| Method no.: |
PV2071 |
| |
|
| Control no.: |
T-PV2071-01-8901-CH |
| |
|
| Matrix: |
Air |
| |
|
| Target concentration: |
0.5 mg/m3 (arbitrary) There is no OSHA permissible exposure level (PEL) or ACGIH threshold limit value (TLV) for pirimiphos-methyl. |
| |
|
| Procedure: |
Samples are collected by drawing known volumes of air through OSHA versatile sampler (OVS-2) tubes, each containing a glass fiber filter and two sections of XAD-2 adsorbent. Samples are desorbed with acetonitrile and analyzed by gas chromatography (GC) using an electron capture detector (ECD). |
| |
|
| Recommended air volume and sampling rate: |
120 L at 1.0 L/min |
| |
|
| Detection limit of the overall procedure (based on the recommended air volume and the analytical detection limit): |
13 μg/m3 |
| |
|
| Status of method: |
Partially validated method. This method has been partially evaluated and is presented for information and trial use only. |
| |
|
| Date: January 1989 |
Chemist: Duane Lee |
| |
|
Carcinogen And Pesticide Branch
OSHA Analytical Laboratory
Salt Lake City, Utah |
1. General Discussion
1.1 Background
1.1.1 History of procedure
The OSHA Analytical Laboratory received a set of samples requesting
the analysis of pirimiphos-methyl and other pesticides. The samples had
been collected on OVS-2 tubes. This report describes the analytical
method developed.
1.1.2 Toxic effects (This section is for information only and should
not be taken as the basis of OSHA policy.)
The oral LD50 for rats is 1250 mg/kg. ( Ref. 5.1) The
acute dermal LD50 for female rats is greater than 4592 mg/kg.
(Ref. 5.2)
1.1.3 Potential workplace exposure
Pirimiphos-methyl is a fast acting and broad-spectrum insecticide.
No estimate of worker exposure to pirimiphos-methyl could be found.
(Ref. 5.1)
1.1.4 Physical properties (Ref. 5.1 to 5.3)
| Molecular weight: |
305.37 |
| Molecular formula: |
C11H20N3O3PS |
| CAS number: |
29232-93-7 |
|
IMIS number:
|
P309 |
| Specific gravity: |
1.157 |
| Vapor pressure: |
0.015 Pa (1.1×10-4 mmHg) at 30°C |
| Solubility: |
miscible with most organic solvents; in water, 5
mg/L at 30°C
|
| Chemical name: |
O-(2-diethylamino-6-methylpyrimidin-4-yl)O,O-dimethyl
phosphorothioate |
| Synonyms: |
Actelic, Actellic, Actellifog,
Blex, PP-511, Silosan, Plant Protection
PP511, Pyridimine Phosphate |
| Description: |
straw colored liquid |
| Structure: |
 |
1.2 Limit defining parameters
The detection limit of the analytical procedure,
including a 15:1 split ratio, is 0.04 ng per injection. This is the amount
of analyte which will give a peak whose height is approximately five times
the baseline noise.
2. Sampling procedure
2.1 Apparatus
2.1.1 A personal sampling pump that can be calibrated to within ± 5%
of the recommended flow rate with the sampling device in line.
2.1.2 OVS-2 tubes, which are specially made 13 mm o.d. glass tubes that are tapered to 6 mm
o.d., packed with a 140-mg backup section, a 270-mg sampling section of cleaned XAD-2
adsorbent and a 13 mm diameter glass fiber filter. The backup section is retained by two foam
plugs and the sampling section is between one foam plug and the glass fiber filter. The glass
fiber filter is held next to the sampling section by a polytetrafluoroethylene (PTFE) retainer.

Figure 2.1.2
OVS-2 Sampling Tube
|
2.2 Reagents
No sampling reagents are required.
2.3 Sampling technique
2.3.1 Immediately before sampling, remove the plastic caps
from the OVS-2 tube.
2.3.2 Attach the small end of the tube to the sampling pump
with flexible tubing.
2.3.3 Attach the tube vertically in the employee's
breathing zone in such a manner that it does not impede work
performance.
2.3.4 After sampling for the appropriate time, remove the
tube and seal with plastic caps.
