2-ETHYLHEXYL ACRYLATE
| Method number: |
PV2026 |
| |
| Matrix: |
Air |
| |
| Target concentration: |
5 ppm (38 mg/m3) (Arbitrary). There is no OSHA Permissible exposure level (PEL) or ACGIH Threshold limit value (TLV) for 2-Ethylhexyl acrylate. |
| |
| Procedure: |
Samples are collected by drawing a known volume of air through a Coated CT(coconut shell charcoal coated with 4-tert-butylcatechol-TBC) tube. Samples are desorbed with 1 mL of carbon disulfide for 30 minutes with shaking on a rotator and analyzed by gas chromatography using a flame ionization detector. |
| |
Recommended air volume and sampling rate: |
12 L at 0.1 L/min |
| |
| Reliable quantitation limit: |
0.01 ppm (0.075 mg/m3) |
| |
| Status of method: |
Partially Evaluated Method. This method has been subjected to established evaluation procedures, and is presented for information and trial use. |
| |
| Date: January 1995 |
Chemist: Yogi Shah |
Organic Service Branch I
OSHA Salt Lake Technical Center
Salt Lake City, UT 84165-0200
1. General Discussion
1.1 Background
1.1.1 History
OSHA method 92 uses TBC coated CT for ethyl acrylate and methyl acrylate, therefore it
was decided to try TBC coated CT for 2-ethylhexyl acrylate (EHA).
1.1.2 Toxic effects (This section is for information only and should not be taken as the basis of
OSHA policy.) (Ref. 5.4)
Inhalation of EHA vapors can irritate the nose, throat and lungs. Vapor concentrations of
25 ppm may not be tolerated for more than a few minutes. Higher concentration may
cause drowsiness, dizziness, tiredness, headache, nausea, difficulty in breathing and
convulsions. Vapors can be irritating to the eyes and cause tears. Liquid can cause severe
burns of the eyes. Skin contact by pure liquid or concentrated solution of EHA can cause
severe irritation and burns. There is no OSHA PEL, but toxic effects were identical to
ethyl acrylate and methyl acrylate and also manufacturer recommended 5 ppm limit to the
analyte, there for it was selected. This is a recommendation only this does not reflect OSHA policy.
1.1.3 Workplace exposure (Ref. 5.2, 5.3)
Monomer is mainly used for plastics, protective coatings, paper treatment and water
based paints.
1.1.4 Physical properties and other descriptive information (Ref. 5.2 and 5.4)
| Synonyms: |
2-ethylhexyl-2-propeonate; octyl acrylate |
| CAS number: |
103-11-7 |
| IMIS: |
1055 |
| RTECS: |
AT0855000; 4244 |
| Molecular weight: |
184 |
| Flash point: |
82.2°C |
| Boiling point: |
213.5°C |
| Freezing point: |
-90°C |
| Odor: |
acrid, musty |
| Color: |
colorless |
| Density: |
0.8865 |
| Molecular formula: |
C11H20O2 |
| Structural formula: |
CH2CHCOOCH2CH(C2H5)C4H9 |
The analyte air concentrations throughout this method are based on the recommended sampling and analytical
parameters. Air concentrations listed in ppm are referenced to 25°C and 101.3 kPa (760 mmHg).
1.2 Limit defining parameters
1.2.1 Detection limit of the overall procedure (DLOP)
The detection limit of the overall procedure is 0.31 µg per sample (0.025 mg/m3). This is
the amount of analyte spiked on the sampler that will give a response that is significantly
different from the background response of a sampler blank.
The DLOP is defined as the concentration of analyte that gives a response (YDLOP) that is
significantly different (three standard deviations (SDBR)) from the background response
(YBR).
YDLOP - YBR = 3(SDBR)
The direct measurement of YBR and SDBR in chromatographic methods is typically
inconvenient, and difficult because YBR is usually extremely low. Estimates of these
parameters can be made with data obtained from the analysis of a series of samples whose
responses are in the vicinity of the background response. The regression curve obtained
for a plot of instrument response versus concentration of analyte will usually be linear.
Assuming SDBR and the precision of data about the curve are similar, the standard error of
estimate (SEE) for the regression curve can be substituted for SDBR in the above equation.
