pesticides parallel those of organophosphorus pesticides. Both classes of compounds inhibit cholinesterase, thereby allowing the accumulation of large amounts of acetylcholine. The major difference being that this
inhibition is reversible for carbamates and irreversible for organophosphates. (Ref. 5.2)
The following paragraph describing the results of this cholinesterase inhibition
is excerpted from the book OCCUPATIONAL DISEASES, A Guide To Their Recognition
and is applicable to both carbamates and organophosphates. (Ref. 5.2)
When a critical level of cholinesterase depletion is reached, usually about
20% of normal, symptoms and signs of acetylcholine accumulation poisoning
become manifest. Symptoms may include blurred vision, weakness, nausea, head-
ache, abdominal cramps, chest discomfort, and diarrhea. Signs may include miosis,
muscle twitching, salivation, sweating, tearing, cyanosis, convulsions, and
coma.
Carbamate pesticides can have low oral LD50s, but in general their
dermal LD50s are higher than other cholinesterase inhibiting pesticides, such as
organophosphates.
Carbofuran has an acute oral LD50, 0 of 8 to 9 mg/kg for rats and an
acute dermal LD50 of 10,200 mg/kg for rabbits. (Refs. 5.3-5.4)
In addition, carbofuran is highly toxic by the inhalation route with
significant cholinesterase depression in the Rhesus monkey at 0.86 mg/m3 for a
75% wettable powder. (Ref. 5.3)
Due to these factors carbofuran has been given a TLV-TWA of 0.1 mg/m3
by the ACGIH. (Ref. 5.3)
OSHA adopted this same value as its PEL in March 1989.
Editorial Note: These March 1989 PELs were vacated on July 7, 1992 and ceased to be enforceable on March 23, 1993 (FR 58:35338-35351, 6/30/1993).
1.1.3 Potential workplace exposure
No estimate of worker exposure to carbofuran could be found. Carbofuran is used
as an insecticide. (Ref. 5.4)
1.1.4 Physical properties (Ref. 5.3-5.5)
|
Molecular weight: |
221.26
|
|
Molecular formula: |
C12H15NO3 |
| CAS #: |
1563-66-2 |
| IMIS #: |
0526 |
| Melting point: |
150 to 152°C |
| Vapor Pressure: |
0.0027 Pa (0.00002 mm Hg) at 33°C |
| Appearance: |
white crystalline
solid |
| Solubility: |
700 ppm in water at 25°C
unstable in alcohols |
| Synonyms: |
Bay 70143, Crisfuran, Curaterr, D 1221, ENT 27164, FMC 10242, Furadan, NIA 10242, Pillarfuran, Yalox |
| Chemical name: |
2,3-Dihydro-2,2-dimethyl-7-benzofu- ranyl- methylcarbamate |
| UV spectrum: |
See Figure 1. |
| Structure: |
 |
| |
| |
1.2 Limit defining parameters
The detection limit of the analytical procedure is 3.1 ng per injection. This
is the amount-of analyte which will give a peak whose height is approximately
five times the baseline noise.
2. Sampling Procedure
2.1 Apparatus
2.1.1 A personal sampling pump that can be calibrated to within
±5% of the recommended flow rate with the sampling device in line.
2.1.2 OVS-2 tubes, which are specially made 13-mm o.d. glass tubes that are
tapered to 6-mm o.d. They are packed with a 140-mg backup section and a 270-mg
sampling section of cleaned XAD-2. The backup section is retained by two foam
plugs and the sampling section is between one foam plug and a 13-mm diameter
glass fiber filter. The glass fiber filter is held next to the sampling section
by a poly-tetrafluoroethylene (PTFE) retainer. (See
Figure 2.)
2.2 Reagents
No sampling reagents are required.
2.3 Sampling technique
2.3.1 Attach the small end of the OVS-2 sampling tube to the
sampling pump with flexible, plastic tubing such that the large, front section
of the sampling tube is exposed directly to the atmosphere. Do not place any
tubing in front of the sampler.
2.3.2 Attach the sampler vertically (large end down) in the worker's breathing
zone in such a manner that it does not impede work performance.
2.3.3 After sampling for the appropriate time, remove the sampling device and
seal the tube with plastic end caps.
2.3.4 Wrap each sample end-to-end with an OSHA seal (Form 21).
2.3.5 Submit at least one blank with each set of samples. Handle the blank
the same as the other samples but do not draw air through it.
2.3.6 Submit any bulk samples in a separate container. Do not ship them with
the air samples.
2.4 Extraction and desorption efficiencies
2.4.1 Glass fiber filter
Six 13-mm glass fiber filters were placed in separate
4-mL vials. Five of these filters were each liquid spiked with 26 µL of a 1.84
mg/mL solution of carbofuran
in acetonitrile. These six vials were then sealed with PTFE-lined septa and
stored overnight, in a drawer at room temperature. They were then extracted with
2.0 mL of acetonitrile and analyzed as in Section 3.5.
