1.1. Background
1.1.1. History of procedure
The OSHA Analytical Laboratory received a set of samples requesting the analysis of
bendiocarb along with other pesticides. The samples had been collected on OVS-2 tubes.
This report describes the analytical method developed for bendiocarb.
1.1.2. Toxic effects (This section is for information only and should not be taken as
the basis of OSHA policy.)
The oral LD50 of bendiocarb is 35-100 mg/kg in mammals. (Ref.
5.3.) From another scource, the oral LD50 for rats is 40-156
mg/kg depending on strain. (Ref. 5.2.)
1.1.3. Potential workplace exposure
Bendiocarb is used as an insecticide. (Ref. 5.2.) There was no information available on
the number of workers potentially exposed to bendiocarb.
1.1.4. Physical properties (Ref. 5.1. to 5.3.)
| CAS number: |
22781-23-3 |
| |
| IMIS number: |
F108 |
| |
| Molecular weight: |
223.23 |
| |
| Molecular formula: |
C11H13NO4 |
| |
| Vapor pressure: |
5.0 × 10-6 mm Hg at 25°C |
| |
| Melting point: |
129-130°C |
| |
| Solubility: |
Soluble in acetonitrile (observed); Soluble in hexane 350 ppm; soluble in
water 40 ppm. |
| |
| Chemical name: |
2,2-Dimethyl-1,3-benzodioxol-4-yl methylcarbamate |
| |
| Synonyms: |
Bencarbate; 1,3-bezodioxole; 2,2-dimethyl-4-(N-methylcarbamato)-; Dycarb;
Ficam; Ficam D; Ficam ULV; Garvox;. Multamat; Multimet; Niomil; Ratate; Tattoo; Turcam. |
| |
| Description: |
white crystalline solid |
| |
| UV Scan: |
Figure 1. |
| |
| Structure: |
 |
1.2. Limit defining parameters
The detection limit of the analytical procedure is 2.4 ng per injection. This is the
amount of analyte which will give a peak whose height is approximately five times the
baseline noise.
2. Sampling Procedure
2.1. Apparatus
2.1.1. A personal sampling pump that can be calibrated to within ± 5% of the
recommended flow rate with the sampling device in line.
2.1.2. OVS-2 tubes, which are specially made 13-mm o.d. glass tubes that are tapered to
6 mm o.d., packed with a 140-mg backup section, a 270-mg sampling section of cleaned XAD-2
adsorbent and a 13-mm diameter glass fiber filter. The backup section is retained by two
foam plugs and the sampling section is between one foam plug and the glass fiber filter.
The glass fiber filter is held next to the sampling section by a polytetrafluoroethylene
(PTFE) retainer. (Figure 2.)
2.2. Reagents
No sampling reagents are required.
2.3. Sampling technique
2.3.1. Immediately before sampling, remove the plastic caps from the OVS-2 tube.
2.3.2. Attach the small end of the tube to the sampling pump with flexible tubing.
2.3.3. Attach the tube vertically in the employee's breathing zone in such a manner
that it does not impede work performance.
2.3.4. After sampling for the appropriate time, remove the tube and seal with plastic
caps.
2.3.5. Wrap each sample end-to-end with an OSHA seal (Form 21).
2.3.6. Record the air volume for each sample, and list any possible interferences.
2.3.7. Submit at least one blank for each set of samples. Handle the blank in the same
manner as the samples, except no air is drawn through it.
2.3.8. Submit bulk samples for analysis in a separate container. Do not ship with air
samples.
2.4. Desorption efficiency (glass fiber filter and XAD-2 adsorbent)
Six vials each containing a 13-mm glass fiber filter and 270-mg of XAD-2 adsorbent were
each liquid spiked on the glass fiber filter with 50 µL of a 2.16 mg/mL solution of
bendiocarb. These samples were allowed to equilibrate overnight in a drawer at ambient
temperature. The next day each sample was desorbed with 2.0 mL of acetonitrile, shaken for
30 min and analyzed as in Section 3. The results are listed in Table 2.4.
Table 2.4.