The following calculations derive a formula for the DLOP:
| Yobs |
= |
observed response |
| Yest |
= |
estimated response from regression curve |
| n |
= |
total no. of data points |
| k |
= |
2 for a linear regression curve |
At point YDLOP on the regression curve
| YDLOP = A(DLOP) + YBR |
A = analytical sensitivity (slope) |
therefore
Substituting 3(SEE) + YBR for YDLOP gives
The DLOP is measured as mass per sample and expressed as equivalent air concentrations, based on the
recommended sampling parameters. Ten samplers were spiked with equal descending increments of
analyte, such that the highest sampler loading was 10 µg/sample. This is the amount, when spiked on a
sampler, that would produce a peak approximately 10 times the background response for the sample blank.
These spiked samplers, and the sample blank were analyzed with the recommended analytical parameters,
and the data obtained used to calculate the required parameters (A and SEE) for the calculation of the
DLOP. DLOP was calculated to be 0.31 µg/sample (0.025 mg/m3).
Table 1.2.1 Detection Limit of the Overall Procedure
|
| mass per sample |
area counts |
| (µg) |
(µV-s) |
|
| 0 |
0 |
| 1 |
1223 |
| 2 |
2408 |
| 3 |
3275 |
| 4 |
4398 |
| 5 |
5420 |
| 6 |
6488 |
| 7 |
7662 |
| 8 |
8644 |
| 9 |
9616 |
| 10 |
10717 |
|
Figure 1.2.1. Plot of data to determine the DLOP/RQL.
1.2.2 Reliable quantitation limit (RQL)
The reliable quantitation limit is 0.9 µg per sample . This is the amount of analyte spiked
on a sampler that will give a signal that is considered the lower limit for precise quantitative
measurements.
The RQL is considered the lower limit
for precise quantitative
measurements. It is determined from
the regression line data obtained for
the calculation of the DLOP (Section
1.2.1), providing at least 75% of the
analyte is recovered. The RQL is
defined as the concentration of
analyte that gives a response (YRQL)
such that
YRQL - YBR = 10(SDBR)
therefore
The RQL is the lowest loading at which 75% of the analyte can be recovered as determined
from the regression line of the plotted data.
RQL = 0.9 µg per sample (0.075 mg/m3)
The recovery at this concentration was 94%.
Figure 1.2.3. Chromatogram of the RQL.
2. Sampling Procedure
2.1 Apparatus
2.1.1 Samples are collected using a personal sampling pump calibrated, with the sampling device attached, to within ±5% of the recommended flow rate.
Samples are collected with 4 mm i.d. × 6 mm o.d. glass sampling tubes packed with two
sections of coconut shell charcoal that has been coated with TBC, 10% by weight. The front
section contains 110 mg and the back section contains 55 mg of TBC-coated coconut
shell charcoal. The sections are held in place with glass wool plugs and are separated by
a urethane foam plug. For this evaluation, commercially prepared sampling tubes were
purchased from SKC, inc (Catalog No. 226-73).
2.2 Technique
2.2.1 Immediately before sampling, break off the ends of the sampling tube. All tubes should be
from the same lot.
2.2.2 Attach the sampling tube to the pump with flexible tubing. It is desirable to utilize sampling
tube holders which have a protective cover to shield the employee from the sharp, jagged
end of the sampling tube. Position the tube so that sampled air passes through the front
section of the tube first.
2.2.3 Air being sampled should not pass through any hose or tubing before entering the sampling
tube.
2.2.4 Attach the sampler vertically with the front section pointing downward, in the worker's
breathing zone, and positioned so it does not impede work performance or safety.
2.2.5 After sampling for the appropriate time, remove the sample and seal the tube with plastic
end caps. Wrap each sample end-to-end with a Form OSHA-21 seal.
2.2.6 Submit at least one blank sample with each set of samples. Handle the blank sampler in
the same manner as the other samples except draw no air through it.
2.2.7 Record sample volumes (in liters of air) for each sample, along with any potential
interferences.
2.2.8 Ship any bulk samples separate from the air samples.
2.2.9 Submit the samples to the laboratory for analysis as soon as possible after sampling. If
delay is unavoidable, store the samples in a refrigerator.
2.3 Desorption efficiency
The desorption efficiencies of 2-ethylhexyl acrylate were determined by liquid-spiking the TBC
coated charcoal tubes with the analytes at 0.1 to 2 times the target concentration. The loadings on
the tubes were 46.1, 230.5, 461 and 922 mg of 2-ethylhexyl acrylate. These samples were stored
overnight at ambient temperature and then desorbed and analyzed. The average desorption
efficiency over the studied range was 100%.