Table 2.4.1
Glass Fiber Filter Extraction Study
|
|
| Filter # |
Amount Spiked |
Amount Recovered |
% Recovery |
|
| F1 |
47.84 µg |
47.12 µg |
98.5 |
|
| F2 |
47.84 µg |
47.12 µg |
98.5 |
|
| F3 |
47.84 µg |
47.46 µg |
99.2 |
|
| F4 |
47.84 µg |
48.56 µg |
101.5 |
|
| F5 |
47.84 µg |
47.12 µg |
98.5 |
|
| F6 |
0.00 µg |
0.00 µg |
Blank |
|
|
| Average recovery is 99.2% |
2.4.2 XAD-2 adsorbent
An amount of XAD-2 adsorbent equal to the sampling section (270 mg) of an
OVS-2 tube was placed in each of six 4-mL vials which were then sealed with PTFE-lined
septa.
Five of these vials were then each liquid spiked with 26 µL of a 1.84 mg/mL
solution of carbofuran in acetonitrile by injecting the solution onto the
adsorbent through the septum. After replacing the punctured septa, these vials
were allowed to equilibrate overnight in a drawer at room temperature. They were
then desorbed with 2.0 mL of acetonitrile and analyzed as in Section 3.5.
Table 2.4.2
XAD-2 Adsorbent Desorption Study
|
|
| Adsorbent# |
Amount spiked |
Amount recovered |
% Recovery |
|
| AD1 |
47.84 µg |
45.69 µg |
95.5 |
| AD2 |
47.84 µg |
45.69 µg |
95.5 |
| AD3 |
47.84 µg |
45.69 µg |
95.5 |
| AD4 |
47.84 µg |
45.69 µg |
95.5 |
| AD5 |
47.84 µg |
46.40 µg |
97.0 |
| AD6 |
0.00 µg |
0.00 µg |
Blank |
|
| Average recovery is 95.8% |
2.5 Retention efficiency
Six OVS-2 tubes were each liquid spiked with 26 µL of a 1.84 mg/mL solution of
carbofuran in acetonitrile by spiking the glass fiber filter. These tubes were
then sealed with plastic end caps and placed in a drawer at room temperature.
After overnight storage, 480 liters of humid air (approximately 70% relative
humidity) were drawn through each tube. Three of these tubes, along with a blank
tube, were then desorbed and analyzed as in Section 3. No carbofuran was
recovered from the backup section of these tubes.
Table 2.5
Retention Efficiency Study
|
|
| Tube # |
Amount spiked |
Amount recovered |
% Recovery |
|
| RET1 |
47.84 µg |
45.16 µg |
94.4 |
| RET2 |
47.84 µg |
45.64 µg |
95.4 |
| RET3 |
47.84 µg |
45.93 µg |
96.0 |
| RET4 |
0.00µg |
0.00 µg |
Blank |
|
| Average recovery 95.3% |
2.6 Sample storage
The remaining three spiked tubes from Section 2.5 (and a blank tube) were stored
for a total of 7 days in a drawer at room temperature. They were then
desorbed and analyzed as in Section 3. No carbofuran was recovered from the
backup section of these tubes.
Table 2.6
Storage Study
|
|
| Tube # |
Amount spiked |
Amount recovered |
% Recovery |
|
| ST1 |
47.84 µg |
48.51 µg |
101.4 |
|
| ST2 |
47.84 µg |
44.40 µg |
92.8 |
|
| ST3 |
47.84 µg |
46.79 µg |
97.8 |
|
| ST4 |
0.00 µg |
0.00 µg |
Blank |
|
|
| Average recovery is 97.3% |
2.7 Recommended air volume and sampling rate
2.7.1 The recommended air volume is 480 L.
2.7.2 The recommended flow rate is 1.0 L/min.
2.8 Interferences (sampling)
It is not known if any compounds will interfere with the collection of
carbofuran. Suspected interferences should be reported to the laboratory with
submitted samples.
2.9 Safety precautions (sampling)
2.9.1 Attach the sampling equipment in such a manner that it will not interfere
with work performance or employee safety.
2.9.2 Follow all safety practices that apply to the work area being sampled.
3. Analytical Procedure
3.1 Apparatus
3.1.1 An HPLC equipped with a UV detector, and a manual or
automatic injector. A Waters 510 pump, Waters 712 autosampler and Waters
490E UV detector were used in this evaluation.
3.1.2 An HPLC column capable of separating carbofuran from any interferences. A
(25 cm x 4.6 mm i.d.) Chromasil C18 (5 micron) column was used in this
evaluation.
3.1.3 An electronic integrator or other suitable means of measuring detector
response. A Hewlett-Packard 3357 data system was used in this evaluation.