2.3.5 Wrap each sample end-to-end with an OSHA seal (Form
21).
2.3.6 Record the air volume for each sample, and list any
possible interferences.
2.3.7 Submit at least one blank for each set of samples.
Handle the blank in the same manner as the samples, except no
air is drawn through it.
2.3.8 Submit bulk samples for analysis in a separate
container.
2.4 Desorption efficiency (glass fiber filter and XAD-2 adsorbent)
Six vials each containing a 13-mm
glass fiber filter and 270-mg of XAD-2 adsorbent were each liquid
spiked on the glass fiber filter with 74.57 μg
of pirimiphos-methyl and allowed to dry for 2 hours. These samples
were each desorbed with 3 mL of acetonitrile, shaken for 30 min
and analyzed as in Section 3. The results are listed in the Table
below. |
|
Table 2.4
Desorption Efficiency
|
amount
spiked,
μg |
amount
found,
μg |
%
recovered |
|
74.57
74.57
74.57
74.57
74.57
74.57 |
69.74
66.56
63.42
65.66
71.57
80.32 |
93.5
89.3
85.0
88.1
96.0
07.8 |
| |
 |
93.3 |
|
|
2.5 Retention efficiency
Eighteen
OVS-2 tubes were each liquid spiked with 74.57 μg
of pirimiphos-methyl on the glass fiber filter. These were allowed
to dry and then 240 L of humid air (~80%
relative humidity) were drawn through each tube at 1 L/min. Six of
the tubes were each desorbed with 3 mL of acetonitrile, shaken for
30 min and then analyzed as in Section 3. The results are listed
in the Table below. The remaining samples were stored, 6 in a
drawer at ambient temperature and 6 in a freezer.
|
|
Table 2.5
Retention Efficiency
|
amount
spiked,
μg |
amount
found,
μg |
%
recovered |
|
74.57
74.57
74.57
74.57
74.57
74.57 |
73.25
67.58
66.17
65.19
75.58
77.21 |
98.2
90.6
88.7
87.4
100.0
103.5 |
| |
 |
94.7 |
|
|
2.6 Sample storage
After 4 days of storage, 6 tubes, 3 from the ambient storage group
and 3 from the freezer storage group, were each desorbed with 3 mL of
acetonitrile, shaken for 30 min and then analyzed as in
Section 3. The remaining tubes were
desorbed and analyzed after 7 days of storage. The results are given in
the Tables below.
Table 2.6.1
Ambient Storage
|
days
stored |
amount
spiked,
μg |
amount
found,
μg
|
%
recovered |
|
4
4
4
7
7
7 |
74.57
74.57
74.57
74.57
74.57
74.57 |
69.68
67.76
72.34
71.12
67.66
72.53 |
93.4
90.9
97.0
95.4
90.7
97.3 |
| |
|
of 4
of 7 |
93.8
94.5 |
|
|
|
Table 2.6.2
Freezer Storage
|
days
stored
|
amount
spiked,
μg |
amount
found,
μg
|
%
recovered |
|
4
4
4
7
7
7 |
74.57
74.57
74.57
74.57
74.57
74.57 |
68.31
74.22
74.00
63.91
68.25
69.10 |
91.6
99.5
99.2
85.7
91.5
92.7
|
| |
|
of 4
of 7 |
96.8
90.0 |
|
|
2.7 Recommended air volume and sampling rate
2.7.1 The recommended air volume is 120 L.
2.7.2 The recommended flow rate is 1.0 L/min.
2.8 Interferences (sampling)
It is not known if any compounds will interfere with the
collection of pirimiphos-methyl. Any suspected interference should
be reported to the laboratory.
2.9 Safety precautions (sampling)
2.9.1 Attach the sampling equipment in such a manner that
it will not interfere with work performance or employee
safety.
2.9.2 Follow all safety practices that apply to the work
area being sampled.
3. Analytical procedure
3.1 Apparatus
3.1.1 A balance capable of weighing to the nearest tenth of
a milligram. A Mettler HL52 balance was used in this
evaluation.
3.1.2 A mechanical shaker.
3.1.3 A GC equipped with an ECD. A Hewlett Packard (HP)
5890 was used in this evaluation.