Table 2.3 Desorption Efficiency of 2 ethylhexyl acrylate
|
|
% Recovered |
|
0.1× |
0.5× |
1.0× |
2.0× |
| Tube # |
µg |
µg |
µg |
µg |
|
| 1 |
102.8 |
103.6 |
100.1 |
99.9 |
| 2 |
100.0 |
103.4 |
99.0 |
99.6 |
| 3 |
99.1 |
102.6 |
98.4 |
96.9 |
| 4 |
102.2 |
99.7 |
98.6 |
100.3 |
| 5 |
101.9 |
102.1 |
99.9 |
100.7 |
| 6 |
103.1 |
102.5 |
99.3 |
99.0 |
| average |
101.5 |
102.3 |
99.2 |
99.4 |
| overall average |
100.0. |
|
| SD |
±1.7 |
|
|
2.4 Retention efficiency
The sampling tubes were spiked with 922 µg (10 ppm) 2-ethylhexyl acrylate, allowed to equilibrate
for 24 hours at room temperature, and then had 12 L humid air (80% RH at 22°C) pulled through
them at 0.1 Lpm. They were opened, desorbed, and analyzed by GC-FID. The retention efficiency
averaged 99.1%. There was no 2-ethylhexyl acrylate found on the back sections of the tubes.
Table 2.4 Retention Efficiency of 2 ethylhexyl acrylate
|
| Tube # |
% Recovered |
|
Front section |
Back section |
Total |
|
| 1 |
99.3. |
0.0 |
99.3 |
| 2 |
.98.1 |
0.0 |
98.1 |
| 3 |
100.1 |
0.0 |
100.1 |
| 4 |
99.2 |
0.0 |
99.2 |
| 5 |
97.7 |
0.0 |
97.7 |
| 6 |
100.1 |
0.0 |
100.1 |
|
|
average |
99.1 |
|
2.5 Sample storage
The front sections of six sampling tubes were each spiked with 461 µg (5 ppm) of 2-ethylhexyl
acrylate. Six more tubes had 12 liters of humid air (80% RH at 22°C) drawn through them before
they were spiked with 461 µg (5 ppm) of 2-ethylhexyl acrylate. They were sealed and stored at
room temperature. Three of each type of samples were analyzed after 7 days and the remaining
three samples of each type after 14 days. The amounts recovered indicate good storage stability
for the time petiod studied and had an average recovery of 98.8%.
Table 2.5 Storage Test for _______
|
Time (days) |
Recovery Humid (%) |
Recovery Dry (%) |
|
| 7 |
98.2 |
99.4 |
| 7 |
99.6 |
101.3 |
| 7 |
100.9 |
98.2 |
| 14 |
97.5 |
97.7 |
| 14 |
96.0 |
97.9 |
| 14 |
99.6 |
98.2 |
|
| average |
98.6 |
98.9. |
|
2.6 Recommended air volume and sampling rate.
Based on the data collected in this evaluation, 12 L air samples should be collected at a sampling
rate of 0.1 L/min.
2.7 Interferences (sampling)
2.7.1 It is not known if any compounds will severely interfere with the collection of 2-ethylhexyl
acrylate on the sampling tubes. In general, the presence of other contaminant vapors in
the air will reduce the capacity of the TBC coated charcoal tube to collect 2-ethylhexyl
acrylate.
2.7.2 Suspected interferences should be reported to the laboratory with submitted samples.
2.8 Safety precautions (sampling)
2.8.1 Attach the sampling equipment to the worker in such a manner that it will not interfere with work performance or safety.
2.8.2 Follow all safety practices that apply to the work area being sampled.
2.8.3 Wear eye protection when breaking the ends of the glass sampling tubes.
3. Analytical Procedure
3.1 Apparatus
3.1.1 The instrument used in this study was a gas chromatograph equipped with a flame
ionization detector (FID), specifically a Hewlett-Packard model 5890.
3.1.2 A GC column capable of separating the analyte from any interferences. The column used
in this study was a SB-1, 30 m, 0.32 mm, 4 mm.
3.1.3 An electronic integrator or some suitable method of measuring peak areas.
3.1.4 Two milliliter vials with TeflonTM-lined caps.
3.1.5 A 10 µL syringe or other convenient size for sample injection.
3.1.6 Pipets for dispensing the desorbing solution. A 1 ml dispenser was used in this study.
3.1.7 Volumetric flasks - 5 or 10 mL and other convenient sizes for preparing standards.
3.2 Reagents
3.2.1 GC grade nitrogen, hydrogen, and air.
3.2.2 2-ethylhexyl acrylate, Reagent grade
3.2.3 carbon disulfide (cs2) (desorbing solvent)
3.2.4 n-hexyl benzene (internal standard)
3.2.5 Desorbing solution was 1 ml with 0.25 µL/mL n-_hexyl benzene as internal standard.
3.3 Standard preparation
3.3.1 At least two separate stock standards are prepared by diluting a known quantity of
ethylhexyl acrylate with the desorbing solution of carbon disulfide. The concentration of these stock standards was 1 µL/mL or 886.5 µg/mL.