3.1.4 Vials, 4-mL glass with PTFE-lined septa.
3.1.5 Volumetric flasks, pipets,
and syringes.
3.2 Reagents
3.2.1 Acetonitrile, HPLC grade.
3.2.2 Water, HPLC grade. A Millipore Milli-Q system was used
to prepare the water in this evaluation.
3.2.3 Carbofuran. A 99.6% pure standard from EPA was used in this evaluation.
3.3 Standard preparation
Prepare stock, standard solutions by adding acetonitrile to preweighed
amounts of carbofuran. Prepare working range standards by diluting stock
solutions with acetonitrile. Store stock and dilute standards in a freezer.
3.4 Sample preparation
3.4.1 Transfer the 13-mm glass fiber filter and the 270-mg sampling section of
the OVS-2 tube to a 4-mL vial. Place the first foam plug and the 140-mg backup
section in a
separate vial. A small glass funnel can be used to facilitate the transfer of
the adsorbent. Discard the rear foam plug. Do not discard the glass sampling
tube; it
can be reused.
3.4.2 Add 2.0 mL of acetonitrile to each vial.
3.4.3 Seal the vials with PTFE-lined septa and allow them to
desorb for one hour. Shake the vials by hand periodically during the one hour
desorption time.
3.5 Analysis
3.5.1 Liquid chromatographic conditions
| Column: |
25 cm x 4.6 mm i.d. stainless steel
column packed with 5 micron Chromasil C18 |
| Mobile Phase: |
25% acetonitrile / 75% water |
| Flow rate: |
1 mL/min |
| UV detector |
210 nm |
| Retention time: |
13 min |
| Injection volume: |
10 µL |
3.5.2 Chromatogram (See
Figure 3.)
3.6 Interferences (analytical)
3.6.1 Any compound having a retention time similar to that of the analyte is a
potential interference. Generally, chromatographic conditions can be altered to
separate interferences from the analyte.
3.6.2 Retention time on a single column is not proof of chemical identity.
Analysis by an alternate HPLC column, detection at another wavelength (for
comparison of absorbance response ratios) and confirmation by mass spectrometry
are additional means of identification.
3.7 Calculations
3.7.1 Construct a calibration curve by plotting detector response versus
standard concentration.
3.7.2 Determine the concentration of carbofuran in each sample from the
calibration curve. If carbofuran is found on the backup section, make blank
corrections for each section separately before adding the results together.
3.7.3 Determine the air concentration by the following formula.
| mg/m3 = |
(µg/mL in sample) x (desorption volume, mL)
(air volume, L) x (desorption efficiency, decimal) |
3.8 Safety precautions
(analytical)
3.8.1 Avoid exposure to all standards.
3.8.2 Avoid exposure to all solvents.
3.8.3 Wear safety glasses at all times.
4. Recommendations for Further Study
4.1 Even though the UV spectrum indicates higher absorbance at either
280 nm or 230 nm (See Figure 1), 210 nm was a more sensitive wavelength for the
analysis of carbofuran. This behavior should be investigated further. For
example, analysis on a diode array UV detector to determine the most sensitive
analytical wavelength.
4.2 A desorption study determining the recovery from a 13-mm glass fiber filter
in combination with 270 mg of XAD-2 should be done. This is the desorption
efficiency used in Section 3.7.3.
4.3 The literature indicates that carbofuran is unstable in alcohols. It was
observed in the HPLC analysis of a carbofuran standard in acetonitrile that the
analyte peak got smaller and a secondary peak grew in size. This did not occur
until after several weeks at room temperature. Carbofuran appears stable enough
in acetonitrile but this slow decomposition with time should be kept in
mind.
4.4 This method should be fully validated.

Figure 1. UV Spectrum of Carbofuran in the HPLC Mobile Phase. |
Figure 2. OVS-2 Sampling Device |

Figure 3. Chromatogram of Carbofuran |
5. References
5.1. Burright, D.; Method #63, "Carbaryl"; OSHA Analytical Laboratory, unpublished, 1987.
5.2. "OCCUPATIONAL DISEASES, A Guide to their Recognition"; U.S. Department of Health, Education, and Welfare; Public Health Service,
Public Health Service Publication No. 1097, U.S. Government Printing Office: Washington, D.C., 1964; p 245.
5.3. "Documentation of the Threshold Limit Values and Biological Exposure Indices", 5th ed.; American Conference of Governmental
Industrial Hygienists: Cincinnati, OH, 1986; p 100.
5.4. "Farm Chemicals Handbook"; Meister Publishing Co.: Willoughby, OR, 1986; p C48.
5.5. Windholz, M., Ed.; "Merck Index", 10th ed.; Merck and Co.: Rahway, NJ, 1983; p 250.
|