3.1.4 A GC column capable of separating pirimiphos-methyl
from any interferences. A 15 m × 0.32 mm i.d. (1.0 μm
film) DB-5 capillary column was used in this evaluation.
3.1.5 An electronic integrator, or some other suitable
means for measuring detector response. The Hewlett-Packard
3357 Laboratory Data System was used in this evaluation.
3.1.6 Volumetric flasks and pipets.
3.1.7 Vials, 2-mL.
3.2 Reagents
3.2.1 Acetonitrile, reagent grade.
3.2.2 Pirimiphos-methyl, reagent grade. A standard obtained
from EPA (EPA # 5643, 100% purity) was used in this
evaluation.
3.3 Standard preparation
Prepare pirimiphos-methyl stock standards by weighing 10 to
15 mg of pirimiphos-methyl. Transfer the pirimiphos-methyl to
separate 10-mL volumetric flasks, and add acetonitrile to the
mark. Make working range standards of 0.5 to 50 μg/mL
by pipet dilutions of the stock standards with acetonitrile.
Store stock and dilute standards in a freezer.
3.4 Sample preparation
3.4.1 Transfer the 13-mm glass fiber filter and the
270-mg sampling section of the tube to a 4-mL vial. Place
the first foam plug and the 140-mg section in a separate
4-mL vial. A small glass funnel can be used to facilitate
the transfer of the adsorbent. Discard the rear foam plug.
Do not discard the glass sampling tube; it can be reused.
3.4.2 Add 3.0 mL of acetonitrile to each vial and seal
with a Teflon-lined cap.
3.4.3 Shake the vials for 30 minutes on a mechanical shaker.
3.4.4 Transfer the samples to 2-mL vials for use on an HP
autosampler.
3.5. Analysis
3.5.1 Instrument conditions
| Column: |
DB-5, 15 m × 0.32 mm i.d., 1.0-μm film |
| Injector temperature: |
275°C |
| Column temperature: |
220°C |
| Detector temperature: |
300°C |
| Gas flows: |
|
| Column: |
4 mL/min hydrogen |
| ECD make up: |
42 mL/min nitrogen |
| Injection volume: |
1.0 μL |
| Split ratio: |
15:1 |
| Retention time: |
3.06 min |
3.5.2 Chromatogram

Figure 3.5.2
Chromatogram of Pirimiphos-methyl |
3.6 Interferences (analytical)
3.6.1 Any collected compound having a similar
retention time to that of the analyte is a potential
interference.
3.6.2 GC conditions may be varied to circumvent
interferences.
3.6.3 Retention time on a single column is not proof of
chemical identity. Analysis by an alternate GC column,
detection on a flame photometeric detector and
confirmation by mass spectrometry are additional means of
identification.
3.7 Calculations
3.7.1 Construct a calibration curve by plotting detector
response versus concentration (μg/mL)
of pirimiphos-methyl.
3.7.2 Determine the μg/mL
of pirimiphos-methyl in both sections of each sample and blank
from the calibration curve.
3.7.3 Blank correct each section by subtracting the μg/mL
found in the blank section from the μg/mL
found in the sample section and then add the sections together.
3.7.4 Determine the air concentration by using the following formula.
|

Figure 3.7.1
Calibration Curve |
| mg/m3 = |
(mg/mL blank corrected)(desorption volume, mL)
(air volume, L)(desorption efficiency, decimal) |
3.8 Safety precautions (analytical)
3.8.1 Avoid skin contact and air exposure to pirimiphos-methyl.
3.8.2 Avoid skin contact with all solvents.
3.8.3 Wear safety glasses at all times.
4. Recommendation for further study
This method should be fully validated.
5. References
5.1 Farm Chemicals Handbook; Berg, Gordon L. Ed.; Meister: Willoughby,
Ohio, 1986; p C6.
5.2 Registry of Toxic Effects of Chemical Substances 1985-86 Edition; U.S. Department of Health
and Human Services: Cincinnati, OH, 1987; DHHS(NIOSH) Publication No. 87-114, p 3425.
5.3 Merck Index, 10th ed.; Windholz, Martha ED.; Merck: Rathway,
N.J., 1983; p 1082.
|