Desorption Efficiency
|
Amount |
Amount |
% |
| |
| Sample # |
Spiked, µg |
Found, µg |
Recovered |
| |
| Ex1 |
108 |
99.37 |
92.0 |
| |
| Ex2 |
108 |
100.24 |
92.8 |
| |
| Ex3 |
108 |
94.53 |
87.5 |
| |
| Ex4 |
108 |
96.71 |
89.5 |
| |
| Ex5 |
108 |
97.94 |
90.7 |
| |
| Ex6 |
108 |
95.81 |
88.7 |
| |
|
| |
|
|
Average = |
90.2 |
2.5. Retention efficiency
Eighteen OVS-2 tubes were each liquid spiked with 50 µL of a 2.16 mg/mL solution of
bendiocarb by spiking the glass fiber filter. These were allowed to equilibrate overnight
in a drawer at ambient temperature. The next day 240 L of humid air (~80% relative
humidity) were drawn through each tube at 1 L/min. Six of the tubes were each desorbed
with 2.0 mL of acetonitrile, shaken for 30 min and then analyzed as in Section 3. The
results are listed in Table 2.5. No bendiocarb was found on the backup sections of these
tubes. The remaining samples were stored, six in a drawer at ambient temperature and six
in a freezer, for use in the storage study below.
Table 2.5.
Retention Efficiency
|
Amount |
Amount |
% |
| |
| Sample # |
Spiked, µg |
Found, µg |
Recovered |
| |
| R1 |
108 |
108.02 |
100.0 |
| |
| R2 |
108 |
103.23 |
95.6 |
| |
| R3 |
108 |
104.13 |
96.4 |
| |
| R4 |
108 |
102.32 |
94.7 |
| |
| R5 |
108 |
105.72 |
97.9 |
| |
| R6 |
108 |
107.41 |
99.5 |
| |
|
| |
|
|
Average = |
97.4 |
2.6. Sample storage
After four days of storage, six tubes were each desorbed with 2.0 mL of acetonitrile,
shaken for 30 min and then analyzed as in Section 3. Three of the tubes were from ambient
storage and the other three were from the freezer storage samples. The remaining tubes
were analyzed after eight days of storage. The results are given in Tables 2.6.1. and 2.6.2.
Table 2.6.1.
Ambient Storage
| Days |
Amount |
Amount |
% |
| |
| Stored |
Spiked, µg |
Found, µg |
Recovered |
| |
| 4 |
108 |
97.98 |
90.7 |
| |
| 4 |
108 |
104.71 |
97.0 |
| |
| 4 |
108 |
98.56 |
91.3 |
| |
| 8 |
108 |
96.54 |
89.4 |
| |
| 8 |
108 |
102.35 |
94.8 |
| |
| 8 |
108 |
86.04 |
79.7 |
| |
|
| |
|
|
Average of four days = |
93.0 |
| |
|
|
Average of eight days = |
88.0 |
Table 2.6.2.
Freezer Storage
| Days |
Amount |
Amount |
% |
| |
| Stored |
Spiked, µg |
Found, µg |
Recovered |
| |
| 4 |
108 |
94.55 |
87.5 |
| |
| 4 |
108 |
100.29 |
92.9 |
| |
| 4 |
108 |
99.73 |
92.3 |
| |
| 8 |
108 |
102.0 |
94.4 |
| |
| 8 |
108 |
103.46 |
95.8 |
| |
| 8 |
108 |
111.83 |
103.5 |
| |
|
| |
|
|
Average of four days = |
90.9 |
| |
|
|
Average of eight days = |
97.9 |
2.7. Recommended air volume and sampling rate
2.7.1. The recommended air volume is 240 L.
2.7.2. The recommended flow rate is 1.0 L/min.
2.8. Interferences (sampling)
It is not known if any compounds will interfere with the collec tion of bendiocarb. Any
suspected interferences should be reported to the laboratory.
2.9. Safety precautions (sampling)
2.9.1. Attach the sampling equipment in such a manner that it will not interfere with
work performance or employee safety.
2.9.2. Follow all safety practices that apply to the work area being sampled.