3.3.2 A third standard at a higher concentration was prepared to check the linearity of the
calibration. For this study, two analytical standards were prepared at a concentration of 1
µL/mL (886.5 µg/mL), and one at 2 µL/mL (1773 µg/mL) 2-ethylhexyl acrylate in the
desorbing solution.
3.4 Sample preparation
3.4.1 Sample tubes are opened and the front and back section of each tube are placed in
separate 2 mL vials.
3.4.2 Each section is desorbed with 1 mL of the desorbing solution of carbon disulfide with 0.25
µL/mL n-hexyl benzene internal standard.
3.4.3 The vials are sealed immediately and allowed to desorb for 30 minutes with constant
shaking on rotator.
3.5 Analysis
3.5.1 Gas chromatography conditions.
{insert chromatogram here}
| Injection size: |
1 µL |
|
| Flow rates (mL/min) |
|
| Hydrogen(carrier): |
1.5 |
| Hydrogen(detector): |
30 |
| Air: |
365 |
|
| Temperatures (°C) |
|
| Injector: |
200 |
| Detector: |
240 |
| Column: |
SB-1, 0.32 mm, 4 mm, 30 mts |
3.5.2 Peak areas are measured by an integrator or other suitable means.
3.6 Interferences (analytical)
3.6.1 Any compound that produces a response and has a similar retention time as the analyte
is a potential interference. If any potential interferences were reported, they should be
considered before samples are desorbed. Generally, chromatographic conditions can be
altered to separate an interference from the analyte.
3.6.2 When necessary, the identity or purity of an analyte peak may be confirmed by GC-mass
spectrometer or by another analytical
procedure.
3.7 Calculations
3.7.1 The instrument was calibrated with a standard of 461 µg/mL 2-ethylhexyl acrylate in the
desorbing solution. The linearity of the calibration was checked with a standard of 1773
µg/mL (10 ppm).
3.7.2 If the calibration is non-linear, two or more standard at different concentrations must be
analyzed, bracketing the samples, so a calibration curve can be plotted and sample values
obtained.
3.7.3 To calculate the concentration of analyte in the air sample the following formulas are used:
(µg/m)(desorption volume) (desorption efficiency) |
= mass of analyte in sample |
(mass of analyte in sample) molecular weight |
= number of moles of analyte |
| (number of moles of analyte) |
(molar volume at 25°C & 760mm) |
= |
volume the analyte will occupy at 25°C & 760mm |
(volume analyte occupies)(106) (air volume) |
= ppm |
* All units must cancel.
3.7.4 The above equations can be consolidated to the following formula.
(µg/mL)(DV)(24.45)(106) (100 L)(DE)(MW) |
× |
(g) (1000 mg) |
× |
(mg) (1000 µg) |
= ppm |
| µg/mL |
= |
concentration of analyte in sample or standard |
| 24.45 |
= |
Molar volume (liters/mole) at 25°C and 760 mm |
| MW |
= |
Molecular weight (g/mole) |
| DV |
= |
Desorption volume |
| 10 L |
= |
10 liter air sample |
| DE |
= |
Desorption efficiency |
3.7.5 This calculation is done for each section of the sampling tube and the results added
together.
3.8 Safety precautions (analytical)
3.8.1 Avoid skin contact and inhalation of all chemicals.
3.8.2 Wear safety glasses, gloves and a lab coat at all times while in the laboratory areas.
4. Recommendations for Further Study
Collection studies need to be performed from a dynamically generated test atmosphere.
5. References (THESE ARE EXAMPLES AND ACTUAL REFERENCES MAY BE DIFFERENT.)
5.1 Lewis, R., "Hawley's Condensed Chemical Dictionary", Twelfth Edition, Van Nostrand Reinhold
Co. NewYork, 1993, p.338.
5.2 Windholz, M., "The Merck Index", Eleventh Edition, Merck & Co., Rahway N.J., 1989, p. 427.
5.3 "Documentation of the Threshold Limit Values and Biological Exposure Indices", Fifth Edtion,
American Conference of Governmental Industrial Hygienists Inc., Cincinnati, OH, 1986, p.161.
5.4 Occupation Health Services, Material Safety Data Sheets, New York, NY 10036.
Page last updated: 03/29/2010
|